Skip to main content

Table 2 Primary endpoint: “mucus” mass suctioned in 30 s in Task 2 (N = 15)

From: Performance of a new single-use bronchoscope versus a marketed single-use comparator: a bench study

Comparison of aScope 4 size versus EXALT Model B size

Mean mass of “mucus” suctioned in 30 s (g)

P value

aScope 4

EXALT Model B

Large versus Regulara

41.5 ± 5.7

41.8 ± 8.3

0.914

Slim versus Slim

1.5 ± 0.6b

4.3 ± 1.3

 < .001

Regular versus Regular

15.4 ± 2.9

41.8 ± 8.3

 < .001

Large versus Large

41.5 ± 5.7

72.9 ± 9.0

 < .001

  1. aBased on a preliminary experiment, the study hypothesis was that the EXALT Model B regular scope would suction significantly more “mucus” than the aScope 4 large scope (not confirmed)
  2. bThere were 14 observations for aScope 4 Slim because one physician was called away for patient care and could not complete this task