Optimization of sniff nasal inspiratory pressure (SNIP) measurement methodology in healthy subjects

Abstract

Background

Maximal inspiratory pressure (MIP) is currently the most commonly used measure for respiratory muscle strength (RMS) estimation, however, requires significant effort. Falsely low values are therefore common, especially in fatigue-prone subjects, such as neuromuscular disorder patients. In contrast, sniff nasal inspiratory pressure (SNIP) requires a short, sharp sniff; this is a natural manoeuvre, decreasing required effort. Consequently, it has been suggested that use of SNIP could confirm the accuracy of MIP measurements. However, no recent guidelines regarding the optimal method of SNIP measurement exist, and varied approaches have been described.

Objectives

We compared SNIP values from three conditions, namely with 30, 60 or 90 s time intervals between repeats, the right (SNIPR) and left (SNIPL) nostril, and the contralateral nostril occluded (SNIPO) or non-occluded (SNIPNO). Additionally, we determined the optimal number of repeats for accurate SNIP measurement.

Method

52 healthy subjects (23 males) were recruited for this study, of which a subset of 10 subjects (5 males) completed tests comparing the time interval between repeats. SNIP was measured from functional residual capacity via a probe in one nostril, while MIP was measured from residual volume.

Results

There was no significant difference in SNIP depending on the interval between repeats (P = 0.98); subjects preferred the 30 s. SNIPO was significantly higher than SNIPNO (P < 0.00001) but SNIPL and SNIPR did not significantly differ (P =  0.60). There was an initial learning effect for the first SNIP test; SNIP did not decline during 80 repeats (P =  0.64).

Conclusions

We conclude that SNIPO is a more reliable RMS indicator than SNIPNO, as there is reduced risk of RMS underestimation. Allowing subjects to choose which nostril to use is appropriate, as this did not significantly affect SNIP, but may increase ease of performance. We suggest that twenty repeats is sufficient to overcome any learning effect and that fatigue is unlikely after this number of repeats. We believe these results are important in aiding the accurate collection of SNIP reference value data in the healthy population.

Peer Review reports

Introduction

Estimation of respiratory muscle strength (RMS) by measuring pressures within the thoracic cavity aids diagnosis and monitoring of several conditions, such as neuromuscular disorders (NMDs; [1]). Such techniques are beneficial, as pressure-based values typically decline before volume-based measures such as vital capacity [1], allowing more timely diagnosis and treatment initiation, which can improve prognosis and quality of life. Maximal inspiratory pressure (MIP) is currently the most commonly used RMS measure [2], however it demands significant effort, as it requires a maximum inspiration from residual volume [3]. This is disadvantageous to NMD patients, as they are prone to fatigue, meaning RMS underestimation is common when measuring MIP in these patients [4]. MIP performance also requires a tight seal to be formed between the mouthpiece and the lips; facial muscle weakness is common in NMDs, and may impair adequate seal formation, again conferring risk of RMS underestimation [4]. This can result in overdiagnosis and inappropriate treatment, impairing the quality of life of patients.

In contrast to MIP, sniff nasal inspiratory pressure (SNIP) measurement does not require formation of such a seal, and requires relatively little effort [5]. SNIP is a non-invasive measure of RMS, measured via a probe inserted into one nostril during a short, sharp sniff [6]. The sniff is dependent on inspiratory muscle contraction, predominately the diaphragm, therefore the pressures recorded indicate RMS [7]. SNIP measurements have been correlated with quality of life and mortality risk in conditions including amyotrophic lateral sclerosis and are an indication for therapy initiation [1, 8]. Several groups have compared SNIP and MIP; one has shown SNIP to be higher [9] while others have shown no difference [10, 11]. It has been suggested that SNIP is unlikely to replace MIP, however several groups have concluded that use of both tests in a complementary manner could be beneficial [9, 12, 13]. For example, recording a low MIP yet a normal SNIP could indicate issues such as suboptimal effort, fatigue or lack of a complete seal during MIP performance, helping to rule out inspiratory muscle weakness and minimizing false positives. This would mean clinicians gain a more accurate idea of patients’ RMS, increasing diagnostic and prognostic accuracy and ensuring the most appropriate therapy is administered [14].

