Diagnostic accuracy of nasal nitric oxide for establishing diagnosis of primary ciliary dyskinesia: a meta-analysis

Abstract

Background

To date, diagnosis of Primary Ciliary Dyskinesia (PCD) remains difficult and challenging. We systematically evaluated the diagnostic performance of nasal Nitric Oxide (nNO) measurement for the detection of PCD, using either velum-closure (VC) or non-velum-closure (non-VC) techniques.

Methods

All major electronic databases were searched from inception until March 2015 using appropriate terms. The sensitivity and specificity of nNO measurement was calculated in PCD patients diagnosed by transmission electron microscopy, high speed video-microscopy or genetic testing. Summary receiver operating characteristic (HSROC) curves were drawn using the parameters of the fitted models.

Results

Twelve studies provided data for 13 different populations, including nine case–control (n = 793) and four prospective cohorts (n = 392). The overall sensitivity of nNO measured by VC techniques was 0.95 (95 % CI 0.91–0.97), while specificity was 0.94 (95 % CI 0.88–0.97). The positive likelihood ratio (LR+) of the test was 15.8 (95 % CI 8.1–30.6), whereas the negative likelihood ratio (LR-) was 0.06 (95 % CI 0.04–0.09). For non-VC techniques, the overall sensitivity of nNO measurement was 0.93 (95 % CI 0.89–0.96) whereas specificity was 0.95 (95 % CI 0.82–0.99). The LR+ of the test was 18.5 (95 % CI 4.6–73.8) whereas the LR- was 0.07 (95 % CI 0.04–0.12).

Conclusions

Diagnostic accuracy of nNO measurement both with VC and non-VC maneuvers is high and can be effectively employed in the clinical setting to detect PCD even in young children, thus potentiating early diagnosis. Measurement of nNO merits to be part of a revised diagnostic algorithm with the most efficacious combination of tests to achieve PCD diagnosis.

Peer Review reports

Background

Primary ciliary dyskinesia (PCD) is a rare, hereditary disorder characterized by impaired mucociliary clearance [1]. Apart from situs inversus in ~50 % of the cases, the main manifestations of the disease are not specific. Nevertheless, the associated recurrent sinopulmonary infections eventually lead to severe chronic lung disease and development of bronchiectasis [2, 3].

While some centers began using targeted genetic testing [4], the diagnosis of PCD in the majority of centers currently relies on an array of different sophisticated tests namely the High Speed Video Microscopy (HSVM) for ciliary motility assessment [5], Transmission Electron Microscopy (TEM) for the examination of cilia ultrastructure [6] and nasal nitric oxide (nNO) measurement [7]. The diversity of the employed diagnostic tests reflects the lack of a golden diagnostic standard and the weaknesses and inaccuracies that characterize each of these tests. In particular, TEM examination of ciliary axonemes exhibits normal ultrastructure in confirmed patients with biallelic mutations in certain disease-causing genes such as DNAH11 [8], while the motility patterns observed by HSVM vary widely depending on the implicated genetic variant [9, 10].

Nasal nitric oxide (nNO) is abnormally low in PCD patients [11] and it has been part of the diagnostic work-up in many PCD centers [12]. Current American Thoracic Society/European Respiratory Society (ATS/ERS) guidelines for nNO measurements recommend air aspiration via a nasal probe while the subject exhales through the mouth against resistance in order to maintain velum closure. Alternative techniques to maintain velum closure such as breath hold or pursed- lip breathing via the mouth are also acceptable [13]. However, velum closure requires cooperation and this precludes the performance of these techniques in young children. Few reports have investigated the discriminative ability of nNO measurements with the velum open as in the case of tidal breathing [3, 14] with encouraging findings for the usefulness of this technique in screening for PCD in younger children and adults unable to perform velum closure.

In view of the above specific restrictions and weaknesses, for the clinicians and the patients it remains of key importance to appraise the potential diagnostic value of each of the available diagnostic tests for PCD, in order to find its place in the armamentarium for elicitation of the diagnosis of the disease. A recent systematic review and meta-analysis summarized the published evidence on the measurement of nNO in PCD and reported on the mean difference of nNO production values obtained during velum closure techniques in PCD patients versus healthy controls (231 nL/min, 95 % CI: 193.3–268.9) and cystic fibrosis patients (114.1 nL/min, 95 % CI: 101.5–126.8) [15]. However, that report did not perform a meta-analysis on the diagnostic accuracy of nNO measurements in order to provide synthesized data on the potential diagnostic value this test may have in future algorithms for PCD diagnosis, which would be particularly informative in clinical decision making. The aim of this study was to systematically evaluate the diagnostic performance of nNO measurement as obtained either with a velum-closure or a non-velum-closure technique in screening for PCD so as to provide appropriate summary estimates of diagnostic accuracy with each breathing technique and demonstrate the summary trade-off between sensitivity and specificity across the included studies.

