Skip to main content

Domiciliary monitoring of exhaled nitric oxide in the management of asthma: a pilot study

Abstract

Background

Whether asthma patients could benefit from home monitoring for fractional exhaled nitric oxide (flow of 50 mL/s, FeNO50) is unknown. We explore the application value of home monitoring FeNO50 in daily asthma management.

Methods

Twenty-two untreated, uncontrolled asthma patients were selected. Medical history, blood and sputum samples, pulmonary function, Asthma Control Test (ACT), and other clinical data of the subjects were collected. All subjects underwent daily monitoring for four weeks using a FeNO50 monitor and mobile spirometry (mSpirometry). The diurnal differences and dynamic changes were described. Compare the effect-acting time and the relative plateau of treatment between FeNO50 and mSpirometry monitoring.

Results

In the first two weeks, the morning median (IQR) level of FeNO50 was 44 (35, 56) ppb, which was significantly higher than the evening median level [41 (32, 53) ppb, P = 0.028]. The median (IQR) effect-acting time assessed by FeNO50 was 4 (3, 5) days, which was significantly earlier than each measure of mSpirometry (P < 0.05). FeNO50 reached the relative plateau significantly earlier than FEV1 (15 ± 2 days vs. 21 ± 3 days, P < 0.001). After treatment, the daily and weekly variation rates of FeNO50 showed a gradually decreasing trend (P < 0.05). The ACT score, sputum eosinophils, and blood eosinophils also significantly improved (P ≤ 0.01).

Conclusions

The daily home monitoring of FeNO50 in asthmatic patients showed significant circadian rhythm, and the sensitivity of FeNO50 in evaluating the response to treatment was higher than mSpirometry. The daily and weekly variation rates of FeNO50 change dynamically with time, which may be used to assess the condition of asthma.

Peer Review reports

Introduction

Bronchial asthma is a common chronic airway inflammatory disease that causes substantial economic and social burdens [1, 2]. Global Initiative for Asthma (GINA) proposes that the long-term goal of asthma treatment is to achieve symptom control and reduce the risk of acute exacerbations, irreversible airflow limitation, and treatment side effects [3]. At the same time, assessment, adjustment, and monitoring form a continuous cycle in asthma treatment and management strategy. Thus, effective self-assessment and monitoring are vital for asthma patients to achieve long-term treatment goals [4]. However, previous studies have shown that the natural history of asthma is heterogeneous and complex, characterized by circadian rhythms and long-term dynamic changes [5]. Therefore, daily asthma management should include dynamic changes, not just the absolute value of a single measurement.

Studies have shown inconsistencies in symptoms, lung function, and airway inflammation of asthma patients [6]. Self-management based on symptoms and peak expiratory flow (PEF) may still miss patients at risk of severe acute exacerbations in the future [7, 8]. There are also inconsistencies between physician and patient in the assessment of asthma control [9]. So, perhaps more methods are needed for daily asthma assessment. Fractional exhaled nitric oxide (flow of 50 mL/s, FeNO50) is one of the tools to assess airway inflammation, which is non-invasive, simple, rapid, and is currently mainly used in hospital scenarios [10]. Studies have shown that increasing the number and frequency of FeNO50 monitoring helps predict asthma control status [11]. The daily fluctuation of FeNO50 with different asthma control statuses is different [12]. Studying the long-term variation pattern of FeNO50 measurements makes it possible to identify the risk of acute exacerbations [13,14,15]. Simultaneous daily monitoring of FeNO50 can also assess compliance [10, 16, 17] and responsiveness to inhaled corticosteroids (ICS) for Type 2 asthma [18,19,20,21]. In summary, applying FeNO50 in daily asthma self-monitoring may be significant for asthma management.

