Skip to main content
  • Research Article
  • Open access
  • Published:

Inspiratory flow profile and usability of the NEXThaler, a multidose dry powder inhaler, in asthma and COPD

Abstract

Background

Inhaler selection is important when managing respiratory conditions; a patient’s inhalation technique should be appropriate for the selected device, and patients should ideally be able to use a device successfully regardless of disease severity. The NEXThaler is a multidose dry-powder inhaler with a breath-actuated mechanism (BAM) and dose counter that activates only following inhalation, so effectively an ‘inhalation counter’. We assessed inspiratory flow through the NEXThaler in two studies and examined whether inhalation triggered the BAM.

Methods

The two studies were open-label, single-arm, and single visit. One study recruited patients with asthma aged ≥ 18 years; the other recruited patients with chronic obstructive pulmonary disease (COPD) aged ≥ 40 years. All patients inhaled twice through a placebo NEXThaler. The inspiratory profile through the device was assessed for each inhalation using acoustic monitoring, with flow at and time to BAM firing, peak inspiratory flow (PIF), and total inhalation time assessed.

Results

A total of 40 patients were enrolled in the asthma study: 20 with controlled asthma and 20 with partly controlled/uncontrolled asthma. All patients were able to trigger the BAM, as evidenced by the inhalation counter activating on closing the device. Mean flow at BAM firing following first inhalation was 35.0 (range 16.3–52.3) L/min; mean PIF was 64.6 (35.0–123.9) L/min. A total of 72 patients were enrolled in the COPD study, with data analysed for 69 (mean forced expiratory volume in 1 s 48.7% predicted [17–92%]). As with the asthma study, all patients, regardless of airflow limitation, were able to trigger the BAM. Mean flow at BAM firing following first inhalation was 41.9 (26.6–57.1) L/min; mean PIF was 68.0 (31.5–125.4) L/min. Device usability was rated highly in both studies, with 5 min sufficient to train the patients, and a click heard shortly after inhalation in all cases (providing feedback on BAM firing).

Conclusions

Inhalation flows triggering the BAM in the NEXThaler were similar between patients with controlled and partly controlled/uncontrolled asthma, and were similar across COPD airflow limitation. All enrolled patients were able to activate the device.

Peer Review reports

Background

Inhaler selection is an important consideration in overall therapy choice when managing respiratory conditions [1, 2], and the choice of device should be tailored to the patient [3]. The ideal inhaler should be breath-activated, should provide feedback that the inhalation manoeuvre has been successful [4], and should indicate the number of remaining doses [5]. In addition, a patient’s inhalation technique should be appropriate for the device [6]. Successful use of a dry-powder inhaler (DPI) requires that a patient can generate sufficient inspiratory flow [6]; ideally patients with all levels of disease severity should be able to use a device successfully.

The NEXThaler is a multidose DPI with a breath-actuated mechanism (BAM) that has an inspiratory flow resistance of 0.036 kPa½ L/min (i.e., medium-to-high resistance). To inhale a dose of medication, the patient only has to open the cover, which makes the dose available. When the patient then inhales through the device, a click is felt or heard as the BAM fires, indicating successful dose release. Then, when the cover is closed, if the dose has been inhaled the counter will count down. If the inhalation was not successful, for example if the patient did not generate sufficient inspiratory flow, just exhaled through the device, or only opened and closed the cover, the counter will not count down, and the dose is not wasted. Data on dose- and flow-independency of the NEXThaler across various inspiratory flows have previously been published [7, 8], as have lung deposition [9] and usability data [10].

The NEXThaler was designed to have an inspiratory flow at BAM firing of 35 L/min. In this manuscript, we describe the results of two studies that assessed the actual inspiratory flow profile generated by patients with asthma and chronic obstructive pulmonary disease (COPD) through the NEXThaler device, and examined whether this manoeuvre triggered the BAM. The studies also evaluated device usability as perceived by the patients.

Methods

Trial design

The two studies were of similar design, being open-label, single-arm, and single visit, with both conducted at a single centre (a specialist investigation unit in Italy). Males or females were eligible. Other than the diagnosis, the main differences in inclusion criteria were age (minimum 18 years for the asthma study; 40 years for the COPD study) and that the COPD study recruited current or ex-smokers (there were no smoking-related criteria in the asthma study). Both studies excluded patients with a diagnosis of any restrictive lung disease. Other main reasons for exclusion from the asthma study were significant seasonal variation in their symptoms, asthma that occurred only during episodic exposure to an allergen, or a history of near-fatal asthma; the main exclusion criterion for the COPD study was a diagnosis of asthma. All patients provided written informed consent prior to any study-related procedure. The studies were approved by an independent ethics committee (Ospedale Maggiore di Parma), and were performed in accordance with the principles of the Declaration of Helsinki, and the International Conference on Harmonisation notes for guidance on Good Clinical Practice (ICH/CPMP/135/95). Both studies were registered in the EudraCT database (asthma study: registration number 2012-000039-22, registered 18 Jun 2012; COPD study: registration number 2013-000262-11, registered 16 May 2013).

