Skip to main content
  • Research article
  • Open access
  • Published:

Biological quality control for cardiopulmonary exercise testing in multicenter clinical trials

Abstract

Background

Precision and accuracy assurance in cardiopulmonary exercise testing (CPET) facilitates multicenter clinical trials by maximizing statistical power and minimizing participant risk. Current guidelines recommend quality control that is largely based on precision at individual testing centers (minimizing test–retest variability). The aim of this study was to establish a multicenter biological quality control (BioQC) method that considers both precision and accuracy in CPET.

Methods

BioQC testing was 6-min treadmill walking at 20 W and 70 W (below the lactate threshold) with healthy non-smoking laboratory staff (15 centers; ~16 months). Measurements were made twice within the initial 4 weeks and quarterly thereafter. Quality control was based on: 1) within-center precision (coefficient of variation [CV] for oxygen uptake [V̇O2], carbon dioxide output [V̇CO2], and minute ventilation [V̇E] within ±10 %); and 2) a criterion that V̇O2 at 20 W and 70 W, and ∆V̇O2/∆WR were each within ±10 % predicted. “Failed” BioQC tests (i.e., those outside the predetermined criterion) prompted troubleshooting and repeated measurements. An additional retrospective analysis, using a composite z-score combining both BioQC precision and accuracy of V̇O2 at 70 W and ∆V̇O2/∆WR, was compared with the other methods.

Results

Of 129 tests (5 to 8 per center), 98 (76 %) were accepted by within-center precision alone. Within-center CV was <9 %, but between-center CV remained high (9.6 to 12.5 %). Only 43 (33 %) tests had all V̇O2 measurements within the ±10 % predicted criterion. However, a composite z-score of 0.67 identified 67 (52 %) non-normal outlying tests, exclusion of which coincided with the minimum CV for CPET variables.

Conclusions

Study-wide BioQC using a composite z-score can increase study-wide precision and accuracy, and optimize the design and conduct of multicenter clinical trials involving CPET.

Trial registration

ClinicalTrials.gov identifier: NCT01072396; February 19, 2010.

Peer Review reports

Background

Cardiopulmonary exercise testing (CPET) is a non-invasive, sensitive test to evaluate cardiopulmonary (patho-) physiology. CPET assesses the physiological basis of functional capacity and exercise intolerance, and plays a valuable role in diagnosis and clinical decision-making [1]. CPET is also used to test intervention efficacy, e.g., exercise training in cardiovascular disease [2, 3], pulmonary rehabilitation [48], and bronchodilator therapy in multicenter trials [911]. These applications require strong agreement between consecutively performed tests within and among investigative centers. However, precision (reproducibility, or test–retest variability) and accuracy (trueness) of CPET depend on interactions among: testing equipment variability, calibration, and maintenance; physiological factors; participants’ cooperation, motivation, and effort during testing; and knowledge, skills, and training of laboratory personnel [12]. Most factors cannot be controlled by a simple system calibration, emphasizing the importance of standardization and quality control (QC). Clinicians can rely on trial results only if interpretation is not biased by measurement error [12, 13]. Therefore, assurance of study-wide precision and accuracy has a major impact on the design and conduct of multicenter trials, by maximizing statistical discriminatory power, and minimizing laboratory burden and participant risk.

The American Thoracic Society (ATS)/American College of Chest Physicians (ACCP) [4], American Heart Association [3], and European Respiratory Society (ERS) [7, 12] have published CPET recommendations and standards. These give “best practice” for calibration and QC, and provide typical coefficients of variation (CV) for physiological measurements. ATS/ACCP [4] recommends a biological QC (BioQC) procedure [14] whereby a healthy subject on a stable diet performs regular exercise tests at work rates (WR) below the lactate threshold [12]. As physiologic responses are typically highly reproducible [3, 7, 1417], the use of healthy individuals performing BioQC can assure reproducibility of the integrated CPET measurement system for patient testing within and between testing centers [18]. Jones and Kane [19] used cycle ergometers and subjects who travelled to each participating center for testing, to demonstrate the efficacy of BioQC to minimize the study-wide CV for oxygen uptake (V̇O2; 5.6 %), carbon dioxide output (V̇CO2; 8.4 %), and minute ventilation (V̇E; 8.2 %) by excluding the least precise results. Similarly, while Gagnon et al. [20] showed good agreement among five testing centers for V̇O2,peak, this approach did not determine the precision and accuracy of measured exercise responses to a standardized protocol. Importantly, the intraclass correlation coefficient at low WR was <0.7 [20], emphasizing the need for QC of the relationship between mechanical and metabolic power output. Brawner et al. [21] used a standardized treadmill protocol, where quality assurance was accepted if at least two of three exercise stages fell within a “target range” for V̇O2. Rather than assuring the precision and accuracy of the individual mechanical–metabolic power coupling, the basis of the acceptability criteria (the “target range”) in Brawner et al. [21] was wide, in part due to differences in weights of the volunteers used to develop the normative data.

In this study, we sought to establish new CPET BioQC acceptability criteria that considered both precision and accuracy of the mechanical–metabolic power coupling during a standardized treadmill protocol, and which would be suitable for quality assurance in multicenter studies.

