Anesthesia considerations to reduce motion and atelectasis during advanced guided bronchoscopy

Abstract

Partnership between anesthesia providers and proceduralists is essential to ensure patient safety and optimize outcomes. A renewed importance of this axiom has emerged in advanced bronchoscopy and interventional pulmonology. While anesthesia-induced atelectasis is common, it is not typically clinically significant. Advanced guided bronchoscopic biopsy is an exception in which anesthesia protocols substantially impact outcomes. Procedure success depends on careful ventilation to avoid excessive motion, reduce distortion causing computed tomography (CT)-to-body-divergence, stabilize dependent areas, and optimize breath-hold maneuvers to prevent atelectasis. Herein are anesthesia recommendations during guided bronchoscopy. An FiO2 of 0.6 to 0.8 is recommended for pre-oxygenation, maintained at the lowest tolerable level for the entire the procedure. Expeditious intubation (not rapid-sequence) with a larger endotracheal tube and non-depolarizing muscle relaxants are preferred. Positive end-expiratory pressure (PEEP) of up to 10–12 cm H2O and increased tidal volumes help to maintain optimal lung inflation, if tolerated by the patient as determined during recruitment. A breath-hold is required to reduce motion artifact during intraprocedural imaging (e.g., cone-beam CT, digital tomosynthesis), timed at the end of a normal tidal breath (peak inspiration) and held until pressures equilibrate and the imaging cycle is complete. Use of the adjustable pressure-limiting valve is critical to maintain the desired PEEP and reduce movement during breath-hold maneuvers. These measures will reduce atelectasis and CT-to-body divergence, minimize motion artifact, and provide clearer, more accurate images during guided bronchoscopy. Following these recommendations will facilitate a successful lung biopsy, potentially accelerating the time to treatment by avoiding additional biopsies. Application of these methods should be at the discretion of the anesthesiologist and the proceduralist; best medical judgement should be used in all cases to ensure the safety of the patient.

Peer Review reports

Background

Advanced guided bronchoscopic biopsy is a minimally invasive method to diagnose suspicious lung lesions [1, 2]. Guidance modalities include not only intraprocedural image-guided methods (e.g., augmented fluoroscopy, tomosynthesis-based fluoroscopic navigation), but also electromagnetic navigation bronchoscopy (ENB), electromagnetic-based and fiber-optic shape-sensing robotic bronchoscopy, virtual bronchoscopic navigation (VBN), radial endobronchial ultrasound (EBUS), and cone-beam computed tomography (CBCT) [1, 2]. Guided bronchoscopy has a lower risk of complications than transthoracic biopsy [3, 4], and is particularly useful for small lesions in the periphery of the lung that cannot be reached by traditional bronchoscopy [5]. Early diagnosis and treatment of malignant lung lesions can substantially increase survival rates [6, 7].

Despite these advantages, the diagnostic yield of all guided bronchoscopy systems has historically been limited by a phenomenon called computed tomography (CT)-to-body divergence [8]. This effect refers to a mismatch between the preprocedure CT-based virtual map used to plan the navigation route to the lung lesion and the actual dynamic lung anatomy during bronchoscopy. This divergence can be caused by a number of factors [8] and has been addressed, in part, by advanced bronchoscopy systems that provide real-time visualization and/or positional correction during the bronchoscopy procedure [9,10,11,12,13,14,15,16,17,18]. However, anesthesia-induced atelectasis remains a challenge that can obscure lesion visibility and cause inaccurate localization, even with intraprocedural positional correction.

While it has long been known that nearly all patients experience atelectasis within minutes after general anesthesia induction [19,20,21,22,23], atelectasis during bronchoscopy and its impact on outcomes has been underappreciated until recently. Because atelectasis is not visible on standard fluoroscopy [24, 25], its prevalence during bronchoscopy was not realized until the use of CBCT became more common. Using CBCT, Casal et al. were the first to report atelectasis in dependent areas in 40% of peripheral lung biopsies, completely obscuring the target in 20% of cases [24]. Avasarala et al. reported atelectasis in 75% of patients, obscuring the lesion in 38% [26]. In the first prospective study specifically designed to assess atelectasis during bronchoscopy (I-LOCATE), Sagar et al. reported that 89% of patients had atelectasis in at least one bronchial segment with a prevalence greater than 50% in dependent lower lobe segments. Atelectasis was observed within 30 min of anesthesia induction. Increased body mass index and time to atelectasis assessment were significant predictors of atelectasis risk [25]. General anesthesia was used with a laryngeal mask airway (LMA) and neuromuscular blocking agents, and most patients were ventilated with 100% fraction of inspired oxygen (FiO2) and zero to minimal PEEP [25].