At the single study visit, patients were initially trained on correct use using an empty NEXThaler DPI. The instructions for use during this training were as follows. “Hold your inhaler firmly in the upright position. Open the cover fully, check the dose counter window. Before inhaling, breathe out as far as comfortable; do not breathe out through the inhaler. Bring the inhaler up to your mouth and place your lips around the mouthpiece. Do not cover the air vent when holding the inhaler. Do not inhale through the air vent. Take a deep and forceful breath through your mouth. On inhalation check that an audible click is heard. Remove the inhaler from your mouth. Hold your breath for 5–10 s or as long as is comfortable. Breathe out. Move the inhaler back to the upright position and close the cover fully. Check that the counter has gone down by one.” All patients then inhaled twice through a placebo NEXThaler device, with the inhalations separated by a maximum of 5 min. The inspiratory profile through the device was assessed for each inhalation using acoustic monitoring (Sensohaler, Sagentia Inc., Cambridge, UK), with acoustic signals recorded using a condenser microphone placed within the device located so as not to impact the operation of the device, and then analysed by specialised software using a set of algorithms. An additional manoeuvre could be performed if the first manoeuvre was judged unacceptable by the investigator (for example if the microphone disconnected during the inhalation, there was an operator-related error in recording, or artifacts). The acoustic monitoring was used to assess the flow at and time to BAM firing, peak inspiratory flow (PIF), total inhalation time, and total inhaled volume, all of which were automatically calculated by the software. The overall performance of the acoustic monitoring system was validated by comparing the flow measured by the system against a completely independent flow measurement by a calibrated hot-wire anemometer. The estimated flow accuracy was within ± 5 L/min of the actual flow from 30 to 130 L/min and within ± 10 L/min from 0 to 30 L/min. Device usability was assessed by means of a physician-assessed questionnaire that comprised 10 questions (see supplement), each with a yes/no answer, with the responses entered directly in the patient’s case report form.

The primary objective of both studies was to assess the inspiratory profile through the NEXThaler, in patients with different levels of asthma control, or patients with COPD who had varying levels of airflow limitation.

Sample size and statistical methods

There was no formal sample size calculation for either study. For the asthma study, 40 patients were considered to be sufficient to assess inhalation profile through the device, 20 with controlled asthma, and 20 with partly controlled asthma (where ‘controlled’ was defined as occurrence of daytime symptoms and rescue medication use twice or less per week, no limitation of activities, no nocturnal symptoms or awakening, and normal lung function [pre-bronchodilator forced expiratory volume in 1 s [FEV1] or peak expiratory flow [PEF] ≥ 80% predicted]). For the COPD study, the aim was to recruit 10–20 patients with Global Initiative for Chronic Obstructive Lung Disease (GOLD) Grade 1 airflow limitation (post-bronchodilator FEV1 ≥ 80% predicted), and 20 patients with Grades 2 to 4 (Grade 2, 50–80%; Grade 3 30–50%; Grade 4 < 30%). Assuming a screening failure rate of 10%, it was estimated that 44 patients would need to be screened for the asthma study, and 89 for the COPD study. Descriptive statistics are provided for each variable, analysed separately for the first and second inhalation.

Results

Asthma

Participants

A total of 40 patients were enrolled: 20 with controlled asthma and 20 with partly controlled or uncontrolled asthma. The mean ± SD age of these enrolled patients was 43.2 ± 16.5 years (range 18–77 years); 60% were female. Mean FEV1 at screening was 92.5 ± 12.5% predicted (95.1 ± 9.8% in the controlled asthma group and 89.8 ± 14.5% in the partly controlled/uncontrolled group), with mean PEF 91.6 ± 13.1% predicted (95.1 ± 11.3 and 88.0 ± 14.2%, respectively), ranging from 60.9 to 116.0%.

Outcomes

All patients were able to trigger the BAM, as evidenced by the inhalation counter activating on closing the device. Inspiratory flow through the NEXThaler was higher in patients with controlled asthma than in those with partly controlled or uncontrolled asthma, both by individual timepoint (Fig. 1) and by PIF (Table 1). However, mean flows at BAM firing were similar, both for the two groups and the two inhalations (Table 1), and all patients were able to generate a PIF above the value required to fire the BAM (data from the first inhalation are shown in Fig. 2). Furthermore, the mean time to BAM firing was similar in the two groups and for the two inhalations (Table 1). Usability was rated highly, with 5 min sufficient to train all patients on device use, a click heard on opening and soon after inhalation, and the mouthpiece fitting well; the inhalation counter appearance was considered clear by all but one patient (Additional file 1: Table S1). Given the air vent is just next to the mouthpiece on the NEXThaler, we included a question specifically on whether the patients blocked these vents during use; none blocked the vents.

