Table 2 Reliability analysis (intraclass correlation coefficient for individual measures) for the Vt estimation in cycles with excess of leaks during inspiratory and expiratory cycle

From: Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench study

 

Excess of inspiratory leak

Excess of expiratory leak

Software

ICC

95% CI

p value

ICC

95% CI

p value

Trilogy

0.938

0.852-0.975

<0.001

0.840

0.481-0.958

<0.001

PB 560

0.687

0.362-0.863

<0.001

0.976

0.908-0.994

<0.001

Stellar

0.932

0.837-0.973

<0.001

0.969

0.882.-0.992

<0.001

Vivo 50

0.652

0.306-0.846

<0.01

0.713

0.492-0.929

<0.01

BetterCare

0.972

0.930-0.989

<0.001

0.997

0.986-999

<0.001