Atelectasis can have a significant impact on the success of guided bronchoscopy. Atelectasis decreases the visibility of lung lesions on imaging and changes the conformation of the airways (Fig. 1), exacerbating any mismatch between the virtual lung map and the patient’s anatomy. Atelectasis reduces the distance between the lesion and the pleura, increasing the risk of pneumothorax from instrumentation. Because of its increased density compared to aerated lung, atelectasis can also create false positive findings on radial EBUS that mimic a lung lesion [25]. The result is inaccurate localization and reduced diagnostic yield [27]. Significant changes in inflation pressure are required to open atelectatic lung units. Because recruitment maneuvers to temporarily increase PEEP and tidal volumes often fail to resolve the atelectasis once it has appeared [24], it is important to prevent atelectasis from occurring and maintain optimal conditions throughout the bronchoscopic navigation and biopsy procedures.

Fig. 1
figure 1

AC Computed tomography (CT) and cone-beam CT (CBCT) scans during image-guided bronchoscopy without an optimized ventilation protocol. Significant atelectasis and ghosting artifact were observed. DF CT and CBCT scans from a different patient using a ventilation protocol designed to prevent atelectasis

Literature review

A systematic literature search of MEDLINE and Embase was conducted for papers evaluating the impact of anesthesia methods on outcomes during advanced peripheral bronchoscopy (ENB, radial EBUS, VBN, CBCT, augmented fluoroscopy, robotic bronchoscopy). Detailed methods are provided in Additional file 1: Figure S1, Additional file 1: Table S1 and Additional file 1: Table S2. After abstract and full-text review, there were 11 papers (8 original studies and 3 meta-analyses) that specifically evaluated the impact of the anesthesia method on atelectasis, safety, or diagnostic yield during advanced bronchoscopy [4, 27,28,29,30,31,32,33,34,35,36]. Bowling et al. compared ENB using general anesthesia versus intravenous sedation and observed diagnostic yields of 70% and 78%, respectively (P = NS) [29]. As will be described in greater detail below, Bhadra et al. 2021 observed less atelectasis and a trend toward greater diagnostic yield using an optimized ventilation protocol compared to conventional ventilation [28]. Minami et al. (2017) reported that fentanyl and midazolam sedation was useful for peripheral bronchoscopy [32]. Webb et al. (2020) reported that jet ventilation was associated with reduced target displacement and increased diagnostic yield compared to traditional ventilation during ENB [35]. General anesthesia and conscious sedation were used with approximately equal frequency across studies in a 2020 meta-analysis with no significant differences in diagnostic yield [4], consistent with results from the AQuIRE registry [33], a 2015 meta-analysis [36], and the NAVIGATE study [30]. One 2014 meta-analysis observed significantly higher diagnostic yield with general anesthesia use compared to conscious sedation [31] and the NAVIGATE study found a significant effect of general anesthesia versus moderate sedation on complication rates [34]. Finally, Tanner et al. found no significant difference in diagnostic yield using moderate versus deep sedation using standard bronchoscopy or radial EBUS [27].

However, with one exception [28], the studies described above used navigation bronchoscopy systems without intraprocedural advanced imaging. Newer systems such as tomosynthesis-based fluoroscopic navigation [10] and CT augmented fluoroscopy [28] require a breath hold to minimize lung movement during intraprocedural imaging and location adjustments. Furthermore, CT-to-body divergence and atelectasis are likely strong contributors to the sub-optimal diagnostic yield seen with earlier generations of navigation bronchoscopy technology in large multicenter studies [30].

Ventilation recommendations

The following is a recommended protocol for ventilation during advanced guided bronchoscopic biopsy (Table 1). These recommendations were derived from the anesthesia literature regarding the principles of minimizing atelectasis during positive pressure ventilation [37,38,39,40,41,42], the authors’ published work [28], and the authors’ combined experience of over 1000 CBCT navigation cases assessing the presences of atelectasis and CT-to-body divergence. While the use of higher PEEP and lower FiO2 during induction may contradict traditional anesthesiology practices, they are important for guided bronchoscopy where the success of the biopsy is highly dependent upon stabilizing dependent areas of the lung during intraprocedural imaging, particularly in obese patients and those with posterior or lower lobe lesions [43, 44]. These techniques will help to reduce movement and minimize atelectasis, thereby increasing the chances of a successful biopsy and potentially accelerating the time to treatment by reducing the need for repeat biopsy procedures. These methods should be used at the discretion of the anesthesiologist and the proceduralist based on the individual risk factors of each patient. Standard monitoring procedures should be followed according to the American Society of Anesthesiologists (ASA) recommendations.