Accurate SNIP measurement is vital due to the importance of the values in diagnosing and monitoring disease, both alone and alongside measures such as MIP. However, varied measurement methodologies have been described. Initial SNIP methodology stated the contralateral nostril should be non-occluded [6], however, several groups have subsequently detailed methodology with occlusion of this nostril [2, 10, 15,16,17]. Additionally, no consensus exists regarding which nostril should be used; often the nostril appearing most patent is chosen [10, 17] or where a few SNIP tests show a higher value [18], however the impact of nostril choice on SNIP values has received limited attention. The optimal number of repeats has also been a subject of debate. Lofaso et al. reported 10 repeats is sufficient to complete the learning effect [19], however other groups have detailed methodology using both more [4, 20, 21] and less repeats [8, 16, 22]. In addition to repeat number, the optimal rest time between repeats is poorly defined. The impact of this variable is important to consider, as too short an interval could increase risk of fatigue, whereas too long an interval could pose risk of the subject becoming distracted and consequently applying submaximal effort. The present study will determine the impact of the above variables on SNIP values, thus clarifying these discrepancies and increasing SNIP reliability. Better standardizations would enable the collection of accurate reference values in healthy subjects, which will allow better assessment of patient populations.

Our aims included to test the impact of the time interval between SNIP repeats, compare right and left nostril data and determine whether occluding the contralateral nostril influenced results. Additionally, we investigated if there was a learning effect and whether subjects were likely to experience fatigue.

Methods

The study was approved by the Faculty of Science and Engineering Research Ethics Committee, University of Plymouth. The subjects were recruited randomly from staff and students of the University of Plymouth. All subjects gave informed and written consent before participation. The experimental protocol was explained to all subjects.

Participants

52 healthy subjects (23 males) participated in the study. Exclusion criteria included being under 18 or over 65 years of age, smokers, patients with cardiorespiratory or neuromuscular disease or any previous major cardiothoracic surgery. We verified that subjects met these criteria via a questionnaire. One subject’s results were excluded from analyses due to consistent SNIP readings of zero, indicating inadequate SNIP performance.

Measurements

Prior to SNIP performance, anthropometric and other measurements were recorded, including height (Seca, Germany), weight (Marsden, UK), blood pressure, heart rate and SpO2; the last three measurements were recorded using a Vital Signs Monitor 300 (Welch Allyn, USA). MIP and maximal expiratory pressure (MEP) were recorded using a MicroRPM device (Care Fusion, UK) and variability was less than 20% [3]. For MIP, subjects exhaled to residual volume before inspiring through a mouthpiece until recording a maximum value, and for MEP, subjects inhaled to total lung capacity before expiring through the device [3, 18]. Pulmonary function tests (PFTs) were performed using a Micro Loop spirometer (Care Fusion, UK), and repeated until meeting British Thoracic Society criteria [23].

SNIP was measured using the MicroRPM device whilst subjects remained seated upright with both feet on the floor. Instructions were given to breathe normally between tests, and, on cue, exhale to functional residual capacity before sharply sniffing inwards with the mouth closed [18]. With a subset of ten subjects (five males), SNIP was measured via a probe inserted into the right nostril, with the left non-occluded. Subjects performed three sets of ten repeats; during each, either a 30, 60 or 90 s rest was given between repeats (Fig. 1). The order of tests was randomized for each subject. Data from these experiments determined that 30 s was an appropriate rest interval for all remaining experiments. After an interval of at least one week, all 51 subjects performed four sets of 20 SNIP tests (Fig. 1). Each set used a different technique to measure SNIP; via the right nostril with the left non-occluded (RNLNO), the right nostril with the left occluded (RNLO), the left nostril with the right non-occluded (LNRNO) or the left nostril with the right occluded (LNRO). The order of tests was randomized for each subject by giving them four numbered cards, where each number corresponded to one of the four techniques, and asking the subjects to pick these in a sequence (whilst the cards were face down). The contralateral nostril was occluded by subjects placing their thumb over their nostril for the RNLO and LNRO techniques.

Fig. 1
figure 1

Subject recruitment and study design. RNLO: right nostril, left occluded; RNLNO: right nostril, left non-occluded; LNRO: left nostril, right occluded; LNRNO: left nostril, right non-occluded