We report the outcome of a precision-based approach to CPET QC in a multicenter trial (ClinicalTrials.gov identifier: NCT01072396) and propose new QC procedures based on both precision and accuracy to minimize variability in multicenter trials. The patient characterization and treatment phases of the parent trial have been reported [22, 23]. Some of the results of this QC study have been previously reported in the form of an abstract [24].

Methods

Individuals involved in BioQC procedures

The primary independent Institutional Review Board (IRB), Chesapeake Research Review, Inc. (Columbia, MD, USA) approved the host trial protocol (NCT01072396) for five sites, which could utilize the study’s central IRB, while the remaining 10 participating centers obtained individual IRB/Independent Ethics Committee (IEC) approval (for details see Additional file 1: Table S1). The approved protocol included detailed exercise BioQC procedures in an Exercise and Quality Control Procedure Manual in accordance with ATS/ACCP recommendations [4]. During the trial, de-identified BioQC data from a non-smoking, healthy member of the laboratory staff at each center were submitted to a central reader. These laboratory staff members were required to fast and not consume caffeinated drinks for at least 2 h prior to testing, and their age, height, and weight (dressed, wearing shoes) were recorded. This manuscript describes an exempt retrospective study of this de-identified physiologic BioQC data from NCT01072396.

Equipment used at the study centers

Details of the CPET equipment, software versions, flow sensors, and treadmill details used in the study are shown in Additional file 2: Table S2. The treadmills were run from exercise software, with the exception of one center, where the treadmill was manually adjusted from its own controller using a preapproved procedure.

Manual of procedures

All centers selected to take part in the trial were provided with an Exercise and Quality Control Procedure Manual, which detailed a standardized approach to exercise testing, calibration, and QC. The main objectives of the manual were to: (1) provide information about the available guidelines to promote QC; (2) standardize technical procedures in CPET in order to minimize variation within and between participating centers; and (3) outline and standardize specific procedures involved in the clinical study.

Staff training, and equipment calibration and verification

Prior to enrolment of patients in the clinical trial, each participating center was visited by a consultant (TechEd Consultants, Inc., Mason, MI, USA) to evaluate and verify the equipment acceptability, and to standardize all QC and test procedures by providing specific training for staff. Training included an initial submaximal incremental exercise test and a BioQC constant work rate test. Details on staff training, and equipment calibration and verification are provided in Additional file 3.

A participating laboratory was only released for patient testing when all CPET technical and equipment performance qualifying criteria were met.

BioQC procedures

CPET systems were calibrated according to manufacturer’s instructions immediately prior to BioQC testing (see Additional file 3). All centers measured V̇O2, V̇CO2, and V̇E breath-by-breath via a mouthpiece, and heart rate from the electrocardiogram. The BioQC was a two-stage constant work rate treadmill exercise test. The 18-min protocol consisted of: 3 min of standing rest; 3 min of slow walking (0.8 mph); and 6 min each at 20 W (1.0 mph) and 70 W (1.8 mph). Treadmill grades for each WR were calculated based on participant’s clothed weight [25].

BioQC procedures were carried out at the on-site training visit, within 4 weeks post-training to verify validity of results, and quarterly thereafter for the duration of the study. Systems that required servicing (analyzer replacement, software updates, etc.) underwent an out-of-schedule BioQC test and patient testing was resumed only when the system passed the QC criteria. A maintenance and troubleshooting log was used to record all preventative maintenance activities, as well as faults and repairs (see Additional file 4).

Accuracy and precision of gas exchange and ventilation measurements

The BioQC tests were analyzed at a central reading site. Physiologic responses were submitted as 10-s bin averages. Steady-state V̇O2, V̇CO2, and V̇E values were calculated by averaging the last 3 min of exercise at 20 W and 70 W. Acceptability of each BioQC test was established by three methods: the central reader method, the z-score method, and the criterion method.

Central reader method

During the multicenter trial, BioQC acceptability was based upon within-center precision. A single central reader compared a BioQC result with the initiation and accumulated quarterly tests within each center. The BioQC was accepted if V̇O2, V̇CO2, and V̇E were within ±10 % of the initial value based on the expected normal variance [2629] and ATS/ACCP requirements [4]. This method established precision over time (after an initial accuracy check at the staff-training visit) [4]. A measurement not meeting these criteria was repeated after troubleshooting (see Additional file 4).

Z-score method

After trial completion, all BioQC results were assessed en bloc to establish generalizable acceptability criteria. Accuracy and precision of BioQC data were based on “position” and “slope” of the linear relationship between V̇O2 and WR [30], as predicted by the following [27]:

$$ \overset{.}{\mathrm{V}}{\mathrm{O}}_{\mathtt{2}}=\left(\mathtt{5.8}\times \mathrm{weight}\left[\mathrm{kg}\right]\right)+\mathtt{151}+\left(\mathtt{10.1}\times \mathrm{W}\right) $$
(1)