Table 1 Anesthesia for advanced guided bronchoscopy

Preprocedure incentive spirometry

Preprocedure incentive spirometry has been recommended to recruit lung volume and prevent atelectasis [45]. The incentive spirometry maximum value will also allow the anesthesiologist to anticipate any risks of the higher intraprocedural tidal volumes recommended below. This is especially important for patients at particular risk of intraprocedural atelectasis, such as obese patients or those with lesions in the dependent areas of the lung.

Preoxygenation

The benefits of pre-oxygenation prior to anesthetic induction and tracheal intubation are widely accepted. However, nitrogen washout during pre-oxygenation promotes loss of gas from the lung to the bloodstream, resulting in alveolar collapse and absorption atelectasis. Gas absorption accelerates airway collapse by a combination of decreased functional residual capacity and compression atelectasis [46]. Therefore, the use of 100% oxygen during induction and anesthesia maintenance is a major cause of atelectasis [41]. Given the risk of absorption atelectasis and lung injury, the lowest tolerable FiO2 is recommended for preoxygenation [46] as guided by oxygen saturation. Edmark et al. demonstrated that in most patients, an FiO2 of 0.8 to 0.6 was associated with little to no atelectasis compared to an FiO2 of 1.0 [37]. Inspired oxygen content should also be kept as low as tolerated during the procedure to minimize absorption atelectasis.

If it is not possible to decrease the FiO2 to less than 1.0 during pre-induction, the FiO2 should be kept at the lowest tolerable level immediately after the endotracheal tube (ETT) cuff is inflated, and recruitment maneuvers should be performed with the lowered FiO2. This lowered FiO2 should be maintained for the remainder of the procedure, including during the biopsy, as the patient’s oxygen saturation permits.

Anesthesia type and intubation

The need for a completely motionless patient during guided bronchoscopy necessitates general anesthesia, paralysis, and intubation. The largest ETT size feasible based on patient anatomy will enable gas passage around the bronchoscope (which ranges in diameter from 5.9 to 6.3 mm) with the least increase in circuit pressure [47]. An ETT is preferred over an LMA to accommodate higher airway pressures (larger tidal volumes and increased PEEP). LMA use may increase the risk of gastric insufflation with higher pressures. This in turn may increase the risk for aspiration, although the risk remains low based on previous studies [48]. Following bronchoscopy, the ETT can be exchanged for an LMA to facilitate complete EBUS staging if desired. LMA may be feasible in smaller patients (e.g., < 80 kg) provided that a good seal can be obtained whilst providing the recommended tidal volumes and PEEP. Jet ventilation has also been used to minimize motion and CT-to-body divergence [35, 49], but may be limited by availability and expertise.

General anesthesia using total intravenous anesthesia (TIVA) with propofol and muscle paralysis is optimal [47]. Lengthy intubation times may increase the risk of atelectasis, thus, avoid “traditional” rapid-sequence intubation (i.e., avoid the use of FiO2 of 1.0). Instead, perform an expeditious intubation using non-depolarizing muscle relaxants rather than suxamethonium. Application of PEEP throughout induction has also been shown to prevent atelectasis [42]. Whilst the use of volatile anesthetics is not contraindicated, repeated opening of the circuit for the passage of the scope poses additional challenges to maintaining the depth of anesthesia and volatile gas pollution in the operating room.

Recruitment maneuvers

Reducing the FiO2 to the lowest tolerable percentage after intubation will improve airway visualization. Unless contraindicated (e.g., due to acute respiratory distress syndrome, ventilator-induced lung injury, recent segmentectomy, structural lung disease, or surgery), conduct up to four recruitment maneuvers immediately after intubation [46] to reverse any intubation atelectasis and to assess hemodynamic stability during higher PEEP (Fig. 2). This is especially important if intubation was prolonged due to a difficult intubation. Hemodynamic instability may preclude the use of traditional recruitment maneuvers, such as the use of 40 cm H2O for 40 s (the ‘40 for 40 hold’). Judicious use of PEEP from the pre-induction phase and throughout the procedure is recommended [21, 38, 42], but may vary based on the patient’s hemodynamic tolerance. Maintain higher PEEP with the lowest tolerable FiO2 as guided by oxygen saturation. PEEP of up to 10–12 cm H2O may be beneficial for upper lobe biopsies, and even higher PEEP may be required for lower lobe biopsies (particularly in obese patients) due to the lower functional residual capacity and increased risk of atelectasis in the dependent areas. An increase in tidal volumes may be considered if tolerated.

