From: A novel mechanical lung model of pulmonary diseases to assist with teaching and training

Figure 2

Schematic drawing of model. The lung is represented by 6 bellows connected to a ventilator. A platform is placed on top of rubber bellow to support weights, used to vary the driving pressure of the bellow. A volume limiter sets the maximum height of the platform, thus limits the height which the bellow is allowed to expand vertically. Each bellow is connected to the larger common tube through an adjustable valve.