More convenient products are applied to self-monitor chronic diseases with the development of science and technology and improved economic levels [7]. The emergence of daily home monitoring devices and innovative applications has made it possible to monitor circadian rhythms and daily changes for asthma patients. However, whether asthma patients could benefit from it is not fully known due to the lack of prognostic data [22]. We attempted to analyze the pattern of longitudinal dynamic changes in FeNO50 after treatment in uncontrolled asthmatic patients who were not regularly treated. To initially explore the value of FeNO50 domiciliary monitoring in the daily management of asthma.

Methods

Subjects

Twenty-two asthma patients who visited the respiratory outpatient of China-Japan Friendship Hospital from October 2019 to December 2021 were prospectively included.

Inclusion criteria: (1) age ≥ 18 years old; (2) fulfilled the diagnostic criteria of bronchial asthma defined by GINA 2018 [3]; (3) asthma symptom control was assessed as uncontrolled according to GINA2018 [3].

Exclusion criteria: (1) subjects underwent other interventional clinical trials 30 days before enrollment; (2) subjects with other pulmonary diseases or other severe system diseases that may affect the conduct of the study; (3) subjects with a smoking index > 10 pack-years and a history of smoking for nearly one year; (4) subjects had been on regular asthma therapy within 12 weeks before enrollment; (5) subjects had respiratory tract infection within four weeks before enrollment.

Written informed consent was obtained from each participant. The China-Japan Friendship Hospital ethics committee approved this study (No. 2018-19-k14, approval date: February 6th, 2018).

Study design

Enrollment stage: All subjects performed the asthma control test (ACT) [23], mini-asthma quality of life questionnaire (mini-AQLQ) [24], differential blood count, serum total IgE (enzymatic chemiluminescence, Beckmen Coulter, USA), spirometry, FeNO50 (NIOX Vero, Circassia (Beijing) Medical Device Co., Beijing, China), differential induced sputum count. Mobile spirometry (A1, Breath Home, China) [25] and FeNO50 monitor (NIOX Vero, Circassia (Beijing) Medical Device Co., Beijing, China) were provided to each subject to measure FeNO50 and mSpirometry twice a day over four weeks at home. On the day of enrollment, subjects were trained on using the equipment mentioned above (viewing usage videos and on-site instruction). Subjects were contacted during subsequent use to ensure they correctly mastered the usage methods.

The treatment strategy is not affected by the study, and the patient's treatment plan is formulated by the physician, with medication recommendations based on the GINA 2018 guidelines [3].

Domiciliary monitor stage: Subjects were asked to measure FeNO50 and mSpirometry twice daily during the same period, between 06:00 to 08:00 and 20:00 to 22:00, respectively. Before medication, FeNO50 measurement was first taken, and the results were automatically recorded on the monitor. When the subjects returned the device, all measurements were transmitted to the computer. Mobile spirometry was taken three times, and the best of the three readings were automatically uploaded to an electronic diary card. Peak expiration flow (PEF), forced expiratory volume in 1 second (FEV1), forced vital capacity (FVC), the maximum expiratory flow rate at 75%/50%/25% of the vital capacity (MEF75, MEF50, and MEF25) were collected by mSpirometry.

End of follow-up: ACT, mini-AQLQ, spirometry, FeNO50, differential blood, and induced sputum count were reviewed again.

Statistical Analysis

The normal distribution data were represented by mean ± standard deviation (sd). Non-normally distributed data were shown as median (interquartile range, IQR). A comparison of each measure before and after follow-up was performed using paired samples t-test and Wilcoxon signed-rank test. The categorical variables were expressed by frequency (composition ratio or percentage) and compared by the chi-square test.

Indicators and calculation formulas representing the variation of FeNO50 and mSpirometry: diurnal variation rate = (highest in a day - lowest in a day)/ (mean of highest versus lowest in a day)×100; mean daily variation rate = mean of diurnal variation rate over 1week; weekly variation rate = (highest over two weeks - lowest over two weeks)/ (mean of highest versus lowest over two weeks)×100.

