Skip to main content

Table 6 Frequency and current in relation to the skin temperature

From: Pulsed direct and constant direct currents in the pilocarpine iontophoresis sweat chloride test

Frequency (Hz)*

Current#

Skin temperature

Mean

Standard deviation

0 (CDC)

0.5

Before

30.9

0.4

 

After

29.3

1.2

0.75

Before

31.0

0.3

 

After

30.6

1.5

1

Before

30.5

0.9

 

After

30.4

0.7

1.5

Before

31.0

0.3

 

After

30.6

1.5

200

0.5

Before

30.9

1.2

 

After

29.6

0.4

0.75

Before

31.5

3.2

 

After

29.6

1.4

1

Before

31.6

2.3

 

After

31.2

1.8

1.5

Before

29.4

1.8

 

After

28.0

1.4

1,000

0.5

Before

31.8

0.9

 

After

30.7

0.1

0.75

Before

31.3

0.8

 

After

30.1

0.8

1

Before

30.4

2.3

 

After

31.3

1.0

1.5

Before

30.8

1.0

 

After

31.1

2.0

5,000

0.5

Before

31.4

1.6

 

After

29.3

1.4

0.75

Before

30.2

0.8

 

After

29.9

0.7

1

Before

31.6

2.1

 

After

30.3

2.1

1.5

Before

32.8

1.5

 

After

31.4

0.5

  1. CDC, constant direct current; Hz, Hertz, mA, milliampere. *In the analysis considering the frequency in relation to skin temperature, p-value = 0.4102; #In the analysis considering the current in relation to skin temperature, p-value = 0.5361; for interaction between current and frequency in relation to skin temperature, p-value = 0.0884. The skin temperature before the experiment was higher than after, p-value <0.0001.