Fig. 2
figure 2

Recruitment maneuvers after intubation may reverse any induction-related atelectasis and assess hemodynamic stability during higher PEEP. Hemodynamic instability may limit use of the traditional ‘40 for 40’ hold. Higher PEEP for shorter duration may be considered

Use of higher PEEP and tidal volumes will help maintain optimal lung inflation [44]. However, use of these higher settings should be guided by what the patient is able to tolerate hemodynamically, as demonstrated in the post-intubation recruitment maneuvers. This is especially important in obese patients or those with structural lung disease who may be at increased risk of barotrauma or volutrauma. Also, the risk of barotrauma is minimized when the sheer stress of repeated alveolar re-expansion is avoided, and the alveoli are held on a favorable part of the compliance curve (Fig. 3). While intuitively contradictory to the classical teaching of low tidal volume to prevent barotrauma, airway visualization, accurate lesion localization, and successful biopsy require these preprocedural conditions to be replicated as closely as possible.

Fig. 3
figure 3

Mechanisms leading to atelectasis during bronchoscopy and increased airway resistance. A Surface tension caused by water molecules leads to attraction. B Once the airflow is disrupted in atelectatic lung units, turbulent airflow leads to increased airway resistance. C As the bronchi decrease in diameter, there is a substantial increase in airways resistance. D Compliance curve for alveoli. As the lung deflates, resistance increases and gas flow through the airways decreases. Atelectatic lung units have poor compliance and require significant changes in inflation pressure to result in minor changes in volume

Breath-hold for image acquisition

Timing the breath-hold

A breath-hold is required to reduce motion artifact and provide clearer, more accurate images during certain image-guided bronchoscopy procedures (e.g., fluoroscopic navigation, CBCT). These procedures require a complete lack of movement beyond cardiac pulsations; therefore, a carefully timed breath-hold is essential. The breath-hold should be performed at peak inspiration, not at end-expiration. This does not require a vital capacity maneuver, rather, it should occur at the end of a normal inspiratory breath (Fig. 4).

Fig. 4
figure 4

A breath-hold is required to reduce motion artifact during intraprocedural imaging (e.g., CBCT, digital tomosynthesis). The breath-hold should be performed at peak inspiration, not at end-expiration. This does not require a vital capacity maneuver but should occur at the end of a normal tidal breath. Adapted from Kapwatt at English Wikipedia (https://commons.wikimedia.org/w/index.php?curid=74891988) and used with permission under the terms of Creative Commons License CC BY-SA 3.0 ()

Maintaining a constant circuit pressure and PEEP

The breath-hold may be achieved by either the anesthesia machine’s automatic feature or manually. The automatic feature sets and maintains the appropriate pressure using the Vital Capacity Hold Function. In the manual method, the breath-hold is initiated by switching the ventilator to manual mode. Adjusting the adjustable pressure-limiting (APL) valve during breath hold maneuvers is required to maintain a constant circuit pressure and PEEP and reduce diaphragmatic movement once a leak is created in the circuit with the passage of the bronchoscope. The measured circuit pressure reflects intrapulmonary pressure (PEEP) and is a balance between fresh gas inflow and leak from the circuit plus outflow via the APL. Therefore, it’s preferable to focus on the circuit pressure reading (shown either on the digital screen or on the circuit pressure manometer) and adjust the APL manually to maintain this pressure at the designated PEEP, rather than on an absolute APL number. Vigilance is required to ensure that there is a balance of fresh gas inflow to circuit leak. These measures will minimize motion artifact of the diaphragm during imaging which can cause blurring and thereby hinder image interpretation by the clinician.

Maintaining the breath-hold

The breath-hold should be maintained for a sufficient time to allow pressures to plateau and equilibrate throughout the bronchial tree prior to beginning intraprocedural imaging (Fig. 5). The proceduralist should wait 5–10 s before beginning the imaging sweep after the anesthesiologist initiates the breath-hold (once PEEP is stabilized as measured by constant circuit pressure). This gives time for the intrapulmonary pressures to equilibrate before the imaging sweep begins. These measures will remove motion artifact, provide clearer images, and allow for more accurate localization of the lung lesion. By maintaining high PEEP, the bronchial tree is also optimally dilated. The imaging sweep will add additional time to the breath-hold (up to 30 s, depending on the system used). Therefore, it is important to be aware of the potential for hemodynamic changes due to prolonged breath-hold during the equilibration and imaging sweep.

Fig. 5
figure 5

To prevent motion artifact, maintain the breath-hold until pressures plateau before beginning the imaging sweep (5–10 s). Be aware of the potential for hemodynamic changes due to prolonged breath-hold

Biopsy procedure

The recommended anesthesia settings described above should be maintained throughout the biopsy procedure to ensure that lung volumes are consistent with those present during intraprocedural imaging. Any changes during the biopsy procedure would necessitate a repeat of both the breath-hold and the imaging sweep.

Post-procedure

Routine reversal and standard assessment of fitness for extubation should be employed, as should suctioning of secretions. Standard post-procedural discharge criteria should be employed, and a postprocedure chest x-ray is recommended to exclude complications such as pneumothorax. Patients should be completely reversed from neuromuscular paralysis.