A repeated-measures analysis of variance was used to compare diurnal differences in FeNO50 logarithmic transformed values and mSpirometry. The least-square method was used to perform curve fitting for each subject's daily monitoring results of FeNO50 and mSpirometry, and the second derivative was used to calculate the inflection point. The inflection point in this study was the transition point for the improvement of FeNO50 and mSpirometry, and its progress slowed down after the inflection point. The relative plateau of treatment was defined as the time to reach the inflection point [26]. The effect-acting time was calculated with a FeNO50 reduction of more than 20% and an improvement in mSpirometry of more than 10% as criteria [10, 27]. ANOVA and Friedman's test were used to compare the relative plateau and the effect-acting of treatment between FeNO50 and mSpirometry. Friedman's test was also used to compare the differences in variation rates of FeNO50 and mSpirometry. A two-tailed p-value of < 0.05 was considered significant. All statistical analyses were performed with SPSS 20 (IBM-SPSS, Armonk, NY, USA) and Matlab software (Mathworks, Inc., Natick, MA, USA).

Results

Characteristics of subjects

A total of 22 subjects were included in the study, with 40 ± 14 years old (range 18-64 years). There were slightly more female subjects (n = 14, 63.6%), and the BMI was 23.74 ± 4.60 kg/m2. The median (IQR) duration of illness was 2.75 (0.84, 8.87) years. The main co-morbidities of the subjects were allergic rhinitis (90.9%), nasal polyps (18.2%), and eczema (22.7%). A history of allergy was present in 59.1% of the subjects, and the median (IQR) serum total IgE level was 185.50 (86.88, 470.75) IU/ml. All subjects received salmeterol/fluticasone (50/250 μg) twice daily, with three subjects receiving montelukast and three subjects receiving tiotropium.

The effective FeNO50 and mSpirometry monitoring were 1035 and 991 times, respectively, and the overall adherence rates were 84.0% (1035/1232) and 80.4% (991/1232), respectively. Of the enrolled subjects, FeNO50 baseline level ≥ 50 ppb in 19 (86.4%) patients; sputum eosinophil baseline level ≥ 3% in 21 (95.4%) patients; and 16 (72.7%) patients had blood eosinophils baseline level ≥ 300/μL; 15 patients (68.2%) had FEV1% baseline level > 80%.

Assessment of the disease condition before and after treatment

After four weeks of treatment, the subjects showed significant improvement in all measurements (Table 1). A significant improvement in the ACT (improvement of more than 3 points) [23] was observed in 17 subjects, with an ACT score ranging from 16 ± 4 points improved to 23 ± 2 points (P < 0.001). In mini-AQLQ, overall score, symptoms, activity, emotion, and environmental scores improved significantly (P < 0.01). The pulmonary function was significantly improved (P ≤ 0.01). In terms of inflammation, either FeNO50 or sputum and blood eosinophils were significantly reduced (P < 0.01) (Table 1).

Table 1 Assessment of the disease condition before and after treatment

Diurnal variation of Fe NO50 and mSpirometry

There was a significant diurnal difference in FeNO50 daily monitoring in the first two weeks. The morning median (IQR) level of FeNO50 was 44 (35, 56) ppb, which was significantly higher than the evening median level [41 (32, 53) ppb, P = 0.03]. However, after two weeks of treatment, the significant difference between day and night disappeared (P = 0.17). In our study, no significant differences were found in any indicators of mSpirometry between day and night (Fig. 1).

Fig. 1
figure 1

Diurnal variation curves of domiciliary monitoring. a FeNO50 diurnal variation curves, b FEV1 diurnal variation curves, c PEF diurnal variation curves, d MEF75 diurnal variation curves, e MEF50 diurnal variation curves, f MEF25 diurnal variation curves. FeNO50: fractional exhaled nitric oxide (flow of 50 mL/s), FEV1: forced expiratory volume in 1-second, PEF: peak expiratory flow, MEF75/ MEF50/ MEF25: maximum expiratory flow rate at 75%/ 50%/ 25% of the vital capacity

Dynamic fluctuation of Fe NO50 and mSpirometry

The best curve fit of the daily monitoring indices of FeNO50 and mSpirometry for the subject is shown in Fig. 2. In general, FeNO50 and Spirometry improved gradually over time with treatment. The median (IQR) effect-acting time assessed by FeNO50 was 4 (3, 5) days, which was significantly earlier than mSpirometry (P < 0.05) (Table 2). After Bonferroni correction, FeNO50 reached the relative plateau significantly earlier than FEV1 (15 ± 2 days vs. 21 ± 3 days, P < 0.001), but there was no statistical difference with other indicators of mSpirometry (Table 2).