Position was established from % predicted V̇O2 at 70 W (V̇O2,70W) and slope from % predicted “functional gain” between 20 W and 70 W (V̇O2,slope) based on a ΔV̇O2/ΔWR of 10.1 mL/min/W [26, 27], which has an established normal range of approximately ±10 % [28, 29]. Thus, accuracy was established by deviation from the predicted value, and precision from the standard deviation (SD) of all BioQC tests. A composite z-score in which position and slope were equally weighted was calculated:

$$ \begin{array}{l}\mathrm{z} = \left(\mathrm{ABS}\left(\%\ \mathrm{predicted}\ \overset{.}{\mathrm{V}}{\mathrm{O}}_{2,70\mathrm{W}}-100\right)/\mathrm{S}\mathrm{D}\ \%\ \mathrm{predicted}\ \overset{.}{\mathrm{V}}{\mathrm{O}}_{2,70\mathrm{W}}\right)/2+\\ {}\left(\mathrm{ABS}\left(\%\ \mathrm{predicted}\ {\mathrm{V}\mathrm{O}}_{2,\mathrm{slope}}-100\right)/\mathrm{S}\mathrm{D}\ \%\ \mathrm{predicted}\ {\mathrm{V}\mathrm{O}}_{2,\mathrm{slope}}\right)/2\end{array} $$
(2)

where SD is the standard deviation of % predicted for each variable. The Shapiro–Wilk test on the population of % predicted V̇O2,70W and V̇O2,slope measurements was used to identify z-score values associated with a systematic departure from normality (W statistic P ≤ 0.05). As the composite z-score criterion is based on % predicted values of all available tests across all centers, it was used to establish the minimum acceptability of both the precision and accuracy of CPET measurements.

Criterion method

Both acceptability methods (reader and z-score) were compared with the application of a rigid criterion that no single V̇O2 measurement (V̇O2,20W, V̇O2,70W, and V̇O2,slope) should deviate by more than ±10 % from predicted. If any one measurement’s deviation exceeded ±10 % predicted, the test was deemed to have failed QC.

Statistical analyses

Mean and SD were used to calculate CV for variables within and between centers. Departure from normality was assessed by the Shapiro–Wilk test. For multiple comparisons in one-way analysis of variance (ANOVA), the Student–Newman–Keuls test was used. Statistical significance was accepted if P ≤ 0.05. All calculations and statistical analyses were performed using MS Excel (Redmond, WA, USA) and SigmaPlot 12 (Systat, San Jose, CA, USA).

Results

BioQC laboratory staff were males (n = 8; age 41.4 ± 11.4 years, weight 86.6 ± 17.3 kg, body mass index [BMI] 28.0 ± 4.3 kg/m2) and females (n = 7; age 41.4 ± 8.4 years, weight 61.2 ± 10.8 kg, BMI 23.7 ± 4.9 kg/m2). The number of BioQC tests per center varied from five to eight, depending on the length of time the center was active in the trial (16.3 ± 3.1 months [range 10 to 20 months]). This resulted in a total of 129 BioQC tests performed (6.5 ± 1.2 per center), with each center having between 10 and 28 non-repeated pairs of BioQC tests. Overall, therefore, there were 523 unique paired differences of BioQC tests, which were used to establish the distribution properties of the measurements.

Within-center variability – central reader method

The central reader method accepted 98 (76 %) BioQC tests; 21 (24 %) initial tests required troubleshooting and repetition to meet QC acceptability criteria. At three centers, all tests were accepted. Four centers required one repeat test, two required two repeats, and five required three or more repeats to bring measurements within acceptable CV limits. Thus, across all centers, 31 repeated tests were necessary to bring BioQC measurements within acceptable CV limits (for a study-wide total of 129 tests). Further details on troubleshooting of the CPET systems during the study are provided in Additional file 4.

The within-center mean and variability of accepted BioQC tests are given in Table 1. The most precise variables were V̇O2 at 70 W (CV = 5.8 %) and ΔV̇O2/ΔWR, which averaged 10.6 ± 0.8 mL/min/W (CV = 5.8 %). The least precise variable was V̇CO2 at 20 W (CV = 9.2 %). The assumption that V̇O2 at 70 W was below each individual’s lactate threshold was supported by attainment of a steady-state within 6 min.

Table 1 Within-center variability of gas exchange and ventilation measurement during treadmill exercise. Measurements were made at work rates of 20 W and 70 W in 98 reader-accepted biological quality control tests

Multicenter precision and accuracy

Z-score method

Study wide “position” accuracy of CPET measurements in the z-score method was assessed using V̇O2,70W because, of the two steady-state V̇O2 measurements, V̇O2,70W showed less variability. Systematic inclusion of non-normal (outlying) BioQC measurements at V̇O2,70W occurred above a z-score of 0.67, while non-normal measurements of ΔV̇O2/∆WR occurred above a z-score of 0.75 (Fig. 1). Thus, using a critical z-score of 0.67, 62 (48 %) tests were deemed acceptable. At z = 0.67, multicenter CV for V̇O2,20W, V̇O 2,70W, and ∆V̇O2/∆WR were 6.2 %, 4.7 %, and 6.0 %, respectively (Table 2). Accuracy was not different from the criterion method (see below), and close to 100 % predicted in all cases (Table 2). The z-score was the only method for which final selected data for all V̇O2 variables were normally distributed (Table 2). Composite z = 0.67 coincided with the lowest CV for absolute measurements in all three variables (Fig. 2a). Systematic increases in V̇O2 measurement CV were observed at a z-score of ~0.9 (Fig. 2a) and CV exceeded ATS/ACCP guideline recommendations at a z-score of ~1.0 (Fig. 2b). The CV of % predicted measurements increased with increasing z-score, and z = 0.67 corresponded to a CV in all three variables less than ~6 % (Fig. 2b) incorporating ~50 % of tests (Fig. 2c). Using the composite z-score, a greater number of tests were excluded than using the central reader method (Table 2). Additionally, there was only 53 % agreement in test acceptability between the central reader and composite z-score methods. Overall, precision and accuracy for V̇O2 measurements using composite z ≤ 0.67 was greater than the central reader method and very similar to the criterion method, despite substantially more tests than the latter being deemed within acceptable limits.