Statistical analysis

For analysis purposes, the mean maximal SNIP (mmSNIP) was calculated. This was calculated by selecting each subject’s maximal SNIP for each technique and then dividing the sum of these by the number of subjects. For example, the maximal SNIP values for the subset investigating the 30 s interval, would be summed and then divided by ten (number of subset subjects). Similarly, when investigating nostril technique, mmSNIP was calculated by summing the maximum SNIP values and dividing by the total subject number (n = 51). Repeated-measures ANOVA tested differences between the mmSNIP obtained from the four techniques and also tested differences across all 20 repeats within each set. Unpaired t-testing compared differences in SNIP depending on gender, ethnicity, nostril used and occlusion. Multiple linear regression was used the individual maximal SNIP values (SNIPmax) as the dependent variable and with gender, age, BMI, physical activity, mean arterial pressure, heart rate, SpO2 and lung function as independent variables in the model. All statistical analyses were performed using SPSS Version 24, with a probability level P < 0.05 considered statistically significant for t-tests and multiple linear regression. When comparing the four techniques, a stricter probability level of P < 0.0125 was used for the repeated-measures ANOVA, with Bonferroni’s correction being used to adjust for type 1 errors. However, with the 10 or 20 repeat experiments, a probability level P < 0.05 was used [24]. Bland–Altman analysis looks for agreement between two methods of measurement. The maximum SNIP data minus maximum MIP was run through a one group t-test, with a population mean of 0. If the values were significantly different from the mean of 0, then it was assumed that the two values differed from each other. Subject data can be viewed in this Additional file 1: Supplementary data file 1.

Results

Participants

Mean questionnaire and lung function data of the 51 study participants are summarised in Table 1. No subject experienced any adverse events whilst participating in the study. Each mean SNIP value in this table was calculated using 2,040 individual SNIP measurements, 51 subjects had 20 tests taken twice. The mean value of the maximal SNIP for each of these 20 tests (102 values) was used.

Table 1 Mean (± SD) data of the 51 participants

Effect of time interval between SNIP tests (n = 10)

The mmSNIP did not significantly differ depending on whether repeats were performed 30, 60 or 90 s apart (84.0 ± 30.9, 84.6 ± 32.6 and 86.9 ± 34.3 cmH2O; P =  0.98, n = 10). In terms of the actual order these tests were performed, there was no significant difference in the mmSNIP obtained from tests in the first, second or third (85.6 ± 35.1, 85.3 ± 30.6 and 84.6 ± 32.1 cmH2O) position of order (P =  0.95), irrespective of which repeat time interval was used.

Impact of nostril and occlusion choice on SNIP

Overall, mmSNIP significantly differed across the four techniques tested (P < 0.00001, n = 51; Fig. 2). Significant differences between tests with the contralateral nostril occluded (SNIPO; RNLO and LNRO) and non-occluded (SNIPNO; RNLNO and LNRNO) were found using t-testing (73.4 ± 33.7 vs. 56.1 ± 27.6 cmH2O, P < 0.00001, n = 102). However, when tests were performed via the left (SNIPL; LNRO and LNRNO) or right (SNIPR; RNLO and RNLNO) nostril, no difference was noted (63.6 ± 31.1 vs. 66.0 ± 32.9 cmH2O, P =  0.59, n = 102). T-testing also revealed significant differences between the RNLO and RNLNO techniques (75.1 ± 34.6 vs. 56.8 ± 28.6 cmH2O; P =  0.005, n = 51), and between the LNRO and LNRNO techniques (71.8 ± 33.1 vs. 55.4 ± 26.8 cmH2O; P =  0.007, n = 51). However, no significant difference was found between the RNLO and LNRO techniques (75.1 ± 34.6 vs. 71.8 ± 33.1 cmH2O; P =  0.54, n = 51), or between the RNLNO and LNRNO techniques (56.8 ± 28.6 vs. 55.4 ± 26.8 cmH2O; P =  0.79, n = 51).

Fig. 2
figure 2

Mean maximal SNIP (± SD) obtained using different nostril techniques (n = 51). RNLO: right nostril, left occluded; RNLNO: right nostril, left non-occluded; LNRO: left nostril, right occluded; LNRNO: left nostril, right non-occluded. * denotes significant differences (P < 0.01) between RNLO vs. RNLNO and LNRO vs. LNRNO

Presence of a learning effect or fatigue

Repeated-measures ANOVA compared mmSNIP data for tests in the actual order they were performed, regardless of technique used. No significant difference was found between the mmSNIP obtained in the first, second, third or fourth set of 20 repeats (63.4 ± 26.6, 62.9 ± 30.0, 67.1 ± 37.4 and 65.6 ± 33.5 cmH2O; P =  0.64; Fig. 3).