Fig. 2
figure 2

Fitting curve of domiciliary monitoring. a Fitting curve of FeNO50 (R2 = 0.698 ± 0.239), b Fitting curve of FEV1 (R2 = 0.476 ± 0.270), c Fitting curve of PEF (R2 = 0.564 ± 0.220), d Fitting curve of MEF75 (R2 = 0.499 ± 0.203), e Fitting curve of MEF50 (R2 = 0.503 ± 0.237), f Fitting curve of MEF25 (R2 = 0.449 ± 0.158). FeNO50: fractional exhaled nitric oxide (flow of 50 mL/s), FEV1: forced expiratory volume in 1 second, PEF: peak expiratory flow, MEF75/ MEF50/ MEF25: maximum expiratory flow rate at 75%/ 50%/ 25% of the vital capacity, R2: coefficient of determination

Table 2 Exhaled nitric oxide fraction and mobile spirometry assessment therapeutic effect

Variation rates of Fe NO50 and mSpirometry

After ICS treatment, the daily variation rates of FeNO50 and mSpirometry showed a gradually decreasing trend. The average daily variation rates of FeNO50 and FEV1 at week 1 were significantly higher than those at week 4 (P < 0.05), while there was no significant difference in PEF, MEF75, MEF50, and MEF25. (Fig. 3). There was a substantial reduction in weekly variation rates of FeNO50 and mSpirometry among the participants (P < 0.05) (Fig. 4).

Fig. 3
figure 3

Daily variation rate curves of domiciliary monitoring. a Daily variation rate curves of FeNO50, b Daily variation rate curves of FEV1, c Daily variation rate curves of PEF, d Daily variation rate curves of MEF75, e Daily variation rate curves of MEF50, f Daily variation rate curves of MEF25. FeNO50: fractional exhaled nitric oxide (flow of 50 mL/s), FEV1 forced expiratory volume in 1 second, PEF: peak expiratory flow, MEF75/ MEF50/ MEF25: maximum expiratory flow rate at 75%/ 50%/ 25% of the vital capacity

Fig. 4
figure 4

Weekly variation rates of domiciliary monitoring. a Weekly variation rate of FeNO50, b Weekly variation rate of FEV1, c Weekly variation rate of PEF, d Weekly variation rate of MEF75, e Weekly variation rate of MEF50, f Weekly variation rate of MEF25. FeNO50 fractional exhaled nitric oxide (flow of 50 mL/s), FEV1: forced expiratory volume in 1-second, PEF: peak expiratory flow, MEF75/ MEF50/ MEF25: maximum expiratory flow rate at 75%/ 50%/ 25% of the vital capacity

Discussion

Through daily domiciliary monitoring of untreated, uncontrolled asthmatic patients, we found significant diurnal differences and daily dynamic changes in FeNO50 and mSpirometry, which can be used to evaluate asthma patients' response to treatment and condition assessment.

FeNO50 is a sensitive biomarker that reflects the inflammation of airway eosinophils [28]. In the early stage of treatment, FeNO50 was significantly higher in the morning than at night in asthmatic patients, consistent with previous findings [21, 29]. Although studies have shown that the decrease of FEV1 affects the results of FeNO50, the effect of the change in airway diameter on FeNO50 will be offset when the airway inflammation is higher [30, 31]. However, the disappearance of diurnal differences in asthmatic patients after treatment may be related to airway inflammation and lung function improvement [32, 33].