Methods

Search strategy and selection criteria

The electronic databases PubMed, SCOPUS, Cochrane Database of Systematic Reviews and Google Scholar were searched from inception until March 2015 using the keywords: ‘nasal nitric oxide’, “nNO”, “nasal NO”, “Primary Ciliary Dyskinesia”, “PCD”, “lung”, “pulmonary”, “pulm*”, “cilia” either in the title or the abstract or using MeSH terms. The references of eligible studies were further examined for possible missing articles. We included studies which were identified after two reviewers (PK, SIP) independently screened the title and abstract of the obtained search results. Final selection was based on full text evaluation. Any disagreements were resolved by discussion and in case of discrepancy, by a third researcher (PKY). As this study is based on a systematic review of the previously published literature, an ethical approval was not obtained, since there is no potential of participant identification and ethical approval and consent was already obtained at the individual study level. The guidelines of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) were followed.

The validity of each primary study was assessed using the Quality Assessment of Diagnostic Accuracy Studies −2 (QUADAS-2) tool [16], that evaluates the risk of bias and applicability of diagnostic accuracy studies. It consists of four key domains: patient selection, index test, reference standard, flow and timing. Each is assessed in terms of risk of bias and the first three in terms of issues regarding applicability.

Studies were considered eligible if they provided data on the sensitivity and specificity of nNO for the diagnosis of PCD in order to construct a 2 × 2 table for each study calculating true positives (TP), false negatives (FN), false positives (FP) and true negatives (FN) for the presence or not of PCD according to nNO values set as a cut-off in each study. In some studies, the numbers were not provided per se but it was possible to extract them from other manuscript data sources. In case of incomplete information, we contacted the authors of the primary studies. Studies that reported only mean values of nNO were not included in our analyses as they did not provide data for computing summary diagnostic accuracy estimates (sensitivity, specificity, positive and negative likelihood ratio). Disease status in each selected study was required to have been confirmed by TEM and/or HSVM or genetic testing. Additional information on NO analyzer type, flow rate and breathing maneuver was also collected and used in data synthesis. Studies that did not report the equipment and flow rate used were not considered eligible as well as studies that may have used flow rate outside of the ATS/ERS recommended range (0.25–3 L/min) [13]. Cut-off values for the nNO test, were usually reported in parts per billion (ppb) and were transformed to NO production rate units (nl/min), using the conversion formula concentration (ppb) × sampling rate (L/min) as used previously [7], in order to account for the used different flow rates. Breathing maneuvers such as breath hold (BH) and exhalation against resistance (ER) were categorized as velum closure (VC) techniques and in case of both maneuvers performed by the study subjects; only results for the ER maneuver were included as the most validated technique according to ATS/ERS guidelines [13]. For studies employing the non-velum closure (non-VC) technique, only results of nNO measurements that were performed during tidal breathing (TB) with mouth open were included in the meta-analysis.

Data extraction

The name of author, study design, publication year, country of origin, study population sample size, age distribution of study population subgroups, nNO cut-off levels, information on the measurement method and the test(s) used for the diagnosis of PCD were recorded for each study. Data on the values of TP, TN, FP, FN were extracted independently by two reviewers (PK, SIP). A third investigator (PKY) settled any discrepancies and consensus was reached for all data.

Analysis

A bivariate model was used to calculate estimates of overall sensitivity and overall specificity. We fitted a two-level mixed logistic regression model conditional on the sensitivity and the specificity of each study and a bivariate normal model for the sensitivity and specificity between studies [17]. This method combines information from multiple thresholds and the output is expressed as a hierarchical summary receiver operator curve (HSROC). The HSROC describes the relationship between sensitivity and specificity derived from the individual receiver operator curves (ROC) of each study. Following this method, it describes the ‘average’ relationship between a continuous cut-off value and discriminatory ability in the ‘average’ population. This maximizes the amount of information used in the evidence synthesis and better represents the available data. The advantage of this method is that it allows clinicians to estimate how changing thresholds will alter the diagnostic utility of the test under study. All calculations are performed using STATA (Version 12, StataCorp, College Station, Texas) with the commands metandi and metandi plots for analyses of four studies and above [18].