Fig. 3
figure 3

Mean maximal SNIP (± SD) obtained from each set of 20 SNIP tests (n = 51); mmSNIP did not significantly differ depending on the set position (P =  0.64)

However, there was an overall significant difference in the mean SNIP between repeats within the first set of 20 SNIPs (P =  0.00006), indicating the presence of a learning effect within the first 20 repeats. Using within-subject pairwise comparisons showed a significant difference between repeat number 1 and repeat numbers 7, 9 and 12–20 (P < 0.05). This was not the case for repeats performed in the second (P =  0.21), third (P =  0.36) or fourth (P =  0.70) sets (Fig. 4).

Fig. 4
figure 4

Mean (± SD) SNIP data obtained by subjects within the a first, b second, c third and d fourth sets for tests in the actual order they were performed, irrespective of technique used (n = 51). * denotes significant difference in mean SNIP between repeat number 1 and 20 (P =  0.006 within (a), the first set)

Relationship between SNIP and gender, ethnicity, MIP and MEP

T-tests revealed that mmSNIP did not significantly differ between males and females (87.7 ± 32.9 vs. 76.1 ± 34.4 cmH2O, P =  0.23). There was also no statistically significant difference in mmSNIP between Caucasians and non-Caucasians (80.5 ± 4.9 vs. 89.5 ± 20.8 cmH2O, P =  0.62), however, there were only four non-Caucasians in our study. Using multiple linear regression with SNIPmax as the dependant variable, our model found MIP to be correlated with SNIP (P =  0.024; Fig. 5). No other non-SNIP variables were significantly correlated with SNIP.

Fig. 5
figure 5

Relationship between a SNIPmax and MIP, showing a significant correlation b SNIPmax and MEP, showing no significant correlation, in our cohort of healthy subjects (n = 51)

Bland–Altman analysis for SNIPmax and MIP

Although linear regression found a relationship between the maximum values of SNIP and MIP, a Bland–Altman analysis was required. From a Bland–Altman plot, we did find three data points being beyond the limits of agreement (Fig. 6). A one sample t-test showed SNIPmax was significantly different, and greater, than MIP (82.0 ± 33.7 vs. 72.8 ± 33.9cmH2O, P =  0.034).

Fig. 6
figure 6

Bland–Altman showing the difference in SNIPmax and MIP, plotted against the mean of these two values (n = 51). The upper dashed line indicates the upper limit of agreement, the continuous line the mean difference and the lower dashed line indicates the lower limit of agreement

Discussion

This study has clearly shown that 30 s intervals between repeat tests are appropriate in SNIP testing in healthy subjects. We have also demonstrated that techniques occluding the contralateral nostril significantly raised SNIP values, but using either the right or left nostril did not significantly differ. There was also a learning effect, where the first test in the first set of 20 SNIP tests (out of four sets), was seen to be lower. Although recent data has been published on the effects of contralateral occlusion on SNIP values, there is a scarcity of data on intervals between tests and differences between right and left nostrils.

Effect of interval between tests on SNIP values

Results obtained from a subset of subjects led us to conclude that the rest time between intervals does not significantly impact SNIP. The shortest rest time (30 s) was tolerable to participants; no subject complained this was too short and no adverse events occurred. Additionally, several subjects thought 90 s intervals were too long and increased the risk of distraction; this could lead to a submaximal effort. This is contrary to our hypothesis that a longer rest time would be preferable to subjects. As far as the authors are aware, we are the first group to investigate whether different time intervals affect SNIP values. We did not investigate how intervals shorter than 30 s affected SNIP or subject comfort, therefore cannot conclude whether use of shorter rest times would be appropriate. However, in our study, we turned our MicroRPM device off after each test, so we could record each SNIP value. If not turned off, the device would only record a value if it were greater than a previous reading, so a repeat lower reading would not be displayed. Although this is clinically useful, we wanted to record each individual value, thus turned the device off, meaning 5 to 10 s was needed for it to restart each time. Taking this into account and the fact the investigator needed to record the measurement, it would have been difficult to complete this process within an interval less than 30 s. As the present study included only healthy subjects, we cannot conclude whether 30 s is an appropriate rest interval for patients who may be prone to fatigue, such as those with NMDs. However, our findings are supported by the fact that 30 s intervals have previously been used in NMD and lung disease patients, with no adverse events or significant impact on SNIP values reported [11, 19] and in healthy children [25].

Comparison of SNIPO and SNIPNO

Our data showed a significant difference between SNIPO and SNIPNO; mean maximum SNIPO was significantly higher than mean maximum SNIPNO, in agreement with previous reports [10, 16, 17, 26]. However, our mmSNIP values were slightly lower than previously reported [10, 16, 26] and one possibility for this could be differences in the age of these cohorts [27]. The percentage difference between occluded and non-occluded SNIP values were similar; ours was 31% and previous groups showed it to be 29% [10] and 26% [16]. However for one study it was 8% [26].