Fig. 1
figure 1

Effect of z-score on normality of oxygen uptake (V̇O2) measurement distribution. Normalcy of biological quality control test measurements selected on the basis of a composite z-score. Above a critical z-score of 0.67, the W statistic drops dramatically and the distribution of paired differences becomes non-normal for V̇O2 at 70 W

Table 2 Characteristics of variability of oxygen uptake (V̇O2) measurements during 129 biological quality control tests, using three different quality control methods
Fig. 2
figure 2

Effect of z = 0.67 cut-off on the CV and number of accepted CPETs. a Coefficient of variation (CV) of the absolute oxygen uptake (V̇O2) at 20 W, 70 W, and the increase in V̇O2 per W (∆V̇O2/∆WR), at different z-scores. b CV of % predicted V̇O2 at 20 W, 70 W, and ∆V̇O2/∆WR, at different z-scores. c % of acceptable tests (n = 129), at different z-scores. The shaded area is the approximate range of z-scores (0.67 to 0.9) over which absolute measurement CV was minimized (based on panel a, and transposed into panels b and c). The critical z-score is the minimum value for which all measurements are normally distributed. It is noted that the absolute CV (panel a) depends on both variability in measurements and differences in weight of individuals performing the biological quality control tests (absolute V̇O2 at 20 W and 70 W treadmill walking is dependent on weight). Weight of individuals did not significantly vary during the trial. Despite this, the absolute CV is useful to isolate the z-score range at which the minimum CV occurred (shaded bar). The variability due to measurement differences among centers is better assessed using % predicted values (panel b)

Criterion method

When a criterion of ±10 % variability in all V̇O2 measurements (% predicted V̇O2,20W, V̇O2,70W, and V̇O2,slope) was applied, only 43 (33 %) tests were acceptable. The criterion method resulted in greater precision (lower CV) and accuracy compared with the central reader method, but was not normally distributed in all variables (Table 2).

Optimizing precision using distribution analysis of paired differences

Analysis of 523 paired differences for V̇O2, V̇CO2, and V̇E is shown in Fig. 3. By including all tests, none of the variables were well described by a Gaussian distribution (Fig. 3, open circles). Whereas, for tests with composite z ≤ 0.67, all gas exchange and ventilation measurements (except V̇E at rest) were distributed normally (Fig. 3, closed circles). A composite z-score based on V̇O2 measurement alone was effective in also excluding tests with outlying, non-normal measurements in V̇CO2 and V̇E. In all CPET variables, a QC method based on V̇O2 precision and accuracy among all centers reduced data variability by ~60 % compared with using no QC method, and by ~50 % compared with the central reader method.

Fig. 3
figure 3

Distribution of relative frequencies of paired differences from zero. Differences from zero for oxygen uptake (V̇O2), carbon dioxide output (V̇CO2), and minute ventilation (V̇E) at rest (ac), 20 W (df), and 70 W (gi). None of the variables show normal distribution when all tests are considered (Shapiro–Wilk test, P < 0.001). Test selection based on z ≤ 0.67 resulted in all data at each level of exercise (with the exception of V̇E at rest) being normally distributed (Shapiro–Wilk test, P > 0.05)

Discussion

To our knowledge, this is the first study to establish study-wide precision and accuracy for CPET measurements in multicenter trials, based on normalcy of the coupling between mechanical and metabolic power output. We found: 1) the CV of CPET measurements was reduced using the central reader assessment of BioQC exercise tests to minimize within-center variability; 2) a rigid application of a ±10 % predicted V̇O2 cut-off criterion excluded 67 % of all measurements, some of which were within the normal distribution of accurate tests; and 3) the application of a composite z-score to identify measurements lying outside normal limits increased the precision and accuracy of multicenter trial CPET measurements by ~50 %.

Efficacy of CPET QC based on within-center precision

Acceptance rate by the central reader method for all BioQC tests was 76 %. The BioQC process identified measurement errors in 12 of 15 centers before measurements of study patients were made in the parent clinical trial. In six centers, measurement error was resolved by standard troubleshooting approaches, and required only one or two additional BioQC tests to demonstrate resolution. The central reader method excluded outlying tests based on reproducibility, and effectively excluded tests departing from normality. Analysis of all tests showed non-normal distribution, whereas including only tests accepted by the central reader assessment resulted in a data set that was not significantly different from a normal distribution (Table 2). The central reading process reduced within-center CV for gas exchange and ventilation measurements to within the range 5.8 to 9.2 % (Table 1), which is within the range generally accepted for CPET studies [3, 4, 12, 31, 32].

Application of a central BioQC reader, therefore, provided a beneficial addition that reduced measurement variability in the multicenter trial. However, while within-center variability was below the upper limits recommended by ATS/ACCP [4] (~10 %), there remained variability between different centers. The between-center CV for gas exchange measurements of all centrally accepted BioQC tests ranged from 9.6 to 12.5 % (Table 2). This residual between-center variability effectively lessened the benefit of the BioQC method, and weakened the statistical power to demonstrate a given change in CPET outcome measures in the parent clinical trial [23].