We also reported the summary likelihood ratios across all studies. These measures also combine in their calculation both sensitivity and specificity. Positive likelihood ratio (LR+) is the ratio of sensitivity/(1-specificity), whereas negative likelihood ratio (LR-) is defined as the ratio of (1-sensitivity)/specificity. When there is absolutely no discriminating ability for a diagnostic test, both ratios are equal to 1. The discriminating ability is better with higher LR+ and lower LR-. A good diagnostic test has typically LR+ greater than 5.0 and LR- less than 0.2 [19].

VC and non-VC measurements were analyzed separately and this allowed us to arrive at estimates on overall sensitivity, specificity and likelihood ratios for nNO depending on VC status. We also performed a sensitivity analysis including only studies in which PCD status was defined by TEM and at least one more diagnostic test, with the rationale to examine whether the diagnostic accuracy of nNO measurement differs with the inclusion of a more representative spectrum of PCD population. Measurements of nNO were compared to PCD diagnosis obtained through a combination of tests which included TEM and HSVM or DNA testing.

Results

Eligible studies

Of the 1940 items retrieved through online search, 1866 were excluded based on the title and abstract and the remaining 74 were downloaded for detailed, full text assessment. Two additional studies were identified through references screening and were also evaluated. Studies with overlapping populations were cross-checked and final selection was based on the largest number of participating PCD patients. In summary, 26 studies did not provide data on sensitivity and specificity, 13 items involved overlapping populations, 15 items were review papers while the remaining items that were excluded were case reports (2), editorials (3) and guidelines papers (2) (Fig. 1). Of the total 76 studies assessed in detail, 15 provided enough data for the construction of a 2×2 table. Among these, two studies did not report type of NO analyzer and flow rate and despite our effort to obtain this information after contacting the authors, this was not feasible and they were excluded from the analysis [20, 21]. Finally, quantitative synthesis included data on 13 different populations from 12 studies (Marthin et al. included data on more than one population) and two separate analyses were carried out, based on the employed breathing maneuver.

Fig. 1
figure 1

PRISMA diagram. PRISMA diagram for the search strategy and selected studies. From: Moher D, Liberati A, Tetzlaff J, Altman DG, The PRISMA Group (2009). Preferred Reporting Items for Systematic Reviews and Meta-Analyses: The PRISMA Statement. PLoS Med 6(6):e1000097. doi:10.1371/journal.pmed1000097. For more information, visit www.prisma-statement.org

Study characteristics

Descriptive characteristics of the included studies are presented in Table 1. From 12 studies, 325 PCD patients and 711 non-PCD subjects were included in the meta-analysis for the diagnostic performance of VC nNO testing. In the case of non-VC nNO testing, 210 PCD patients and 471 non-PCD subjects from seven studies were included. The majority of the studies were performed in Western Europe and only two in North America. Four studies evaluated the diagnostic efficacy of nNO in cohorts of referred suspect patients for PCD testing [7, 2224] whereas the rest of the studies had a case–control design. Controls were non-PCD subjects, either healthy subjects only [22, 2528] or healthy subjects and patients with other respiratory diseases [14, 2931].

Table 1 Characteristics of the included studies

The number of PCD patients (range: 9–59) and controls (range: 14–188) per case–control study varied widely. All case–control studies confirmed PCD status by TEM findings while in 55 % of them HSVM was also performed. Of the four prospective studies, Beydon et al. used a combination of TEM, HSVM and genetic testing to confirm PCD diagnosis [24] while Marthin et al. used TEM and HSVM in their cohort of consecutive referrals [22]. Leigh et al. confirmed PCD via a combination of ultrastructure assessment and genetic testing [7] while the smallest cohort study confirmed PCD only via ultrastructural assessment [23]. The sensitivity and specificity of each included study with VC and non-VC technique are shown in Fig. 2.

