Skip to main content

Table 1 Calculated stage cut-off diameters (μm) for NGI at 20 L/min, 30 L/min, and 40 L/min

From: Aerosol drug delivery to the lungs during nasal high flow therapy: an in vitro study

Stage

20 L/min

Flow rate

30 L/mina

30 L/minb

40 L/min

1

13.05

11.70

14.59

10.03

2

7.61

6.40

7.90

5.51

3

4.76

3.99

4.88

3.45

4

2.84

2.30

2.78

2.01

5

1.74

1.36

1.68

1.17

6

1.11

0.83

1.06

0.70

7

0.77

0.54

0.71

0.45

  1. aNumbers based on Eq. 2
  2. bNumbers based on Eq. 1