Skip to main content

Table 3 Procedural details and equipment information of ECMO

From: Complications and mortality of venovenous extracorporeal membrane oxygenation in the treatment of neonatal respiratory failure: a systematic review and meta-analysis

Study

VV ECMO (%)

VA ECMO (%)

VV ECMO convert to VA ECMO

ECMO duration (hours)

Site of insertion

Maximum cannula size

Oxygenator type

Cannula type

Pump type

Speggiorin et al. [17]

100%

0

0

90.5

Right internal jugular vein

16Fr

Polymethylpentene hollow fiber membrane

Double-lumen venous cannula

Ccentrifugal pump

Kugelman et al. [18]

100%

0

2

88.5

Right internal jugular vein

14Fr

NA

Double-lumen venous cannula

NA

Knight et al. [20]

100%

0

0

114 ± 9

Right internal jugular vein

14Fr

NA

Double-lumen venous cannula

NA

Chevalier et al. [19]

95.3%

4.7%

5

117.8 ± 84

Right internal jugular vein

10Fr

NA

Double-lumen venous cannula

Non-occlusive roller pump

  1. VV ECMO Venovenous extracorporeal membrane oxygenation, VA ECMO Venoarterial extracorporeal membrane oxygenation, NA Not available