Fig. 2
figure 2

Forest plots for sensitivity and specificity. Forest plot of sensitivity and specificity of nNO for detecting PCD with the 95 % CI for each population of the included studies. a Forest plot for studies employing a VC breathing technique and b Forest plot for studies employing a non-VC breathing technique

Quality assessment

Reporting of the meta-analysis is based on PRISMA guidelines [32]. Based on the QUADAS-2 tool, the quality assessment of the primary studies is shown in Table 2. In general, the analyzed studies had overall reasonably good methodology and this offers relative reassurance that results have not been substantially influenced from bias.

Table 2 QUADAS-2 Quality Assessment results

Data synthesis

The overall sensitivity of abnormal (low) nNO measured by VC techniques for all the included studies was 0.95 (95 % CI 0.91–0.97), while the specificity was 0.94 (95 % CI 0.88–0.97). The LR+ of the test was 15.8 (95 % CI 8.1–30.6), whereas the LR- was 0.06 (95 % CI 0.04–0.09). The HSROC curve is shown in Fig. 3.

Fig. 3
figure 3

VC and non-VC HSROC curves. HSROC curves for the included studies. a HSROC curve for studies employing a VC breathing technique and b HSROC curve for studies employing a non-VC breathing technique

For the non-VC techniques the overall sensitivity of nNO to detect PCD was 0.93 (95 % CI 0.89–0.96) whereas the specificity was 0.95 (95 % CI 0.82–0.99). The LR+ of the test was 18.5 (95 % CI 4.6–73.8) whereas the LR- was 0.07 (95 % CI 0.04–0.12). The HSROC curve is shown in Fig. 3.

When we performed a sensitivity analysis, to calculate the nNO diagnostic accuracy in studies that only included PCD populations diagnosed by more than one test (combination of TEM and HSVM or genetic testing) [7, 14, 22, 24, 26, 27, 30], the results did not change significantly. Similarly, after performing a post hoc sensitivity analysis, with the exclusion of studies that used the electrochemical devise NIOX MINO [14, 27], the resulting estimates of overall sensitivity and overall specificity for the non-VC maneuver do not significantly differ from the estimates of the main analysis.

Discussion

In this meta-analysis we demonstrated that nNO measurement with VC techniques has overall sensitivity of 95 % and a specificity of 94 % whereas nNO measurement with the non-VC technique has comparable and very similar sensitivity (93 %) and specificity (95 %). We applied a different approach to the one employed in a previous report [15] and evaluated nNO diagnostic performance metrics by using all the available evidence in the literature. These summary estimates allow us to make comparisons between the various proposed and established diagnostic tests for PCD, which are essential for clinical decision making. We also provide a graphical representation of our results using the hierarchical summary receiver operating characteristic (HSROC) curve incorporating the different cut-offs between primary studies. The clinical utility of nNO measurement is underlined by the high LR+ (VC: 15.8, non-VC: 18.5) and low LR- (VC: 0.06, non-VC: 0.07), meaning that an abnormal (low) nNO leads to a steep increase in the post-test probability of PCD, compared to the pretest probability, while in the case of a normal nNO measurement the opposite is also true [19]. However, since the sensitivity and specificity of the test are not 100 %, in the presence of strong clinical suspicion for PCD [33], even in the case of a negative nNO test, a more detailed diagnostic work-up (HSVM, TEM, genetics) is indicated.

Current ATS/ERS recommendations for nNO include only VC maneuvers although recent evidence [14, 30, 31], that is supported by the results of this meta-analysis, highlights the discriminative ability of nNO during TB. TB is the only method available to obtain nNO measurements in young children (<5 years), which is particularly important as disease manifestations appear very early in life. Of course, the validity of nNO measurements in infants (<6 months) has been questioned, as nNO output in infancy is reduced due to the partial development of paranasal sinuses [25] where the majority of NO is produced [11] whereas the number of patients under 5 years which were evaluated in these studies [14, 22, 24, 27, 30, 31] is very small . The usefulness of nNO during TB has been demonstrated in the Danish cohort of 117 consecutive referrals with median age 6.9 years, where 83 % were able to perform TB versus 50 % for BH and 31 % for ER [22]. There is evidence that patients that have earlier diagnosis of this disease might have better clinical and functional outcomes [34, 35] and the application of this promising screening method in preschool children could not only lead to diagnosis at an earlier age but could also contribute towards the reduction of unnecessary cilia biopsies. Nevertheless, our results for the non-VC techniques should be interpreted with caution. Their low 95 % CI limit for specificity is at 82 % which suggests that a significant number of suspect PCD referrals is possible to give a falsely low nNO and prompt further diagnostic testing, thus increasing costs both to the healthcare system and the patient. Only seven studies were eligible for inclusion in the meta-analysis of non-VC maneuver, as opposed to 12 studies for the VC maneuver, and the low 95 % CI limit of the former may be due to the limited sample size. Furthermore, two of the non-VC studies used the NIOX MINO portable device which uses electrochemical analysis of NO as opposed to the better validated chemiluminescence method of the stationary devices (NIOX FLEX, Ecomedics CLD88). Nevertheless, we performed a post hoc analysis with the exclusion of these studies which did not influence the diagnostic accuracy of the non-VC maneuver and the quantitative synthesis includes data from all seven non-VC studies.