The primary purpose of this study is to observe the impact of home monitoring on evaluating treatment outcomes. After four weeks of treatment, all participants showed improvement in symptoms and inflammation levels. Therefore, although different treatment plans may cause biases, we believe the impact on the conclusion is relatively tiny.

Peak-trough times of lung function and biomarkers in asthma patients may differ due to individual chronotype differences [34]. Lung function fluctuates between 6-7 days in Fig. 1, but there is no significant difference. Although the measurement period was specified to avoid bias, the personal measurement time or additional measurement period was not selected according to the living habits, thus reducing the reliability of dynamic variation. No significant diurnal differences in lung function were found in this study, which may be related to the relatively good lung function of the included subjects, reducing the sensitivity of diurnal differences.

We characterized the changing trend of FeNO50 in treated Type 2 asthmatic patients. We found that FeNO50 could detect the treatment response for 3-5 days and reached a relative plateau for two weeks. Various indices (PEF, FEV1, MEF75, MEF50, MEF25) of mSpirometry improved gradually, with treatment effect-acting time around 8~12 days and reached a relative plateau of improvement around 2-3 weeks [26, 35]. However, due to the relatively low goodness of fit of the fitting curves for mSpirometry in this study, the treatment turning point of mSpirometry still needs further validation with extensive sample data.

Daily monitoring found that the treatment effect assessed by FeNO50 was significantly earlier than mSpirometry, and the time to reach the relative plateau of treatment was substantially earlier than FEV1. As we can see, FeNO50 was more sensitive than mSpirometry in assessing responsiveness to asthma therapy. Meanwhile, the latest research indicates that FeNO50 is a risk biomarker identifying patients at increased risk of lung function decline [36].

After treatment, the daily and weekly variation rates of FeNO50 and mSpirometry showed a decreasing trend. Our study found that the average daily variation of FeNO50 in the first week was significantly higher than in the fourth week. Therefore, the improvement of diurnal variation in FeNO50 can also be used to evaluate the effectiveness of treatment in asthma. This variability over time suggests that the domiciliary monitoring strategy has the advantage of detecting daily and long-term changes in physiology and inflammation, providing substantial evidence to predict future exacerbations and disease assessment. After four weeks of treatment, the participants showed significant improvements in asthma control, quality of life, lung function, and inflammation. With the considerable improvement of symptoms and clinical indicators, the variation rate of domiciliary monitors gradually decreased, which is an essential clinical signal to evaluate the changes in the conditions of asthma patients.

The study had limitations. Firstly, as our study is a pilot study, all admitted patients were untreated type 2 asthma. We did not include patients with severe asthma. Still, previous studies have shown that FeNO50 can also serve as an inflammatory marker for evaluating treatment response in severe asthma populations. [37]. A large sample study was needed to demonstrate the general generalization of the variation pattern. Secondly, complex device use, frequency of daily readings, and strict monitoring times reduce the completion of measures [38, 39]. Still, previous studies have also shown that remote monitoring devices allow subjects to understand their self-control levels and improve patient compliance [40, 41]. Although there was no monitoring of drug adherence, the completion rate of this study exceeded 80%, indirectly indicating that drug adherence is still acceptable. Thirdly, our study could not compare mSpirometry A1 with the lung function laboratory device. The study found that only a portion of mSpirometry A1 (BreathHome, China) met the quality and performance evaluation standards [25]. However, considering this study mainly observed dynamic changes, the research results are still acceptable.