Clinical evidence

The proposed ventilation protocol has been evaluated in a retrospective, single-center study of subjects with peripheral lung lesions < 30 mm undergoing CBCT with augmented fluoroscopy [28]. In two non-randomized, consecutively enrolled groups (25 subjects per group), a conventional, non-standardized ventilation protocol (typically 100% FiO2 and PEEP set to 0 or 5 cm H2O) was compared to the optimized ventilation protocol. Two independent reviewers observed significantly less atelectasis in dependent areas (P = 0.0001) and in sublobar or lobar regions (P = 0.01) using the optimized protocol. There was also a significantly smaller proportion of lesions obscured by atelectasis using the optimized protocol (8% Reviewer 1, 4% Reviewer 2) than the conventional protocol (36% observed by both reviewers). Furthermore, this significant reduction in atelectasis was accompanied by a higher diagnostic yield, though not statistically significant (70% conventional versus 92% optimized, P = 0.08). Pneumothorax not requiring a chest tube was observed in 1 patient following the optimized protocol compared to 0 patients with conventional ventilation (P = 1.0). A study of tomosynthesis-based fluoroscopic navigation used a similar ventilation protocol, with intubation under TIVA, neuromuscular blockade, a recruitment maneuver after intubation, minimized oxygen, and 15 cm H2O applied throughout the procedure. A diagnostic yield of 77% was observed, with a pneumothorax rate of only 2.5% (8/324) [10]. While future randomized studies are necessary to confirm these results, this data suggests that a ventilation protocol optimized for advanced guided bronchoscopy not only reduces atelectasis, but may also increase in diagnostic yield. The randomized VESPA trial is (NCT04311723) currently enrolling and aims to compare a ventilation strategy designed to prevent atelectasis during bronchoscopy compared to conventional mechanical ventilation [50].

Conclusions

Atelectasis is common during advanced guided bronchoscopic biopsy [25] and can cause CT-to-body divergence [8], interfering with the ability to obtain diagnostic tissue [28]. The anesthesia techniques proposed in this paper are designed to stabilize dependent areas of the lung, reduce atelectasis, minimize motion artifact, and provide more accurate localization during guided bronchoscopy. Future randomized studies are needed to prospectively compare bronchoscopy procedures with and without optimized ventilation strategies.

Availability of data and materials

Not applicable.

Abbreviations

APL:

Adjustable pressure-limiting (valve)

ASA:

American Society of Anesthesiologists

CBCT:

Cone-beam computed tomography

CT:

Computed tomography

EBUS:

Endobronchial ultrasound

ETT:

Endotracheal tube

FiO2 :

Fraction of inspired oxygen

LMA:

Laryngeal mask airway

PEEP:

Positive end-expiratory pressure

TIVA:

Total intravenous anesthesia

References

  1. Cicenia J, Avasarala SK, Gildea TR. Navigational bronchoscopy: a guide through history, current use, and developing technology. J Thorac Dis. 2020;12:3263–71.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Katsis JM, Rickman OB, Maldonado F, Lentz RJ. Bronchoscopic biopsy of peripheral pulmonary lesions in 2020: a review of existing technologies. J Thorac Dis. 2020;12:3253–62.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Heerink WJ, de Bock GH, de Jonge GJ, Groen HJ, Vliegenthart R, Oudkerk M. Complication rates of CT-guided transthoracic lung biopsy: meta-analysis. Eur Radiol. 2017;27:138–48.

    Article  CAS  PubMed  Google Scholar 

  4. Folch EE, Labarca G, Ospina-Delgado D, Kheir F, Majid A, Khandhar SJ, et al. Sensitivity and safety of electromagnetic navigation bronchoscopy for lung cancer diagnosis: systematic review and meta-analysis. Chest. 2020;158:1753–69.

    Article  PubMed  Google Scholar 

  5. Rivera MP, Mehta AC, Wahidi MM. Establishing the diagnosis of lung cancer: diagnosis and management of lung cancer, 3rd edition: American College of Chest Physicians evidence-based clinical practice guidelines. Chest. 2013;143:e142S-e165S.

    Article  PubMed  Google Scholar 

  6. Siegel RL, Miller KD, Fuchs HE, Jemal A. Cancer statistics, 2021. CA Cancer J Clin. 2021;71:7–33.

    Article  PubMed  Google Scholar 

  7. Henschke CI, Yankelevitz DF, Libby DM, Pasmantier MW, Smith JP, Miettinen OS. Survival of patients with stage I lung cancer detected on CT screening. N Engl J Med. 2006;355:1763–71.