NIOX MINO is a simpler and cheaper tool for measuring nNO, and validation studies have already been published [14, 36, 37]. However, as NIOX MINO was designed for exhaled NO measurement in asthmatic individuals, issues relating to its accuracy [37] and repeatability [14] have been reported when used for nNO measurement in subjects referred for PCD evaluation. In addition, while the manufacturer recommends measurements of nNO with BH for at least 45 s, this is usually not possible by many patients and instead NIOX MINO is frequently used with the alternative TB maneuver regardless of patient’s age [14, 37]. These limitations question the suitability of NIOX MINO as a stand-alone diagnostic test and additional studies on the diagnostic accuracy of NIOX MINO measurements during TB are needed to confirm the validity of this method. However, the low cost and simple use potentiate the consideration of NIOX MINO as a promising first line screening test in a future diagnostic algorithm for PCD.

Currently, there is no universally accepted cutoff for abnormally low nNO. The included studies in this meta-analysis proposed a variety of cutoffs for nNO production by VC (7.2–126 nl/min) and non-VC (17.4–60 nl/min) techniques. This variability demonstrates the need for standardization of nNO measurements and agreement on cutoffs for the different breathing maneuvers. A recent, large, multicenter study has proposed a cutoff equal to 77 nl/min for VC [7], whereas the meta-analysis by Collins et al. reported that a cutoff of 75.2 nl/min would include 99.85 % of PCD patients performing VC maneuvers [15]. Regarding the non-VC maneuver however, no cutoff value has been proposed by a large enough study, thus additional studies are needed for the establishment of such cutoffs and further standardization of the technique.

Our study has some limitations. The main limitation is the heterogeneity and the weaknesses of the diagnostic standard that the published studies are employing for the definition of PCD status. As a result, the captured spectrum of the disease might not be totally representative of the true PCD population. TEM, which is the most commonly used test for PCD status definition in published studies, misses approximately 30 % of patients with PCD [38] and this should be taken into account in the assessment of the diagnostic efficacy of nNO. However, in the sensitivity analysis, when we included the studies that had employed more than one (in addition to TEM) diagnostic test to establish PCD diagnosis, our results did not change substantially thus providing relative certainty to the accuracy of nNO as a diagnostic test. There is considerable variation between individual studies in the number of cases, total sample size and cut-off values. However, the bivariate meta-analysis and HSROC curve analyses take explicitly this diversity into account and can accommodate studies with populations of different risks and different definition thresholds. Another issue for the synthesis of the data is that the majority of the included studies were diagnostic case–control studies. Empirical evidence has shown that case–control studies, as opposed to cohort studies, may overestimate the diagnostics Odds Ratio (DOR) [39]. Nevertheless, we think that the possibility of overestimation is limited, as the case–control studies included here were diagnostic studies designed to assess the test accuracy and not to provide evidence on associations between a risk factor and the disease [40]. Additionally, given the rarity of PCD, it is expected that the majority of studies will have a case–control design. Due to the same reason, both case–control and prospective cohort studies included relatively small numbers of subjects. However the synthesis of the included studies led to the inclusion of data for several hundreds of PCD and non-PCD subjects and allowed the use of the appropriate statistical models and provided the estimates we report. It should be underlined of course that these estimates apply provided that the ATS/ERS guidelines are followed for the performance of the test and the obtained values are compared to the normal values obtained from samples of healthy subjects in the respective populations.