The difference between SNIPO and SNIPNO is most likely because occlusion ensures a completely sealed nasal cavity rather than relying on collapse of the contralateral nasal valve, as the open manoeuvre must [28], increasing the recorded pressure. In both the present study and previous studies [10, 16, 17], SNIPNO was lower than MIP. SNIPNO may therefore be more likely to underestimate RMS, therefore may not be a suitable for use after a low MIP has been recorded. In contrast, we found SNIPO to be slightly higher than MIP, similar to recent reports [16]. Kaminska et al. have shown SNIPO to be slightly lower than MIP [10], leading us to conclude that both variables maybe very similar in magnitude. Therefore, SNIPO may help to determine whether a low MIP is due to inspiratory muscle weakness or issues such as difficulty performing MIP. SNIPO may therefore be valuable as part of a multimodal evaluation during diagnosis and monitoring of conditions including NMDs [17].

Despite reduced risk of RMS underestimation, one limitation when using SNIPO is the method of occlusion. During our study, subjects covered the nostril entrance with their thumb. We had decided not to use a plug, as we were concerned this might interfere with the SNIP probe. Tape was also decided against as it allowed leaks and would have been uncomfortable for subjects. In our cohort of healthy subjects, we felt using their thumbs was the most convenient option, however we do understand that this might not be feasible in a patient population with NMD. However, using the thumb method may have decreased reproducibility due to subjects potentially moving their thumb between repeats, reducing certainty that a complete seal was always formed. Although the method of occlusion is not always reported, we noted that Tilanus et al. used a silicone plug for contralateral nostril occlusion [8]. For SNIPO to be performed reliably, a standardized method of occlusion is needed which is compatible with the SNIP probe in use.

Comparison of SNIPL and SNIPR

SNIPL and SNIPR did not significantly differ, suggesting that use of either nostril is appropriate. Other groups have used the nostril appearing most patent [4, 10, 17], however we suggest that giving subjects the choice of nostril for SNIP is a suitable approach. This may be beneficial as some subjects commented that one nostril felt more uncomfortable than the other. Feedback from our subjects suggests that if subjects perform SNIP in the nostril which feels most comfortable, they are more likely to apply maximal effort, increasing SNIP accuracy and hence reducing risk of RMS underestimation. Having a choice would have been particularly useful for two of our subjects who had reported a previous broken nose. Both found difficulty inserting the probe into one nostril, potentially due to previous damage to their nasal structure [29]. In these two cases, the subjects became frustrated and needed more time for nasal probe positioning. To use only the other nostril would have prevented this difficulty, increasing ease of performance and likelihood of maximal effort being applied. Dominant nostril airflow is unlikely to have affected our data, as nostril airflow has been thought to alternate over a short timespan [30].

Impact of a learning effect or fatigue on SNIP

Within our study, there was a learning effect within the first set of twenty SNIPs. We suggest that 20 repeats are necessary for reliable SNIP testing, in slight contrast with Lofaso et al. who recommend that 20 repeats are necessary only when SNIPmax in the first 10 is slightly below normal [19].

SNIP did not decline during 80 repeats, suggesting that healthy subjects are unlikely to experience fatigue during repeat SNIP testing. This confirms and expands upon Uldry and Fitting’s findings that SNIP did not decline during 30 repeats [9]. However, both Uldry and Fitting’s study [9] and the present study included only healthy subjects; we therefore cannot conclude whether fatigue would occur during large numbers of repeats in fatigue-prone subjects, such as NMD patients. Further studies evaluating the impact of fatigue on SNIP in such patients would be beneficial, to ensure that a suitable number of repeats is performed in clinical settings and reliable results are obtained.

Variation of SNIP depending on gender, ethnicity and age

In our study, SNIP did not differ depending on gender, in agreement with previous findings [10, 20]. However, some groups [27, 31] have reported that SNIP is greater in males. This disagreement is likely due to methodological variation and potentially differing subject demographics. Additionally, we found that SNIP did not correlate with age; supporting previous findings [10, 20]. However, two studies reported a negative correlation with age [9, 27]. These latter two studies [9, 27] did use subjects up to the age of 80, while the former two [10, 20] and our study tended to have subjects or patients younger than 65 years old. This could explain this discrepancy with regard to age.

