Table 1 Results of the leak test and volume accuracy test for 16 volume calibration syringes

From: Quality inspection and result analysis of the spirometer calibration cylinder

The evaluation index

NEW-D1

NEW-D2

NEW-D3

NEW-D4

NEW-D5

NEW-D6

NEW-D7

NEW-D8

OLD-D1

OLD-D2

OLD-D3

OLD-D4

OLD-D5

OLD-D6

OLD-D7

OLD-D8

Overall pass rate/%

Use fixed number of years

0

0

0

0

0

0

0

0

2.5

2.5

1

1.5

3

3

2

5

 

Leak test

The target volume (mL)

1000.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

100.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

100.0

 

Objective to reveal (mL/min)

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

 

Actual average leakage (mL/min)

0.1

0.2

0.9

0.5

2.5

0.0

0.8

0.3

0.3

0.1

0.4

0.2

1.0

1.9

0.1

75

Reveal the mean error value (mL/min)

0.1

0.2

0.9

0.5

2.5

0.0

0.8

0.3

0.3

0.1

0.4

0.2

1.0

1.9

0.1

75

Whether through (Y/N)

Y

Y

Y

Y

N

Y

Y

Y

Y

Y

Y

Y

N

N

N

Y

75

Volume accuracy test

The target volume (mL)

1000.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

100.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

3000.0

100.0

 

Average volume measurement (mL)

992.2

2994.7

2996.9

3003.0

2990.9

2995.7

3001.2

96.9

2994.4

2998.6

2992.5

2990.6

2995.4

2998.9

2979.2

92.1

88

Volume mean error range (mL)

0.5

0.4

0.6

0.5

0.8

1.4

0.4

0.4

1.5

0.6

1.0

1.2

0.4

1.8

13.5

0.9

69

Volume mean error (mL)

 − 7.8

 − 5.3

 − 3.1

3.0

 − 9.1

 − 4.3

1.2

 − 3.1

 − 5.6

1.4

 − 7.5

 − 9.4

 − 4.6

 − 1.1

 − 20.8

 − 7.9

88

Whether through (Y/N)

N

Y

Y

Y

Y

Y

Y

N

Y

Y

Y

Y

Y

N

N

N

69

Two tests whether through (Y/N)

N

Y

Y

Y

N

Y

Y

N

Y

Y

Y

Y

N

N

N

N

56

  1. “—” indicates that the leak test could not be completed due to serious gas leakage during the leak test