Conclusions

In summary, measurement of nNO, both with VC and non-VC maneuvers, has high overall diagnostic accuracy and provides a clinically significant diagnostic tool for large uninvestigated populations of suspect cases worldwide where access to TEM and HSVM is not easy. Furthermore, the high overall diagnostic accuracy of nNO calls for re-evaluation of the diagnostic accuracy of each of the available diagnostic tests for PCD (nNO, TEM and HSVM) with the aim to develop an algorithm with the most efficacious combination of tests to achieve PCD diagnosis.

Abbreviations

ATS/ERS:

American Thoracic Society/European Respiratory Society

BH:

Breath hold

ER:

Exhalation against resistance

FN:

False negatives

FP:

False positives

HSVM:

High Speed Video Microscopy

nNO:

nasal Nitric Oxide

PCD:

Primary Ciliary Dyskinesia

ppb:

parts per billion

TB:

Tidal breathing

TEM:

Transmission Electron Microscopy

TP:

True positives

TN:

True negatives

VC:

Velum closure

References

  1. Leigh MW, Pittman JE, Carson JL, Ferkol TW, Dell SD, Davis SD, et al. Clinical and genetic aspects of primary ciliary dyskinesia/Kartagener syndrome. Genet Med. 2009;11(7):473–87.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Noone PG, Leigh MW, Sannuti A, Minnix SL, Carson JL, Hazucha M, et al. Primary ciliary dyskinesia: diagnostic and phenotypic features. Am J Respir Crit Care Med. 2004;169(4):459–67.

    Article  PubMed  Google Scholar 

  3. Boon M, Jorissen M, Proesmans M, De Boeck K. Primary ciliary dyskinesia, an orphan disease. Eur J Pediatr. 2013;172(2):151–62.

    Article  PubMed  Google Scholar 

  4. Kim RH, A Hall D, Cutz E, Knowles MR, Nelligan KA, Nykamp K, et al. The Role of Molecular Genetic Analysis in the Diagnosis of Primary Ciliary Dyskinesia. Ann Am Thorac Soc. 2014;11(3):351–9.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Stannard WA, Chilvers MA, Rutman AR, Williams CD, O’Callaghan C. Diagnostic testing of patients suspected of primary ciliary dyskinesia. Am J Respir Crit Care Med. 2010;181(4):307–14.

    Article  PubMed  Google Scholar 

  6. Papon JF, Coste A, Roudot-Thoraval F, Boucherat M, Roger G, Tamalet A, et al. A 20-year experience of electron microscopy in the diagnosis of primary ciliary dyskinesia. Eur Respir J. 2010;35(5):1057–63.

    Article  CAS  PubMed  Google Scholar 

  7. Leigh MW, Hazucha MJ, Chawla KK, Baker BR, Shapiro AJ, Brown DE, et al. Standardizing nasal nitric oxide measurement as a test for primary ciliary dyskinesia. Ann Am Thorac Soc. 2013;10(6):574–81.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Schwabe GC, Hoffmann K, Loges NT, Birker D, Rossier C, De Santi MM, et al. Primary ciliary dyskinesia associated with normal axoneme ultrastructure is caused by DNAH11 mutations. Hum Mutat. 2008;29(2):289–98.

    Article  CAS  PubMed  Google Scholar 

  9. Lucas JS, Adam EC, Goggin PM, Jackson CL, Powles‐Glover N, Patel SH, et al. Static respiratory cilia associated with mutations in Dnahc11/DNAH11: a mouse model of PCD. Hum Mutat. 2012;33(3):495–503.

    Article  CAS  PubMed  Google Scholar 

  10. Raidt J, Wallmeier J, Hjeij R, Onnebrink JG, Pennekamp P, Loges NT, et al. Ciliary beat pattern and frequency in genetic variants of primary ciliary dyskinesia. Eur Respir J. 2014

  11. Lundberg J, Farkas-Szallasi T, Weitzberg E, Rinder J, Lidholm J, Änggåard A, et al. High nitric oxide production in human paranasal sinuses. Nat Med. 1995;1(4):370–3.

    Article  CAS  PubMed  Google Scholar 

  12. Strippoli MP, Frischer T, Barbato A, Snijders D, Maurer E, Lucas JS, et al. ERS Task Force onPrimary Ciliary Dyskinesia in Children: Management of primary ciliary dyskinesia in European children: recommendations and clinical practice. Eur Respir J. 2012;39(6):1482–91.