    Article  PubMed  Google Scholar 

  8. Pritchett MA, Bhadra K, Calcutt M, Folch E. Virtual or reality: divergence between preprocedural computed tomography scans and lung anatomy during guided bronchoscopy. J Thorac Dis. 2020;12:1595–611.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Katsis J, Roller L, Lester M, Johnson J, Lentz R, Rickman O, et al. High Accuracy of digital tomosynthesis-guided bronchoscopic biopsy confirmed by intraprocedural computed tomography. Respiration. 2021;100:1–8.

    Article  Google Scholar 

  10. Katsis J, Roller L, Aboudara M, Pannu J, Chen H, Johnson J, et al. Diagnostic yield of digital tomosynthesis-assisted navigational bronchoscopy for indeterminate lung nodules. J Bronchol Interv Pulmonol. 2021. https://doi.org/10.1097/LBR.0000000000000766.

    Article  Google Scholar 

  11. Pritchett MA, Bhadra K, Mattingley JS. Electromagnetic navigation bronchoscopy with tomosynthesis-based visualization and positional correction: three-dimensional accuracy as confirmed by cone-beam computed tomography. J Bronchol Interv Pulmonol. 2021;28:10–20.

    Article  Google Scholar 

  12. Pritchett MA. Prospective analysis of a novel endobronchial augmented fluoroscopic navigation system for diagnosis of peripheral pulmonary lesions. J Bronchol Interv Pulmonol. 2021;28:107–15.

    Article  Google Scholar 

  13. Thiboutot J, Lee HJ, Silvestri GA, Chen A, Wahidi MM, Gilbert CR, et al. Study design and rationale: a multicenter, prospective trial of electromagnetic bronchoscopic and electromagnetic transthoracic navigational approaches for the biopsy of peripheral pulmonary nodules (ALL IN ONE Trial). Contemp Clin Trials. 2018;71:88–95.

    Article  PubMed  Google Scholar 

  14. Fielding DIK, Bashirzadeh F, Son JH, Todman M, Chin A, Tan L, et al. First human use of a new robotic-assisted fiber optic sensing navigation system for small peripheral pulmonary nodules. Respiration. 2019;98:142–50.

    Article  PubMed  Google Scholar 

  15. Benn BS, Romero AO, Lum M, Krishna G. Robotic-assisted navigation bronchoscopy as a paradigm shift in peripheral lung access. Lung. 2021;199:177–86.

    Article  PubMed  Google Scholar 

  16. Chen AC, Pastis NJ Jr, Mahajan AK, Khandhar SJ, Simoff MJ, Machuzak MS, et al. Robotic bronchoscopy for peripheral pulmonary lesions: a multicenter pilot and feasibility study (BENEFIT). Chest. 2021;159:845–52.

    Article  PubMed  Google Scholar 

  17. Chaddha U, Kovacs SP, Manley C, Hogarth DK, Cumbo-Nacheli G, Bhavani SV, et al. Robot-assisted bronchoscopy for pulmonary lesion diagnosis: results from the initial multicenter experience. BMC Pulm Med. 2019;19:243.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Cicenia J, Bhadra K, Sethi S, Nader DA, Whitten P, Hogarth DK. Augmented fluoroscopy: a new and novel navigation platform for peripheral bronchoscopy. J Bronchol Interv Pulmonol. 2021;28:116–23.

    Article  Google Scholar 

  19. Gamsu G, Singer MM, Vincent HH, Berry S, Nadel JA. Postoperative impairment of mucous transport in the lung. Am Rev Respir Dis. 1976;114:673–9.

    CAS  PubMed  Google Scholar 

  20. Strandberg A, Tokics L, Brismar B, Lundquist H, Hedenstierna G. Atelectasis during anaesthesia and in the postoperative period. Acta Anaesthesiol Scand. 1986;30:154–8.

    Article  CAS  PubMed  Google Scholar 

  21. Brismar B, Hedenstierna G, Lundquist H, Strandberg A, Svensson L, Tokics L. Pulmonary densities during anesthesia with muscular relaxation—a proposal of atelectasis. Anesthesiology. 1985;62:422–8.

    Article  CAS  PubMed  Google Scholar 

  22. Lundquist H, Hedenstierna G, Strandberg A, Tokics L, Brismar B. CT-assessment of dependent lung densities in man during general anaesthesia. Acta Radiol. 1995;36:626–32.

