Skip to main content

Table 3 Technical validation data of the DCN-CS ELISA assay

From: Cathepsin-S degraded decorin are elevated in fibrotic lung disorders – development and biological validation of a new serum biomarker

Tecnical validation step

DCN-CS performance

Detection range (LLOD-ULOD)

1.2–345.3 ng/mL

Lower limit of quantification (LLOQ)

5.3 ng/mL

Intra-assay variation

3%

Inter-assay variation

13%

Dilution of serum samples

1:4

Dilution recovery (1:4 pre-dilution)a

100% (82–113%)

Freeze/thaw recovery (4 cycles)a

94% (90–97%)

Analyte stability up to 24 h, 4 °Ca

87% (86–90%)

Interference Lipids, low/high

107%/86%

Interference Biotin, low/high

100%/100%

Interference Hemoglobin, low/high

100%/100%

  1. aPercentages are reported as mean with range shown in brackets