    Article  PubMed  Google Scholar 

  13. American Thoracic Society, European Respiratory Society. ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005. Am J Respir Crit Care Med. 2005;171(8):912–30.

    Article  Google Scholar 

  14. Harris A, Bhullar E, Gove K, Joslin R, Pelling J, Evans HJ, et al. Validation of a portable nitric oxide analyzer for screening in primary ciliary dyskinesias. BMC Pulm Med. 2014;14(1):18.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Collins SA, Gove K, Walker W, Lucas JS. Nasal nitric oxide screening for primary ciliary dyskinesia: systematic review and meta-analysis. Eur Respir J. 2014;44(6):1589–99.

    Article  PubMed  Google Scholar 

  16. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529–36.

    Article  PubMed  Google Scholar 

  17. Chu H, Cole SR. Bivariate meta-analysis of sensitivity and specificity with sparse data: a generalized linear mixed model approach. J Clin Epidemiol. 2006;59(12):1331–2.

    Article  PubMed  Google Scholar 

  18. Harbord R. METANDI: Stata module to perform meta-analysis of diagnostic accuracy. In: Statistical Software Components S456932, Boston College Department of Economics, 2008. https://ideas.repec.org/c/boc/bocode/s456932.html. Assessed 15 May 2015.

  19. Deeks JJ, Altman DG. Diagnostic tests 4: likelihood ratios. BMJ. 2004;329(7458):168–9.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Pifferi M, Bush A, Maggi F, Michelucci A, Ricci V, Conidi ME, et al. Nasal nitric oxide and nitric oxide synthase expression in primary ciliary dyskinesia. Eur Respir J. 2011;37(3):572–7.

    Article  CAS  PubMed  Google Scholar 

  21. Morgan L, Buddle L, Rogers P, Seccombe L, MacKenney K, Hughes L. Nasal nitric oxide is useful in an Australian Primary Ciliary Dyskinesia clinic. Cilia. 2012;1 Suppl 1:7.

    Google Scholar 

  22. Marthin JK, Nielsen KG. Choice of nasal nitric oxide technique as first-line test for primary ciliary dyskinesia. Eur Respir J. 2011;37(3):559–65.

    Article  CAS  PubMed  Google Scholar 

  23. Corbelli R, Bringolf-Isler B, Amacher A, Sasse B, Spycher M, Hammer J. Nasal nitric oxide measurements to screen children for primary ciliary dyskinesia. CHEST J. 2004;126(4):1054–9.

    Article  CAS  Google Scholar 

  24. Beydon N, Chambellan A, Alberti C, de Blic J, Clément A, Escudier E, et al. Technical and practical issues for tidal breathing measurements of nasal nitric oxide in children. Pediatr Pulmonol. 2015; :n/a-n/a.

  25. Piacentini GL, Bodini A, Peroni D, Rigotti E, Pigozzi R, Pradal U, et al. Nasal nitric oxide for early diagnosis of primary ciliary dyskinesia: practical issues in children. Respir Med. 2008;102(4):541–7.

    Article  PubMed  Google Scholar 

  26. Narang I, Ersu R, Wilson NM, Bush A. Nitric oxide in chronic airway inflammation in children: diagnostic use and pathophysiological significance. Thorax. 2002;57(7):586–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Montella S, Alving K, De Stefano S, Di Micco LL, Di Giorgio A, Santamaria F. Nasal nitric oxide measurement using continuous aspiration by hand-held device discriminates patients with primary ciliary dyskinesia from healthy subjects. Eur Respir J. 2012;40 Suppl 56:3338.

    Google Scholar 

  28. Santamaria F, De Stefano S, Montella S, Barbarano F, Iacotucci P, Ciccarelli R, et al. Nasal nitric oxide assessment in primary ciliary dyskinesia using aspiration, exhalation, and humming. Med Sci Monit. 2008;14(2):CR80–5.

    CAS  PubMed  Google Scholar 

  29. Moreno Galdo A, Vizmanos Lamotte G, Reverte Bover C, Gartner S, Cobos Barroso N, Rovira Amigo S, et al. Value of nasal nitric oxide in the diagnosis of primary ciliary dyskinesia. An Pediatr (Barc). 2010;73(2):88–93.