    Article  CAS  PubMed  Google Scholar 

  23. Duggan M, Kavanagh BP. Pulmonary atelectasis: a pathogenic perioperative entity. Anesthesiology. 2005;102:838–54.

    Article  PubMed  Google Scholar 

  24. Casal RF, Sarkiss M, Jones AK, Stewart J, Tam A, Grosu HB, et al. Cone beam computed tomography-guided thin/ultrathin bronchoscopy for diagnosis of peripheral lung nodules: a prospective pilot study. J Thorac Dis. 2018;10:6950–9.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Sagar AS, Sabath BF, Eapen GA, Song J, Marcoux M, Sarkiss M, et al. Incidence and location of atelectasis developed during bronchoscopy under general anesthesia: the I-LOCATE trial. Chest. 2020;158:2658–66.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Avasarala SK, Machuzak MS, Gildea TR. Multidimensional precision: hybrid mobile 2D/3D C-arm assisted biopsy of peripheral lung nodules. J Bronchol Interv Pulmonol. 2020;27:153–5.

    Article  Google Scholar 

  27. Tanner NT, Yarmus L, Chen A, Wang Memoli J, Mehta HJ, Pastis NJ, et al. Standard bronchoscopy with fluoroscopy versus thin bronchoscopy and radial endobronchial ultrasound for biopsy of pulmonary lesions: a multicenter, prospective. Randomized Trial Chest. 2018;154:1035–43.

    PubMed  Google Scholar 

  28. Bhadra K, Setser RM, Condra W, Pritchett MA. Lung navigation ventilation protocol to optimize biopsy of peripheral lung lesions. J Bronchol Interv Pulmonol. 2021. https://doi.org/10.1097/LBR.0000000000000756.

    Article  Google Scholar 

  29. Bowling MR, Kohan MW, Walker P, Efird J, Ben OS. The effect of general anesthesia versus intravenous sedation on diagnostic yield and success in electromagnetic navigation bronchoscopy. J Bronchol Interv Pulmonol. 2015;22:5–13.

    Article  Google Scholar 

  30. Folch EE, Pritchett MA, Nead MA, Bowling MR, Murgu SD, Krimsky WS, et al. Electromagnetic navigation bronchoscopy for peripheral pulmonary lesions: one-year results of the prospective, multicenter NAVIGATE study. J Thorac Oncol. 2019;14:445–58.

    Article  PubMed  Google Scholar 

  31. Gex G, Pralong JA, Combescure C, Seijo L, Rochat T, Soccal PM. Diagnostic yield and safety of electromagnetic navigation bronchoscopy for lung nodules: a systematic review and meta-analysis. Respiration. 2014;87:165–76.

    Article  PubMed  Google Scholar 

  32. Minami D, Nakasuka T, Ando C, Iwamoto MdY, Sato K, Fujiwara K, et al. Bronchoscopic diagnosis of peripheral pulmonary lung cancer employing sedation with fentanyl and midazolam. Respir Investig. 2017;55:314–7.

    Article  PubMed  Google Scholar 

  33. Ost DE, Ernst A, Lei X, Kovitz KL, Benzaquen S, Diaz-Mendoza J, et al. Diagnostic yield and complications of bronchoscopy for peripheral lung lesions: results of the AQuIRE registry. Am J Respir Crit Care Med. 2016;193:68–77.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Towe CW, Nead MA, Rickman OB, Folch EE, Khandhar SJ, Perry Y, et al. Safety of electromagnetic navigation bronchoscopy in patients With COPD: results From the NAVIGATE study. J Bronchol Interv Pulmonol. 2019;26:33–40.

    Article  Google Scholar 

  35. Webb TN, Bonta DV, Masters R, Parks C, Bechara R. Jet ventilation decreases target motion and increases yield of ENB especially in lesions with negative bronchus sign. J Bronchol Interv Pulmonol. 2020;27:14–21.

    Article  Google Scholar 

  36. Zhang W, Chen S, Dong X, Lei P. Meta-analysis of the diagnostic yield and safety of electromagnetic navigation bronchoscopy for lung nodules. J Thorac Dis. 2015;7:799–809.

    PubMed  PubMed Central  Google Scholar 

  37. Edmark L, Kostova-Aherdan K, Enlund M, Hedenstierna G. Optimal oxygen concentration during induction of general anesthesia. Anesthesiology. 2003;98:28–33.

    Article  CAS  PubMed  Google Scholar 

  38. Coussa M, Proietti S, Schnyder P, Frascarolo P, Suter M, Spahn DR, et al. Prevention of atelectasis formation during the induction of general anesthesia in morbidly obese patients. Anesth Analg. 2004;98:1491–5.

    Article  PubMed  Google Scholar 

  39. Tusman G, Bohm SH, Warner DO, Sprung J. Atelectasis and perioperative pulmonary complications in high-risk patients. Curr Opin Anaesthesiol. 2012;25:1–10.

    Article  CAS  PubMed  Google Scholar 

  40. Hemmes SN, Serpa Neto A, Schultz MJ. Intraoperative ventilatory strategies to prevent postoperative pulmonary complications: a meta-analysis. Curr Opin Anaesthesiol. 2013;26:126–33.

