Table 3 Diagnostic performance of SpO2 and SaO2

From: A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals

Ability for SpO2 <90% to detect SaO2 <90%

 

SaO2 <90%

Sensitivity: 88.6%

 SpO2 <90%

Yes

No

Specificity: 95.1%

 Yes

31

18

 No

4

347

 Total

35

365

Ability for SpO2 <92% to detect SaO2 <90%

 

SaO2 <90%

Sensitivity: 100%

 SpO2 <92%

Yes

No

Specificity: 84.4%

 Yes

35

57

 No

0

308

 Total

35

365

Ability for SpO2 <90% to detect PaO2 <60 mmHg

 

PaO2 <60 mmHg

Sensitivity: 70.5%

 SpO2 <90%

Yes

No

Specificity: 98.2%

 Yes

43

6

 No

18

333

 Total

61

339

Ability for SaO2 <90% to detect PaO2 <60 mmHg

 

PaO2 <60 mmHg

Sensitivity: 54.1%

 SaO2 <90%

Yes

No

Specificity: 99.4%

 Yes

33

2

 No

28

337

 Total

61

339

  1. PaO2: Partial pressure of oxygen, SaO2: Oxygen saturation measured by arterial blood gas sample, SpO2: Oxygen saturation measured by standard pulse oximeter