Comparison of SNIP and MIP

It is widely reported that SNIP is an easier, more natural test than MIP, which is particularly beneficial when assessing certain subject groups, such as those with NMDs [4]. Previous studies have found such subjects are almost always able to perform SNIP adequately, and always more frequently than MIP [15]. In contrast, all subjects in the present study successfully performed MIP, whereas one subject was unable to successfully perform SNIP. However, the subjects participating in our study were healthy, therefore would not be expected to experience significant difficulty with MIP performance. It is unclear why one subject in our study experienced difficulty with SNIP performance; further studies may help to determine why SNIP may be inappropriate in some subjects or patients and if this could be predicted. Although our study showed SNIP and MIP values were significantly correlated, the Bland–Altman analysis showed they differed, where SNIP was higher. This is supported by previous work, which also found SNIP to be significantly higher than MIP [9]. However, in contrast, work in NMD patients and healthy subjects has shown no significant difference between SNIP and MIP in both groups [10]. Recent work in patients has again shown no significant differences between these two variables [11].

Limitations

One limitation of our study was that after several SNIP repeats, excess mucus was produced in some subjects, increasing the chance of the probe falling out. This necessitated them to push it back in, often inadvertently changing its position. Different positions might produce different results; we noted if the probe hit the back of the nostril, SNIP was zero as no pressure could be recorded. In contrast, if the probe was not pushed in far enough, it might lower SNIP due to an incomplete seal. Therefore, in order to ensure SNIP remains reliable no matter how many repeats are performed, a technique to ensure the probe remains in a constant position should be devised. Another potential solution is to move the probe to the other nostril if excess mucus is produced.

Another limitation is that, although all subjects performed SNIP whilst in an upright position, their exact position was not strictly controlled. Recent work has shown that a slouched sitting position can significantly decrease SNIP compared to an upright sitting position [22], potentially due to reduced diaphragm tension and movement. Thus, variations in subject position may have introduced variability into our results and affected the reliability of our study.

Only healthy subjects were included in the present study. Results would differ significantly in patients with diseases such as NMDs, for example due to muscular weakness and increased likelihood of fatigue, among other reasons [12]. Although our conclusions cannot be extended to patient populations, they do provide a baseline to guide the collection of normal values in healthy subjects, and this in turn will aid diagnosis and prognosis in patients with reduced SNIP values.

Inter-investigator reliability was not an issue, as our study only had one investigator involved in data collection.

Conclusions

In summary, we showed that SNIP testing via either nostril, with the contralateral nostril occluded, is an appropriate and reliable method of assessing respiratory muscle strength and that SNIP values are higher to MIP values. We also conclude that twenty repeats are sufficient to overcome any learning effect and that fatigue is not an issue in healthy subjects even during up to 80 repeats. Our subset study confirmed that SNIP is associated with minimal fatigue as the shortest rest interval between SNIP repeats (30 s) did not affect results, suggesting SNIP performance does not require significant effort in healthy subjects. A larger cohort of patients would be required to ideally optimize measurements of SNIP measurement methodology. We believe our findings are important for methodological studies determining reference values in a healthy population (Additional file 1).

Availability of data and materials

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  1. Sferrazza Papa GF, Pellegrino GM, Shaikh H, Lax A, Lorini L, Corbo M. Respiratory muscle testing in amyotrophic lateral sclerosis: a practical approach. Minerva Med. 2018;109:11–9.

    PubMed  Google Scholar 

  2. Tilanus TBM, Groothuis JT, Ten Broek-Pastoor JMC, Doorduin J, van Engelen BGM, Kampelmacher MJ, Raaphorst J. Respiratory assessment of ALS patients: a nationwide survey of current Dutch practice. J Neuromusc Dis. 2018;5(4):431–8.

    Article  CAS  Google Scholar 

  3. American Thoracic Society and European Respiratory Society. ATS/ERS statement on respiratory muscle testing. Am J Respir Crit Care Med. 2002;166:518–624.

    Article  Google Scholar 

  4. Oliveira MJ, Rodrigues F, Firmino-Machado J, Ladeira IT, Lima R, Conde S, Guimarães M. Assessment of respiratory muscle weakness in subjects with neuromuscular disease. Respiratory Care. 2018;63(10):1223–30.