CPET QC based on study-wide precision and accuracy

To our knowledge, only one previous study attempted a QC regimen based on both precision and accuracy [21]. However, the accuracy criterion used was wide (±25 %) because the approach did not account for differences in mechanical power output during treadmill walking at given speed/grade combinations among individuals differing in weight. Therefore, to develop an approach to CPET QC in multicenter trials that included study-wide precision and accuracy in the coupling of mechanical to metabolic power, we applied two different BioQC methods. The first, a simple criterion approach, rigidly excluded BioQC tests in which accuracy of any one V̇O2 measurement (V̇O2,20W, V̇O2,70W, and V̇O2,slope) lay outside 90 to 110 % of the predicted value at the calculated power output. This approach appears inherently sensible, in that measurements outside this ±10 % range (roughly based on guideline recommendations [4, 31, 32]) are considered as outlying, and thus excluded. However, a limitation is that a relatively small error in only one variable causes a failing BioQC test, even if all other measurements are within the tolerable limit. The very low acceptability rate of tests across the 15 centers (33 %) makes application of this criterion impractical. Indeed, seven study centers would have been completely excluded from the trial, despite reporting demonstrably accurate measurements based on the retrospective analysis of the normal distribution of the BioQC measurements (Fig. 3). Thus, while the rigid criterion dramatically improved the CV of CPET measurements, it excluded data that were within the normal distribution of measurement variability (e.g., Fig. 1, Table 2).

Therefore, we developed a method that allowed small variability outside the 90 to 110 % range for predicted V̇O2, but could successfully identify outlying, non-normal measurements and reduce between-center measurement CV. The composite z-score considered equally, the relative deviation from the mean of “position” (V̇O2,70W) and “slope” (ΔV̇O2/ΔWR) of the highly predictable relationship between V̇O2 and WR. By combining error distribution of these two variables, a small deviation from predicted (i.e., between ~87 % and ~113 %) in one measurement was allowed as long as the other was accurate. We found this method able to strongly predict systematic deviations in non-normal measurements above a composite z-score of 0.67 (based on SD of all tests; Fig. 1). In addition, composite z-score of 0.67 coincided with local minima in CV of absolute V̇O2 measurements (Fig. 2a) and resulted in a study-wide CV of ~6 % (Fig. 2b).

Using z = 0.67 we were able to identify that 52 % of the BioQC tests lay outside the normal distribution of V̇O2 measurements. While this approach excluded more tests than the central reading method, this combined precision- and accuracy-based method achieved three main benefits. Firstly, it had a strong agreement with the rigid criterion-based approach (84 % agreement between methods). Secondly, it had a relatively high acceptance rate (48 %) without compromising narrow measurement CV compared with the criterion method (Table 2). Lastly, it had a low CV of CPET measurements; ~50 % lower CV than that compared with central precision-based QC approaches that form the basis of guideline recommendations. This latter point is of considerable importance for the design and conduct of multicenter clinical trials with CPET measurement outcomes. By applying a z-score-based BioQC method across all centers, we suggest that measurement variability can be reduced by ~50 %, providing an increase in statistical power to detect changes in CPET measurements. Regular BioQC tests are not onerous and, until a larger BioQC data population is established, any CPET laboratory seeking to implement a QC procedure may simply apply the z-score criterion using equation 2, and the study-wide population SD values established in this study (11.0 % and 13.6 % for V̇O2,70W and V̇O2,slope, respectively). While we found the optimal z-score at 0.67, based on distribution normality, the CV of absolute V̇O2 measurements remained close to the minimum up to a z-score of ~0.90, corresponding to ~65 % of all tests and a CV for % predicted below ATS/ACCP guidelines (which occurred at z-score ~1.0) [4]. Further research is required to determine more precisely the optimal z-score within the range of ~0.67 to ~0.90 that balances requirements of normality, minimized CV and the number of accepted tests to inform CPET studies. Nevertheless, while using a combined z-score of 0.67 would minimize CPET measurement variability, any power calculations for future clinical trials should also account for the response variability inherent in the clinical population studied. Thus, the combined z-score approach maximizes statistical discriminatory power within multicenter trials and minimizes laboratory testing burden and study participant risk.

Strategies to minimize measurement variability

We found a relatively high rate of measurement error over 16 months. Importantly, 8 of 15 centers (53 %) required at least one additional validation procedure after an initial failing BioQC test. Each failure triggered a CPET-system troubleshooting process, which included site technicians and central support to identify the error source. Most required involvement of the system manufacturer and eventually led to major equipment service, emphasizing the need for regular maintenance.

In addition, this justifies the recommendation for frequent and rigorously evaluated QC methods in order to prevent large unexplained variability in CPET measurements. The BioQC process used here also identified equipment error prior to equipment failure, and allowed centers to address failing components of CPET systems before trial-related measurements were scheduled. Overall, our results support the view that systematic BioQC is needed to achieve satisfactorily accurate and precise data in multicenter trials employing CPET.

