Skip to main content

Table 4 Visualization ratings (N = 15)

From: Performance of a new single-use bronchoscope versus a marketed single-use comparator: a bench study

Attribute

Median rating (range)

P value

aScope 4 / aView 2 Advance monitor

EXALT model B/EXALT monitor

Definition

8.0 (7.0−10.0)

9.0 (8.0−10.0)

0.001

Color

8.0 (7.0−10.0)

9.0 (7.0−10.0)

0.009

White balance

8.0 (6.0−10.0)

9.0 (7.0−10.0)

0.029

Far field resolution

7.0 (6.0−10.0)

9.0 (7.0−10.0)

0.001

Near field resolution

9.0 (7.0−10.0)

10.0 (8.0−10.0)

0.018