    Article  PubMed  Google Scholar 

  5. Fitting JW. Sniff nasal inspiratory pressure: simple or too simple? Eur Respir J. 2006;27:881–3.

    Article  PubMed  Google Scholar 

  6. Héritier F, Rahm F, Pasche P, Fitting JW. Sniff nasal inspiratory pressure: a noninvasive assessment of inspiratory muscle strength. Am J Respir Crit Care Med. 1994;150:1678–83.

    Article  PubMed  Google Scholar 

  7. Stefanutti D, Fitting JW. Sniff nasal inspiratory pressure reference values in Caucasian children. Am J Respir Crit Care Med. 1999;159:107–11.

    Article  CAS  PubMed  Google Scholar 

  8. Tilanus TBM, Groothuis JT, Ten Broek-Pastoor JMC, Feuth TB, Heijdra YF, Slenders JPL, Doorduin J, van Engelen BGM, Kampelmacher MJ, Raaphorst J. The predictive value of respiratory function tests for non-invasive ventilation in amyotrophic lateral sclerosis. Respir Res. 2017;18(1):144.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Uldry C, Fitting JW. Maximal values of sniff nasal inspiratory pressure in healthy subjects. Thorax. 1995;50:371–5.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Kaminska M, Noel F, Petrof BJ. Optimal method for assessment of respiratory muscle strength in neuromuscular disorders using sniff nasal inspiratory pressure (SNIP). PLoS ONE. 2017;12(5):e0177723.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Janssens J-P, Adler D, Ferfoglia RI, Poncet A, Graf LG, Leuchter I, Imhof ME, Héritier Barras A-C. Assessing inspiratory muscle strength for early detection of respiratory failure in motor neuron disease: should we use MIP, SNIP, or both? Respiration. 2019;98(2):114–24.

    Article  PubMed  Google Scholar 

  12. Hart N, Polkey MI, Sharshar T, Falaize L, Fauroux B, Raphaël JC, Lofaso F. Limitations of sniff nasal pressure in patients with severe neuromuscular weakness. J Neurol Neurosurg Psychiatry. 2003;74(12):1685–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Martínez-Llorens J, Ausín P, Roig A, Balañá A, Admetlló M, Muñoz L, Gea J. Nasal inspiratory pressure: an alternative for the assessment of inspiratory muscle strength? Arch Bronconeumol. 2011;47(4):169–75.

    Article  PubMed  Google Scholar 

  14. Steier J, Kaul S, Seymour J, Jolley C, Rafferty G, Man W, Luo YM, Roughton M, Polkey MI, Moxham J. The value of multiple tests of respiratory muscle strength. Thorax. 2007;62(11):975–80.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Morgan RK, McNally S, Alexander M, Conroy R, Hardiman O, Costello RW. Use of sniff nasal-inspiratory force to predict survival in amyotrophic lateral sclerosis. Am J Respir Crit Care Med. 2005;171(3):269–74.

    Article  PubMed  Google Scholar 

  16. Pinto S, de Carvalho M. Sniff nasal inspiratory pressure (SNIP) in amyotrophic lateral sclerosis: relevance of the methodology for respiratory function evaluation. Clin Neurol Neurosurg. 2018;171:42–5.

    Article  PubMed  Google Scholar 

  17. Murray D, Rooney J, Campion A, Fenton L, Hammond M, Heverin M, Hardiman O. Longitudinal analysis of sniff nasal inspiratory pressure assessed using occluded and un-occluded measurement techniques in amyotrophic lateral sclerosis and primary lateral sclerosis. Amyotrophic Lateral Sclerosis Frontotemp Degen. 2019;20(7–8):481–9.

    Article  CAS  Google Scholar 

  18. Laveneziana D, Albuquerque A, Aliverti A, Babb T, Barreiro E, Dres M, Dubé B, Fauroux B, Gea J, Guenette JA, Hudson AL, Kabitz H, Laghi F, Langer D, Luo Y, Neder JA, O’Donnell D, Polkey MI, Rabinovich RA, Rossi A, Series F, Similowski T, Spengler CM, Vogiatzis I, Verges S. ERS statement on respiratory muscle testing at rest and during exercise. Eur Respirat J. 2019;53(6):1801214.

    Article  Google Scholar 

  19. Lofaso F, Nicot F, Lejaille M, Falaize L, Louis A, Clement A, Raphael JC, Orlikowski D, Fauroux B. Sniff nasal inspiratory pressure: what is the optimal number of sniffs? Eur Respirat J. 2006;27(5):980–2.

