Fig. 4 | BMC Pulmonary Medicine

Fig. 4

From: Radial probe endobronchial ultrasound using a guide sheath for peripheral lung lesions in beginners

Fig. 4

Breakage of the guide sheath (GS). a Forceps biopsy via the GS was performed under fluoroscopic guidance after precise identification of the tumor using a radial probe EBUS (white arrow). b A kink in the GS (arrowhead) resulting in its dislocation was seen on fluoroscopy. The kink may have been caused by a discordance between the long axes of the bronchoscope (dotted line, a) and the GS (black line, a). c To prevent additional breakage of the GS, a thin bronchoscope was introduced as far as possible close to the target lesion (arrow). Thereafter, the two long axes of the bronchoscope and GS were aligned and the procedure was successfully completed