Limitations

A potential limitation relates to the accurate estimation of treadmill WR. External WR is calculated considering a subject’s weight [33]; but does not account for the inertia associated with body movements while walking [25]. These inertial components increase with weight, which may reduce the accuracy of calculated WR and predicted V̇O2 value upon which BioQC is based. A similar phenomenon occurs in cycle ergometry, where V̇O2 is influenced by pedaling frequency. However, a change in treadmill speed between 20 W and 70 W (1.0 mph to 1.8 mph) requires an obligatory cadence increase and thus variable internal work; in cycling, cadence can be effectively controlled [34, 35]. One solution would be to recruit BioQC subjects who are similar in weight to potential trial patients. Another solution would be to use calculations for treadmill WR that incorporate kinetic energy (mv2) instead of momentum (mv).

Another limitation is that we used an equation to predict metabolic rate that was originally developed for cycle ergometry. However, within the speed range used in this study, measured and predicted metabolic rates show strong agreement [25]. V̇O2 prediction was based on a ΔV̇O2/ΔWR of 10.1 mL/min/W [26, 27]. A range of studies support this value, e.g., 10.2 ± 1.0 mL/min/W [28] and 9.9 ± 0.7 mL/min/W [29], although it is recognized that a greater value may be seen in endurance-trained individuals [26]. While, in this study we found that ΔV̇O2/ΔWR averaged 10.6 mL/min/W, this mean was derived from only 15 individuals who performed the BioQC and was within the normal range. We found that post hoc adjustment of the target ΔV̇O2/ΔWR between 10.1 mL/min/W and 10.6 mL/min/W excluded only one additional BioQC test and had no effect on the optimal z-score range. Nevertheless, equations to better calculate treadmill WR to improve accuracy of the ΔV̇O2/ΔWR target, or using exercise modalities such as cycling where WR can be better controlled, would likely further improve precision and accuracy provided by the composite z-score BioQC method.

The BioQC method relies on the attainment of steady-state metabolic responses below the lactate threshold at 70 W. This may require verification by an additional incremental exercise test for non-invasive lactate threshold estimation, and/or the use of a lower WR for less aerobically fit or smaller individuals.

The QC method assesses instrumental measurement precision and accuracy (as opposed to physiologic variability) from submaximal steady-state CPET measurements, because the variability of predicted values for healthy participants is low within this domain. However, clinical trials typically assess both submaximal and maximal values from CPET measurements. Therefore, the instrumental measurement precision determined in this study may not necessarily reflect the instrumental precision of peak measurements, where breathing frequency is greater, and the response times of the gas analyzers become increasingly important. Nevertheless, quality assurance of the integrated CPET measurement system linking the mechanical and metabolic power output within the submaximal domain should also contribute to improving assurance of multicenter instrumental precision and accuracy of maximal CPET measurements.

Conclusions

A central precision-based QC procedure for multicenter studies with CPET as an outcome, reduced measurement variability within center, but was not sufficient to assure between-center measurement accuracy. Thus, we established the distribution of measurements linking mechanical and metabolic power output during multicenter CPET testing, and used this to develop a composite z-score-based method to assess accuracy and precision of CPET measurements. Based on 129 moderate intensity BioQC exercise tests in healthy laboratory staff across 15 centers and 16 months, we found that a composite z-score of 0.67 was able to detect non-normal (outlying) CPET measurements and trigger CPET system troubleshooting, enabling a reduction of multicenter measurement variability by ~50 %. The measurement distribution and z-score method established in this study may be applied to future multicenter studies where CPET variables are measured. Thus, a study-wide precision- and accuracy-based QC process is required to optimize the design, sample size, and conduct of multicenter clinical trials involving CPET measurements.

Availability of data and materials

Not applicable.

Abbreviations

ATS:

American Thoracic Society

ACCP:

American College of Chest Physicians

BioQC:

Biological quality control

BMI:

Body mass index

V̇CO2 :

Carbon dioxide output

CPET:

Cardiopulmonary exercise testing

CV:

Coefficient of variation

ERS:

European Respiratory Society

IRB:

Institutional Review Board

mv2 :

Kinetic energy

VE:

Minute ventilation

mv:

Momentum

V̇O2 :

Oxygen uptake

QC:

Quality control

SD:

Standard deviation

WR:

Work rate

References

  1. Wasserman K, Hansen JE, Sue DY, Stringer WW, Whipp BJ. Clinical applications of cardiopulmonary exercise testing. In: Principles of Exercise Testing and Interpretation Including Pathophysiology and Clinical Applications. 4th ed. Philadelphia: Lippincott Williams & Wilkins; 2005. p. 198–241.

    Google Scholar 

  2. Mezzani A, Corrà U, Giordano A, Colombo S, Psaroudaki M, Giannuzzi P. Upper intensity limit for prolonged aerobic exercise in chronic heart failure. Med Sci Sports Exerc. 2010;42(4):633–9.

    Article  PubMed  Google Scholar 

  3. Myers J, Arena R, Franklin B, Pina I, Kraus WE, McInnis K, et al. Recommendations for clinical exercise laboratories: a scientific statement from the American Heart Association. Circulation. 2009;119(24):3144–61.

    Article  PubMed  Google Scholar 

  4. American Thoracic Society (ATS)/American College of Chest Physicians (ACCP). ATS/ACCP Statement on cardiopulmonary exercise testing. Am J Respir Crit Care Med. 2003;167(2):211–77.

    Article  Google Scholar 

  5. Casaburi R, Porszasz J, Burns MR, Carithers ER, Chang RS, Cooper CB. Physiologic benefits of exercise training in rehabilitation of patients with severe chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 1997;155(5):1541–51.

