Table 2 Comparison of mean PIF and pressure drop values in overall population

From: Comparison of peak inspiratory flow rate via the Breezhaler®, Ellipta® and HandiHaler® dry powder inhalers in patients with moderate to very severe COPD: a randomized cross-over trial

Variable

Breezhaler®

Ellipta®

HandiHaler®

n

97

91

97

R (cmH2O0.5[L/min]−1)a

0.060

0.090

0.163

PIF (L/min)

108 ± 23

78 ± 15

49 ± 9

Range (Min-Max)

54–156

45–109

22–70

Pressure drop at PIF (cmH2O)

44 ± 18

51 ± 19

67 ± 23

ΔPIF vs Breezhaler® (L/min); 95% CI

30; 27 to 32

59; 56 to 62

P-value

 

< 0.0001

< 0.0001

  1. Data are presented as mean ± standard deviation unless stated otherwise. P-values generated from a paired t-test on comparison of PIF values
  2. Poor quality (flat-line) inhalational profiles with erroneous PIF values were not considered in the analysis
  3. CI confidence interval, n number of patients, ΔPIF difference in mean PIF values, PIF peak inspiratory flow rate, R intrinsic airflow resistance of the inhaler
  4. aData published previously [19, 20]