    Article  CAS  Google Scholar 

  20. Prigent H, Orlikowski D, Fermanian C, Lejaille M, Falaize L, Louis A, Fauroux B, Lofaso F. Sniff and Muller manoeuvres to measure diaphragmatic muscle strength. Respir Med. 2008;102(12):1737–43.

    Article  PubMed  Google Scholar 

  21. Nève V, Cuisset JM, Edmé JL, Carpentier A, Howsam M, Leclerc O, Matran R. Sniff nasal inspiratory pressure in the longitudinal assessment of young Duchenne muscular dystrophy children. Eur Respir J. 2013;42(3):671–80.

    Article  PubMed  Google Scholar 

  22. Albarrati A, Zafar H, Alghadir AH, Anwer S. Effect of upright and slouched sitting postures on the respiratory muscle strength in healthy young males. Biomed Res Int. 2018. https://doi.org/10.1155/2018/3058970.

    Article  PubMed  PubMed Central  Google Scholar 

  23. British Thoracic Society, A guide to performing quality assured diagnostic spirometry (2013). https://www.brit-thoracic.org.uk/media/70454/spirometry_e-guide_2013.pdf. Accessed: 19 Oct 2022

  24. Park E, Cho M, Ki C. Correct use of repeated measures analysis of variance. Korean J Lab Med. 2009;29:1–9.

    PubMed  Google Scholar 

  25. Marcelino AA, Fregonezi G, Marques L, Lista-Paz A, Torres-Castro R, Resqueti V. Reliability of SNIP test and optimal number of maneuvers in 6–11 years healthy children. PLoS ONE. 2021;16(5):e0252150. https://doi.org/10.1371/journal.pone.0252150.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Corberó AB, Admetllo M, Sancho-Muñoz A, Alvarado M, Gea J, Ausin P, Llorens JM. Descriptive study of the effect of methodology in the measurement of sniff nasal inspiratory pressure (SNIP) in a healthy population. Arch Bronconeumol. 2020;56(6):398–9. https://doi.org/10.1016/j.arbr.2019.12.005.

    Article  Google Scholar 

  27. De Araújo PR, Resqueti VR, Nascimento J Jr, Carvalho LA, Cavalcanti AG, Silva VC, Silva E, Moreno MA, Andrade AFD, Fregonezi GAF. Reference values for sniff nasal inspiratory pressure in healthy subjects in Brazil: a multicenter study. J Bras Pneumol. 2012;38(6):700–7.

    Article  PubMed  Google Scholar 

  28. Haight JS, Cole P. The site and function of the nasal valve. Laryngoscope. 1983;93(1):49–55.

    Article  CAS  PubMed  Google Scholar 

  29. Mondin V, Rinaldo A, Ferlito A. Management of nasal bone fractures. Am J Otolaryngol. 2005;26(3):181–5.

    Article  PubMed  Google Scholar 

  30. Price A, Eccles R. Nasal airflow and brain activity: is there a link? J Laryngol Otol. 2016;130:794–9.

    Article  CAS  PubMed  Google Scholar 

  31. Kamide N, Ogino M, Yamashina N, Fukuda M. Sniff nasal inspiratory pressure in healthy Japanese subjects: mean values and lower limits of normal. Respiration. 2009;77(1):58–62.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The authors received no grant funding for this work. The study was internally funded by the School of Biomedical Sciences. RW received an internal summer bursary.

Author information

Authors and Affiliations

Authors

Contributions

Conception or design of the work (RW, MT, MMFS). Acquisition, analysis, or interpretation of data for the work (RW, MT, MMFS). Drafting of the work or revising it critically for important intellectual content (RW, MT, MMFS). We can confirm that all authors (RW, MT, MMFS) approved the final version of the manuscript, agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved and that all persons designated as authors qualify for authorship, and all those who qualify for authorship are listed. All authors read and approved the final manuscript.

Corresponding author

Correspondence to M. M. F. Subhan.

Ethics declarations

Ethics approval and consent to participate

Informed consent was obtained from all subjects and/or their legal guardian(s). All procedures performed in studies involving human participants were in accordance with the ethical standards of the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. The study was approved by the Faculty of Science and Engineering Research Ethics Committee University of Plymouth.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Supplementary data file 1. This file show all the data for the 51 subjects who were selected in this study.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wilding, R.J., Thynne, M. & Subhan, M.M.F. Optimization of sniff nasal inspiratory pressure (SNIP) measurement methodology in healthy subjects. BMC Pulm Med 23, 66 (2023). https://doi.org/10.1186/s12890-023-02348-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-023-02348-0

Keywords