From: Performance of a new single-use bronchoscope versus a marketed single-use comparator: a bench study
Comparison of aScope 4 size versus EXALT Model B size | Mean mass of “mucus” suctioned in 30 s (g) | P value | |
---|---|---|---|
aScope 4 | EXALT Model B | ||
Large versus Regulara | 41.5 ± 5.7 | 41.8 ± 8.3 | 0.914 |
Slim versus Slim | 1.5 ± 0.6b | 4.3 ± 1.3 | < .001 |
Regular versus Regular | 15.4 ± 2.9 | 41.8 ± 8.3 | < .001 |
Large versus Large | 41.5 ± 5.7 | 72.9 ± 9.0 | < .001 |