In conclusion, this pilot study of domiciliary monitoring found that FeNO50 in uncontrolled asthma patients has significant diurnal differences and is superior to mSpirometry in assessing sensitivity to treatment response. The dynamic fluctuation of FeNO50 may be available for evaluating disease conditions in asthmatics. Studies with large samples and long observation periods are expected to explore mobile domiciliary monitors in asthma management.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

ACT :

Asthma control test

BMI :

Body mass index

Fe NO50 :

Fractional exhaled nitric oxide (flow of 50 mL/s)

FEV 1 :

Forced expiratory volume in 1 second

FVC :

Forced vital capacity

GINA :

Global initiative for asthma

ICS :

inhaled corticosteroids

IQR :

interquartile range

MEF 75 :

Maximum expiratory flow rate at 75% of the vital capacity

MEF 50 :

Maximum expiratory flow rate at 50% of the vital capacity

MEF 25 :

Maximum expiratory flow rate at 25% of the vital capacity

mini-AQLQ :

mini Asthma Quality of Life Questionnaire

mSpirometry :

mobile spirometry

PEF :

peak expiratory flow

References

  1. Huang K, Yang T, Xu J, et al. Prevalence, risk factors, and management of asthma in China: a national cross-sectional study. Lancet. 2019;394(10196):407–18.

    Article  PubMed  Google Scholar 

  2. Haselkorn T, Szefler SJ, Chipps BE, et al. Disease Burden and Long-Term Risk of Persistent Very Poorly Controlled Asthma: TENOR II. J Allergy Clin Immunol Pract. 2020;8(7):2243–53.

    Article  PubMed  Google Scholar 

  3. Stewart L, Katial RK. Exhaled nitric oxide. Immunol Allergy Clin North Am. 2012;32(3):347–62.

    Article  PubMed  Google Scholar 

  4. Janson SL, McGrath KW, Covington JK, et al. Individualized asthma self-management improves medication adherence and markers of asthma control. J Allergy Clin Immunol. 2009;123(4):840–6.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Wang R, Murray CS, Fowler SJ, et al. Asthma diagnosis: into the fourth dimension. Thorax. 2021;76(6):624–31.

    Article  PubMed  Google Scholar 

  6. Voorend-van Bergen S, Vaessen-Verberne AA, Landstra AM, et al. Fractional Exhaled Nitric Oxide Monitoring Does Not Improve Asthma Management in Children with Concordant and Discordant Asthma Phenotypes. Am J Respir Crit Care Med. 2015;192(8):1016–8.

    Article  CAS  PubMed  Google Scholar 

  7. Castner J, Klingman K, Sullivan S, et al. Hitting home with technology development for asthma. Lancet Respir Med. 2016;4(2):102–3.

    Article  PubMed  Google Scholar 

  8. Lo DK, Beardsmore CS, Roland D, et al. Lung function and asthma control in school-age children managed in UK primary care: a cohort study. Thorax. 2020;75(2):101–7.

    Article  PubMed  Google Scholar 

  9. Fuhlbrigge A, Marvel J, Electricwala B, et al. Physician-Patient Concordance in the Assessment of Asthma Control. J Allergy Clin Immunol Pract. 2021;9(8):3080-3088.e3081.

    Article  PubMed  Google Scholar 

  10. Dweik RA, Boggs PB, Erzurum SC, et al. An official ATS clinical practice guideline: interpretation of exhaled nitric oxide levels (FENO) for clinical applications. Am J Respir Crit Care Med. 2011;184(5):602–15.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Yang S, Park J, Lee YK, et al. Association of longitudinal fractional exhaled nitric oxide measurements with asthma control in atopic children. Respir Med. 2015;109(5):572–9.

    Article  PubMed  Google Scholar 

  12. Saito J, Gibeon D, Macedo P, et al. Domiciliary diurnal variation of exhaled nitric oxide fraction for asthma control. Eur Respir J. 2014;43(2):474–84.

    Article  PubMed  Google Scholar 

  13. van der Valk RJP, Baraldi E, Stern G, et al. Daily exhaled nitric oxide measurements and asthma exacerbations in children. Allergy. 2012;67(2):265–71.

    Article  PubMed  Google Scholar 

  14. Stern G, de Jongste J, van der Valk R, et al. Fluctuation phenotyping based on daily fraction of exhaled nitric oxide values in asthmatic children. J Allergy Clin Immunol. 2011;128(2):293–300.