    Article  PubMed  CAS  Google Scholar 

  6. Emtner M, Porszasz J, Burns M, Somfay A, Casaburi R. Benefits of supplemental oxygen in exercise training in nonhypoxemic chronic obstructive pulmonary disease patients. Am J Respir Crit Care Med. 2003;168(9):1034–42.

    Article  PubMed  Google Scholar 

  7. Palange P, Ward SA, Carlsen KH, Casaburi R, Gallagher CG, Gosselink R, et al. Recommendations on the use of exercise testing in clinical practice. Eur Respir J. 2007;29(1):185–209.

    Article  PubMed  CAS  Google Scholar 

  8. Vogiatzis I. Strategies of muscle training in very severe COPD patients. Eur Respir J. 2011;38(4):971–5.

    Article  PubMed  CAS  Google Scholar 

  9. Maltais F, Celli B, Casaburi R, Porszasz J, Jarreta D, Seoane B, et al. Aclidinium bromide improves exercise endurance and lung hyperinflation in patients with moderate to severe COPD. Respir Med. 2011;105(4):580–7.

    Article  PubMed  Google Scholar 

  10. Neder JA, Fuld JP, Overend T, Thirlwell J, Carter R, Stevenson R, et al. Effects of formoterol on exercise tolerance in severely disabled patients with COPD. Respir Med. 2007;101(10):2056–64.

    Article  PubMed  Google Scholar 

  11. O’Donnell DE, Flüge T, Gerken F, Hamilton A, Webb K, Aguilaniu B, et al. Effects of tiotropium on lung hyperinflation, dyspnoea and exercise tolerance in COPD. Eur Respir J. 2004;23(6):832–40.

    Article  PubMed  Google Scholar 

  12. Porszasz J, Stringer W, Casaburi R. Equipment, measurements and quality control in clinical exercise testing. In: Ward S, Palange P, editors. European Respiratory Monograph, No 40. Volume 12. Sheffield, UK: European Respiratory Society; 2007. p. 108–28.

  13. Lamarra N, Whipp B. Measurement of pulmonary gas exchange. In: Maud PJ, Foster C, editors. Physiological Assessment of Human Fitness. Champaign: Human Kinetics; 1995. p. 19–35.

    Google Scholar 

  14. Revill SM, Morgan MD. Biological quality control for exercise testing. Thorax. 2000;55(1):63–6.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  15. Arena R, Peberdy MA. Short-term reliability of oxygen uptake on-kinetics in apparently healthy subjects. J Cardiopulm Rehabil. 2006;26(4):219–23.

    Article  PubMed  Google Scholar 

  16. Hansen JE, Sun XG, Yasunobu Y, Garafano RP, Gates G, Barst RJ, et al. Reproducibility of cardiopulmonary exercise measurements in patients with pulmonary arterial hypertension. Chest. 2004;126(3):816–24.

    Article  PubMed  Google Scholar 

  17. Johnston KN, Jenkins SC, Stick SM. Repeatability of peak oxygen uptake in children who are healthy. Pediatr Phys Ther. 2005;17(1):11–7.

    Article  PubMed  Google Scholar 

  18. Porszasz J, Woodhouse LI, Casaburi R, Nguyen AHC, Tsurugaya H, Whipp BJ. Use of physiological calibration to validate gas exchange measurements during exercise. Med Sci Sports Exerc. 2004;36(5):S27.

    Google Scholar 

  19. Jones NL, Kane JW. Quality control of exercise test measurements. Med Sci Sports. 1979;11(4):368–72.

    PubMed  CAS  Google Scholar 

  20. Gagnon J, Province MA, Bouchard C, Leon AS, Skinner JS, Wilmore JH, et al. The HERITAGE Family Study: quality assurance and quality control. Ann Epidemiol. 1996;6(6):520–9.

    Article  PubMed  CAS  Google Scholar 

  21. Brawner CA, Ehrman JK, Aldred H, Schairer JR, Keteyian SJ. Quality assurance and cardiopulmonary exercise testing in clinical trials. J Card Fail. 2008;14(4):283–9.

    Article  PubMed  Google Scholar 

  22. O’Donnell DE, Maltais F, Porszasz J, Webb KA, Albers FC, Deng Q, et al. The continuum of physiological impairment during treadmill walking in patients with mild-to-moderate COPD: patient characterization phase of a randomized clinical trial. PloS one. 2014;9(5):e96574.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Casaburi R, Maltais F, Porszasz J, Albers F, Deng Q, Iqbal A, et al. Effects of tiotropium on hyperinflation and treadmill exercise tolerance in mild to moderate chronic obstructive pulmonary disease. Ann Am Thorac Soc. 2014;11(9):1351–61.

    Article  PubMed  Google Scholar 

  24. Blonshine B, Cao R, Casaburi R, Paden H, Porszasz J. Results from a multi-center cardiopulmonary exercise physiological quality control protocol. American Thoracic Society 2012 International Conference, May 18-23, 2012, San Francisco, CA, USA. Am J Resp Crit Care Med, Vol. 185, Meeting Abstracts, 2012. A43 COPD: Functional Assessment and Exercise. p. A1545.

  25. Porszasz J, Casaburi R, Somfay A, Woodhouse LJ, Whipp BJ. A treadmill ramp protocol using simultaneous changes in speed and grade. Med Sci Sports Exerc. 2003;35(9):1596–603.

