Skip to main content

Table 2 Trigger results for shallow nasal breathing, proportion of pulses aligned with inhalation

From: Contemporary portable oxygen concentrators and diverse breathing behaviours -- a bench comparison

POC SettingProportion of POC pulses aligned with inspiration
Device ADevice BDevice C
142%100%99%
240%100%90%
342%100%51%
444%100%78%