    Article  CAS  PubMed  Google Scholar 

  15. Petsky HL, Cates CJ, Kew KM, et al. Tailoring asthma treatment on eosinophilic markers (exhaled nitric oxide or sputum eosinophils): a systematic review and meta-analysis. Thorax. 2018;73(12):1110–9.

    Article  PubMed  Google Scholar 

  16. McNicholl DM, Stevenson M, McGarvey LP, et al. The utility of fractional exhaled nitric oxide suppression in the identification of nonadherence in difficult asthma. Am J Respir Crit Care Med. 2012;186(11):1102–8.

    Article  PubMed  Google Scholar 

  17. Heaney LG, Busby J, Bradding P, et al. Remotely Monitored Therapy and Nitric Oxide Suppression Identifies Nonadherence in Severe Asthma. Am J Respir Crit Care Med. 2019;199(4):454–64.

    Article  CAS  PubMed  Google Scholar 

  18. Buchvald F, Eiberg H, Bisgaard H. Heterogeneity of FeNO response to inhaled steroid in asthmatic children. Clin Exp Allergy. 2003;33(12):1735–40.

    Article  CAS  PubMed  Google Scholar 

  19. Cowan DC, Taylor DR, Peterson LE, et al. Biomarker-based asthma phenotypes of corticosteroid response. J Allergy Clin Immunol. 2015;135(4):877-883.e871.

    Article  CAS  PubMed  Google Scholar 

  20. Kharitonov SA, Donnelly LE, Montuschi P, et al. Dose-dependent onset and cessation of action of inhaled budesonide on exhaled nitric oxide and symptoms in mild asthma. Thorax. 2002;57(10):889–96.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Anderson WJ, Short PM, Williamson PA, et al. Inhaled corticosteroid dose response using domiciliary exhaled nitric oxide in persistent asthma: the FENOtype trial. Chest. 2012;142(6):1553–61.

    Article  CAS  PubMed  Google Scholar 

  22. Carpenter DM, Jurdi R, Roberts CA, et al. A Review of Portable Electronic Spirometers: Implications for Asthma Self-Management. Curr Allergy Asthma Rep. 2018;18(10):53.

    Article  PubMed  Google Scholar 

  23. Nathan RA, Sorkness CA, Kosinski M, et al. Development of the asthma control test: a survey for assessing asthma control. J Allergy Clin Immunol. 2004;113(1):59–65.

    Article  PubMed  Google Scholar 

  24. Juniper EF, Guyatt GH, Cox FM, et al. Development and validation of the Mini Asthma Quality of Life Questionnaire. Eur Respir J. 1999;14(1):32–8.

    Article  CAS  PubMed  Google Scholar 

  25. Wu Z, Huang R, Zhong L, et al. Technical performance analysis of different types of spirometers. BMC Pulm Med. 2022;22(1):23.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Reddel HK, Marks GB, Jenkins CR. When can personal best peak flow be determined for asthma action plans? Thorax. 2004;59(11):922–4.

    Article  CAS  PubMed Central  Google Scholar 

  27. Szefler SJ, Martin RJ, King TS, et al. Significant variability in response to inhaled corticosteroids for persistent asthma. J Allergy Clin Immunol. 2002;109(3):410–8.

    Article  CAS  PubMed  Google Scholar 

  28. Nanda CR, Singapuri A, Soares M, et al. Domiciliary exhaled nitric oxide and eosinophilic airway inflammation in adults with asthma. Eur Respir J. 2016;48(1):242–4.

    Article  PubMed  Google Scholar 

  29. Pijnenburg MW, Floor SE, Hop WC, et al. Daily ambulatory exhaled nitric oxide measurements in asthma. Pediatr Allergy Immunol. 2006;17(3):189–93.

    Article  PubMed  Google Scholar 

  30. Garnier P, Fajac I, Dessanges JF, et al. Exhaled nitric oxide during acute changes of airways calibre in asthma. Eur Respir J. 1996;9(6):1134–8.