    Article  PubMed  Google Scholar 

  26. Wasserman K, Hansen J, Sue D, Stringer W, Whipp B. Oxygen cost of work. In: Principles of Exercise Testing and Interpretation Including Pathophysiology and Clinical Applications. 4th ed. Philadelphia: Lippincott Williams and Wilkins; 2005. p. 19.

    Google Scholar 

  27. Wasserman K, Whipp BJ. Exercise physiology in health and disease. Am Rev Respir Dis. 1975;112(2):219–49.

    PubMed  CAS  Google Scholar 

  28. Hansen JE, Sue DY, Wasserman K. Predicted values for clinical exercise testing. Am Rev Respir Dis. 1984;129(2 Pt 2):S49–55.

    Article  PubMed  CAS  Google Scholar 

  29. Wasserman K, Sue D. Coupling of external to cellular respiration. In: Wasserman K, editor. Exercise Gas Exchange in Heart Disease. New York: Futura Publishing; 1996.

    Google Scholar 

  30. Hansen JE, Casaburi R, Cooper DM, Wasserman K. Oxygen uptake as related to work rate increment during cycle ergometer exercise. Eur J Appl Physiol Occup Physiol. 1988;57(2):140–5.

    Article  PubMed  CAS  Google Scholar 

  31. Gibbons RJ, Balady GJ, Bricker JT, Chaitman BR, Fletcher GF, Froelicher VF, et al. ACC/AHA 2002 guideline update for exercise testing: summary article. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Update the 1997 Exercise Testing Guidelines). J Am Coll Cardiol. 2002;40(8):1531–40.

    Article  PubMed  Google Scholar 

  32. American College of Sports Medicine. ACSM’s Guidelines for Exercise Testing and Prescription. 7th ed. Philadelphia: Lippincott Williams & Wilkins; 2006.

    Google Scholar 

  33. Jones NL. Clinical Exercise Testing. 3rd ed. Philadelphia: W.B Saunders; 1988.

    Google Scholar 

  34. Widrick JJ, Freedson PS, Hamill J. Effect of internal work on the calculation of optimal pedaling rates. Med Sci Sports Exerc. 1992;24(3):376–82.

    Article  PubMed  CAS  Google Scholar 

  35. Zoladz JA, Rademaker AC, Sargeant AJ. Human muscle power generating capability during cycling at different pedalling rates. Exp Physiol. 2000;85(1):117–24.

    Article  PubMed  CAS  Google Scholar 

Download references

Acknowledgments

The authors wish to express their indebtedness to all laboratory staff who sacrificed their time and effort to comply with the special requirements, and spent significant time on performing these tests without complaint in order to achieve the highest possible quality in this trial. Furthermore, we would also like to acknowledge all center principal investigators and personnel who contributed their time and effort to this study. This work was supported by Boehringer Ingelheim Pharmaceuticals, Inc. (BIPI) and Pfizer Inc. Writing, editorial support, and formatting assistance was provided by Jane M. Gilbert, BSc, CMPP, of Envision Scientific Solutions, which was contracted, and compensated by BIPI and Pfizer Inc for these services.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Janos Porszasz.

Additional information

Competing interests

Janos Porszasz reports grants, personal fees, and non-financial support from Boehringer Ingelheim, Inc., grants from Actelion, and personal fees from Actelion and Allergan during the conduct of the study. In addition, Dr. Porszasz has a patent US 7,628,732 issued, and a patent US 7,927,251 issued.

Susan Blonshine reports personal fees from Boehringer Ingelheim during the conduct of the study and personal fees from Boehringer Ingelheim outside the submitted work.

Robert Cao was a consultant for Boehringer Ingelheim Pharmaceuticals, Inc.

Heather A Paden is an employee of Boehringer Ingelheim Pharmaceuticals, Inc.

Richard Casaburi reports grants from Los Angeles Biomedical Research Institute and personal fees from Boehringer Ingelheim during the conduct of this study. He reports grants from Boehringer Ingelheim, Forest, GlaxoSmithKline, and Novartis, and personal fees from AstraZeneca, Boehringer Ingelheim, Forest, GlaxoSmithKline, Novartis, and Pfizer outside the submitted work.

Harry B Rossiter declares that he has no competing interests.

Authors’ contributions

JP, SB, HP, RCas, and HBR participated in the conception, hypotheses delineation, and design of the study. JP, SB, and RCao collected the study data. JP, RCao, and HBR performed the statistical analyses. JP, HBR, and RCas interpreted the results. JP, SB, RCas, and HBR drafted the manuscript. All authors reviewed manuscript drafts, and read and approved the final version.

Additional files

Additional file 1: Table S1.

Independent Ethics Committees and Institutional Review Boards for ClinicalTrial.gov: NCT01072396. (PDF 71 kb)

Additional file 2: Table S2.

Equipment used at the study centers. (PDF 18 kb)

Additional file 3:

Staff training, and equipment calibration and verification. (PDF 75 kb)

Additional file 4:

Maintenance and troubleshooting log, and troubleshooting of CPET systems consequent to a failing BioQC test. (PDF 64 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Porszasz, J., Blonshine, S., Cao, R. et al. Biological quality control for cardiopulmonary exercise testing in multicenter clinical trials. BMC Pulm Med 16, 13 (2016). https://doi.org/10.1186/s12890-016-0174-8

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-016-0174-8

Keywords