Skip to main content

Table 1 Technical validation of the EL-NE assay

From: Serological assessment of neutrophil elastase activity on elastin during lung ECM remodeling

Technical validation step

EL-NE

Detection range

0.90 nM – 119 nM

Lower limit of quantification

1.9 nM

Intra-assay variation

7%

Inter-assay variation

11%

Dilution range of serum samples

1:2 (recommended)

Dilution range of plasma samples

1:2 (recommended)

Dilution recovery in serum

104%

Dilution recovery in plasma (Heparin)

86%

Dilution recovery in plasma (Citrate)

108%

Freeze-thaw recovery in serum

96%

Freeze-thaw recovery in plasma (Heparin)

87%

Freeze-thaw recovery in plasma (Citrate)

104%

Analyte stability serum (24 h, 4°C/20°C)

88%/77%

Analyte stability plasma (Heparin) (24 h, 4°C/20°C)

95%/111%

Analyte stability plasma (Citrate) (24 h, 4°C/20°C)

85%/77%