    Article  CAS  PubMed  Google Scholar 

  31. Haccuria A, Michils A, Michiels S, et al. Exhaled nitric oxide: a biomarker integrating both lung function and airway inflammation changes. J Allergy Clin Immunol. 2014;134(3):554–9.

    Article  CAS  PubMed  Google Scholar 

  32. Gochicoa-Rangel L, Rojas-Cisneros F, Miguel-Reyes JL, et al. Variability of FeNO in healthy subjects at 2240 meters above sea level. J Clin Monit Comput. 2016;30(4):445–9.

    Article  PubMed  Google Scholar 

  33. Ekroos H, Karjalainen J, Sarna S, et al. Short-term variability of exhaled nitric oxide in young male patients with mild asthma and in healthy subjects. Respir Med. 2002;96(11):895–900.

    Article  CAS  PubMed  Google Scholar 

  34. Adan A, Archer SN, Hidalgo MP, et al. Circadian typology: a comprehensive review. Chronobiol Int. 2012;29(9):1153–75.

    Article  PubMed  Google Scholar 

  35. Reddel HK, Jenkins CR, Marks GB, et al. Optimal asthma control, starting with high doses of inhaled budesonide. Eur Respir J. 2000;16(2):226–35.

    Article  CAS  PubMed  Google Scholar 

  36. Pavord ID, de Prado Gómez L, Brusselle G, et al. Biomarkers Associated with Lung Function Decline and Dupilumab Response in Patients with Asthma. Am J Respir Crit Care Med. 2024;209(8):1031–3.

    Article  CAS  PubMed  Google Scholar 

  37. Couillard S, Shrimanker R, Lemaire-Paquette S, et al. Longitudinal changes in sputum and blood inflammatory mediators during FeNO suppression testing. Thorax. 2022;77(9):933–8.

    Article  PubMed  Google Scholar 

  38. Reddel H, Jenkins C, Woolcock A. Diurnal variability–time to change asthma guidelines? Bmj. 1999;319(7201):45–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. Côté J, Cartier A, Malo JL, et al. Compliance with peak expiratory flow monitoring in home management of asthma. Chest. 1998;113(4):968–72.

    Article  PubMed  Google Scholar 

  40. van Vliet D, van Horck M, van de Kant K, et al. Electronic monitoring of symptoms and lung function to assess asthma control in children. Ann Allergy Asthma Immunol. 2014;113(3):257-262.e251.

    Article  PubMed  Google Scholar 

  41. Feldman JM, Kutner H, Matte L, et al. Prediction of peak flow values followed by feedback improves perception of lung function and adherence to inhaled corticosteroids in children with asthma. Thorax. 2012;67(12):1040–5.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

Thanks to Hongyi Wang (Circassia (Beijing) Medical Device Co., Beijing, China), Yang Wang (Circassia (Beijing) Medical Device Co., Beijing, China, and Yang Dong (Breath Home, Beijing, China) provided technical help.

Submission declaration

We confirm that the original manuscript has not been published before, is not currently being considered for publication elsewhere, and has not been posted to a preprint server.

Funding

This work was funded by the Natural Science Foundation of Shandong Province (Youth Program ZR2022QH371).

Author information

Authors and Affiliations

Authors

Contributions

H.L. and J.L. designed the study. J.W. and C.L. acquisitions of data or analysis and interpretation of data. Q.Z. and H.L. statistical analysis. H.L. drafting articles. All authors revise the article critically for important intellectual content and final approval of the version to be published.

Corresponding author

Correspondence to Jiangtao Lin.

Ethics declarations

Ethics approval and consent to participate

The study protocol was approved by the Ethics Committee of China-Japan Friendship Hospital (approval number: 2018-19-k14). Written informed consent was obtained from each participant.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, H., Lin, J., Zhang, Q. et al. Domiciliary monitoring of exhaled nitric oxide in the management of asthma: a pilot study. BMC Pulm Med 24, 244 (2024). https://doi.org/10.1186/s12890-024-03031-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-024-03031-8

Keywords