Fig. 2From: Evaluation of time courses of agreement between minutely obtained transcutaneous blood gas data and the gold standard arterial data from spontaneously breathing Asian adults, and various subgroup analysesBlood gas data at 30 min (transcutaneous data were obtained via the chest or forearm sensors, n = 272). a Concordance between PaCO2 and PtcCO2 data at 30 min. The bias of PCO2 was 4.7 mmHg, with PtcCO2 being higher than PaCO2 and 95% limits of agreement of ±6.3 mmHg. b Concordance between PaO2 and PtcO2 data at 30 min. The bias of PO2 was 12.2 mmHg, with PaO2 being higher than PtcO2 and 95% limits of agreement of ±23.1 mmHgBack to article page