    Article  PubMed  Google Scholar 

  41. Rothen HU, Sporre B, Engberg G, Wegenius G, Reber A, Hedenstierna G. Prevention of atelectasis during general anaesthesia. Lancet. 1995;345:1387–91.

    Article  CAS  PubMed  Google Scholar 

  42. Rusca M, Proietti S, Schnyder P, Frascarolo P, Hedenstierna G, Spahn DR, et al. Prevention of atelectasis formation during induction of general anesthesia. Anesth Analg. 2003;97:1835–9.

    Article  PubMed  Google Scholar 

  43. Eichenberger A, Proietti S, Wicky S, Frascarolo P, Suter M, Spahn DR, et al. Morbid obesity and postoperative pulmonary atelectasis: an underestimated problem. Anesth Analg. 2002;95:1788–92.

    Article  PubMed  Google Scholar 

  44. Magnusson L, Spahn DR. New concepts of atelectasis during general anaesthesia. Br J Anaesth. 2003;91:61–72.

    Article  CAS  PubMed  Google Scholar 

  45. Eltorai AEM, Baird GL, Eltorai AS, Pangborn J, Antoci V Jr, Cullen HA, et al. Perspectives on incentive spirometry utility and patient protocols. Respir Care. 2018;63:519–31.

    Article  PubMed  Google Scholar 

  46. Nimmagadda U, Salem MR, Crystal GJ. Preoxygenation: physiologic basis, benefits, and potential risks. Anesth Analg. 2017;124:507–17.

    Article  CAS  PubMed  Google Scholar 

  47. Galway U, Zura A, Khanna S, Wang M, Turan A, Ruetzler K. Anesthetic considerations for bronchoscopic procedures: a narrative review based on the Cleveland Clinic experience. J Thorac Dis. 2019;11:3156–70.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Brimacombe J. The advantages of the LMA over the tracheal tube or facemask: a meta-analysis. Can J Anaesth. 1995;42:1017–23.

    Article  CAS  PubMed  Google Scholar 

  49. Hohenforst-Schmidt W, Zarogoulidis P, Huang H, Man Y-G, Laskou S, Koulouris C, et al. A new and safe mode of ventilation for interventional pulmonary medicine: the ease of nasal superimposed high frequency jet ventilation. J Cancer. 2018;9:816–33.

    Article  PubMed  PubMed Central  Google Scholar 

  50. Ventilatory strategy for the prevention of atelectasis during bronchoscopy under general anesthesia, VESPA trial (NCT04311723). ClinicalTrials.gov website. 2020. https://clinicaltrials.gov/ct2/show/NCT04311723. Accessed 31 May 2021.

  51. Battisti WP, Wager E, Baltzer L, Bridges D, Cairns A, Carswell CI, et al. Good publication practice for communicating company-sponsored medical research: GPP3. Ann Intern Med. 2015;163:461–4.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors wish to thank Jennifer S. Mattingley MD and Mike Calcutt of Medtronic for review and guidance.

Funding

Supported by Medtronic. Medical writing support was provided by Kristin L. Hood PhD of Medtronic (Minneapolis, MA) in accordance with Good Publication Practice (GPP3) guidelines [51]. The authors had final responsibility for the decision to submit for publication and were not paid to write this article by Medtronic or any other agency.

Author information

Authors and Affiliations

Authors

Contributions

MP and KB conceived of the paper and developed the ventilation methodology; MP, KB, KL, and SS validated the methodology; KP refined the methodology; MP, KB, and KP wrote sections of the paper and directed the content; all authors provided interpretation and critically revised all drafts of the paper. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Michael A. Pritchett.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

K. Bhadra discloses consulting/speaking payments from Medtronic, Boston Scientific, BodyVision, Auris Health, Inc., Intuitive Surgical, Veracyte, Biodesix, Merit Medical Endotek, and Johnson & Johnson. K. Lau discloses consulting/speaking payments from Medtronic, Philips, Intuitive Surgical, and Johnson & Johnson. K. Phillips is a full-time employee of Medtronic. M. Pritchett discloses speaking, consulting, or research payments from Medtronic, Auris Health, Inc, BodyVision, Intuitive Surgical, Philips, Biodesix, AstraZeneca, Johnson & Johnson/NeuWave, Actelion, Inivata, and Boston Scientific. S. Skibo discloses speaking/consulting payments from Medtronic, Biodesix, GlaxoSmithKline and AstraZeneca.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Literature search methods.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pritchett, M.A., Lau, K., Skibo, S. et al. Anesthesia considerations to reduce motion and atelectasis during advanced guided bronchoscopy. BMC Pulm Med 21, 240 (2021). https://doi.org/10.1186/s12890-021-01584-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-021-01584-6

Keywords