    Article  CAS  Google Scholar 

  30. Boon M, Meyts I, Proesmans M, Vermeulen FL, Jorissen M, De Boeck K. Diagnostic accuracy of nitric oxide measurements to detect primary ciliary dyskinesia. Eur J Clin Invest. 2014;44(5):477–85.

    Article  CAS  PubMed  Google Scholar 

  31. Mateos-Corral D, Coombs R, Grasemann H, Ratjen F, Dell SD. Diagnostic value of nasal nitric oxide measured with non-velum closure techniques for children with primary ciliary dyskinesia. J Pediatr. 2011;159(3):420–4.

    Article  CAS  PubMed  Google Scholar 

  32. Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Ann Intern Med. 2009;151(4):264–9.

    Article  PubMed  Google Scholar 

  33. Barbato A, Frischer T, Kuehni CE, Snijders D, Azevedo I, Baktai G, et al. Primary ciliary dyskinesia: a consensus statement on diagnostic and treatment approaches in children. Eur Respir J. 2009;34(6):1264–76.

    Article  CAS  PubMed  Google Scholar 

  34. Yiallouros PK, Kouis P, Middleton N, Nearchou M, Adamidi T, Georgiou A, et al. Clinical features of primary ciliary dyskinesia in Cyprus with emphasis on lobectomized patients. Respir Med. 2015;109(3):347–56.

    Article  PubMed  Google Scholar 

  35. Coren M, Meeks M, Morrison I, Buchdahl R, Bush A. Primary ciliary dyskinesia: age at diagnosis and symptom history. Acta Paediatr. 2002;91(6):667–9.

    Article  CAS  PubMed  Google Scholar 

  36. Maniscalco M, Laurentiis G, Weitzberg E, Lundberg J, Sofia M. Validation study of nasal nitric oxide measurements using a hand‐held electrochemical analyser. Eur J Clin Invest. 2008;38(3):197–200.

    Article  CAS  PubMed  Google Scholar 

  37. Marthin JK, Nielsen KG. Hand-held tidal breathing nasal nitric oxide measurement–a promising targeted case-finding tool for the diagnosis of primary ciliary dyskinesia. PLoS One. 2013;8(2):e57262.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  38. Knowles MR, Leigh MW, Carson JL, Davis SD, Dell SD, Ferkol TW, et al. Genetic Disorders of Mucociliary Clearance Consortium: Mutations of DNAH11 in patients with primary ciliary dyskinesia with normal ciliary ultrastructure. Thorax. 2012;67(5):433–41.

    Article  PubMed  Google Scholar 

  39. Lijmer JG, Mol BW, Heisterkamp S, Bonsel GJ, Prins MH, van der Meulen JH, et al. Empirical evidence of design-related bias in studies of diagnostic tests. JAMA. 1999;282(11):1061–6.

    Article  CAS  PubMed  Google Scholar 

  40. Rutjes AW, Reitsma JB, Vandenbroucke JP, Glas AS, Bossuyt PM. Case–control and two-gate designs in diagnostic accuracy studies. Clin Chem. 2005;51(8):1335–41.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

This work was performed at the Cyprus International Institute for Environmental and Public Health in association with Harvard School of Public Health, Cyprus University of Technology, Limassol, Cyprus.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Panayiotis Kouis.

Additional information

Competing interests

The authors wish to declare that they have no competing interests.

Authors’ contributions

PK performed the search and extraction of information from the papers, organized the material and prepared the first draft of the manuscript. SIP provided advice on the methodology of the search and selection of the papers, contributed to the interpretation of the findings, critically revised the manuscript and contributed towards the final version of the manuscript. PKY conceived the hypothesis of the manuscript, coordinated the whole work, contributed to the interpretation of the findings and contributed intellectually towards the final version of the manuscript. All authors read and approved the final manuscript.

Authors’ information

S.I.P. is the main advisor of P.K for his doctoral studies. PK was supported by the European Union’s Seventh Framework Program EC-GA No. 305404 BESTCILIA. The sponsors had no role or involvement in study design; in the collection, analysis and interpretation of data; in the writing of the report; and in the decision to submit the article for publication.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kouis, P., Papatheodorou, S.I. & Yiallouros, P.K. Diagnostic accuracy of nasal nitric oxide for establishing diagnosis of primary ciliary dyskinesia: a meta-analysis. BMC Pulm Med 15, 153 (2015). https://doi.org/10.1186/s12890-015-0147-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-015-0147-3

Keywords