Fig. 1
figure 1

Asthma study: Mean inspiratory flow over time by asthma control (first inhalation). Values on the right of the plot area are the mean flow at BAM firing in patients with controlled (top) and partly controlled/uncontrolled asthma (bottom). BAM breath-actuated mechanism

Table 1 Asthma study: inspiratory flow profile from acoustic monitoring
Fig. 2
figure 2

Asthma study: scatter plot of PIF versus flow at BAM firing (first inhalation). PIF peak inspiratory flow, BAM breath-actuated mechanism

COPD

Participants

A total of 72 patients were enrolled: 21 with post-bronchodilator FEV1 ≥ 80% predicted (GOLD Grade 1); 20 with FEV1 50–80% (Grade 2); 21 with FEV1 30–50% predicted (Grade 3); and 10 with FEV1 < 30% predicted (Grade 4). Three patients were excluded from the analyses (two did not have inhalation profiles available [one GOLD Grade 1 and one Grade 3], and one due to an inaccurate COPD diagnosis [Grade 1]). For the remaining 69 patients, mean ± SD age was 67.9 ± 8.3 years (range 44–80 years), 58 (84.1%) were male, 21 (30.4%) were current smokers, post-bronchodilator FEV1 was 48.7 ± 20.0% predicted (range 17–92%), with post-bronchodilator FEV1 to forced vital capacity ratio 0.52 ± 0.11 (0.22–0.69).

Outcomes

All patients, regardless of airflow limitation, were able to trigger the BAM, as evidenced by the inhalation counter being activated on closing the device. Inspiratory flow through the device was consistent with the degree of airflow limitation, being slightly higher in patients with GOLD Grades 1 and 2 than Grades 3 and 4 (Fig. 3 and Table 2). Importantly, however, all patients had a PIF above the BAM firing value on both inhalations, indicating that they were able to use the device effectively, regardless of airflow limitation (data from the first inhalation are shown in Fig. 4).

Fig. 3
figure 3

COPD study: mean inspiratory flow over time by GOLD Grade (first inhalation). Values on the right of the plot area are the minimum and the maximum mean flows at BAM firing across GOLD Grades. GOLD Global Initiative for Chronic Obstructive Lung Disease, BAM breath-actuated mechanism

Table 2 COPD study: inspiratory flow profile from acoustic monitoring
Fig. 4
figure 4

COPD study: scatter plot of PIF versus flow at BAM firing (first inhalation). PIF peak inspiratory flow, BAM breath-actuated mechanism

The slope of the curves at the start of inhalation (indicating flow acceleration) was steep and consistent across GOLD Grades (Fig. 3). Furthermore, both the flow at and time to BAM firing were consistent across all GOLD Grades, and similar for the first and second inhalations, indicating highly consistent and reproducible device performance, independent of the severity of airflow limitation (Table 2). In the usability questionnaire, in all cases the inhalation counter appearance was clear, a click was heard on opening the inhaler and after drug inhalation and 5 min was sufficient time to train patients in the use of the device (Additional file 1: Table S2).

Discussion

All patients in both studies were able to generate sufficient inspiratory flow to trigger the BAM, with firing occurring with a flow as low as 14.4 L/min in the asthma study and 18.1 L/min in the COPD study. The mean flow at BAM was similar in the two asthma groups, and was similar across the four airflow limitation categories in the COPD study, as was the time to BAM firing, indicating that disease characteristics did not impact the ability to trigger dose delivery in either disease. Furthermore, the mean time to BAM firing was early in the inhalation manoeuvre. Importantly, the inhalation counter activated on closure for all patients, confirming that all patients were able to activate the device, and a click was heard soon after inhalation in all cases, providing additional confirmation of BAM firing.

Interestingly, the inspiratory flows generated through the NEXThaler in the COPD study were overall larger than those in the asthma study. Although this may appear counter-intuitive (and differs from data in two previous studies with different DPIs [11, 12]), there was no requirement for bronchodilator washout prior to the use of the NEXThaler, so comparing the two populations is difficult, as is comparing these data with standard populations. Furthermore, our data are from two different studies (even though the designs of the studies were very similar), and no formal comparisons between the studies have been performed. In addition, the data are the peak flows achieved through the device, rather than peak flows generated through a spirometer, and so the data could be influenced by the instructions for use and the characteristics of the device itself.

The combination of the BAM and the dose counter only being activated following adequate inhalation means that the dose counter on the NEXThaler is effectively an ‘inhalation counter’ providing feedback on successful dose release. This is a key distinction between the NEXThaler and the Ellipta and the Turbuhaler multi-dose DPIs [13]. In these other devices, the inhalation counter is activated on priming, so that if the patient does not use the dose (either deliberately or accidentally) it is wasted [14, 15]—but more importantly this does not give any indication of correct use. In addition, although we did not evaluate dose delivery in the current studies (given we used placebo devices), a previous in-vitro study has demonstrated flow independency through the NEXThaler for inspiratory flows ranging from 30 to 90 L/min (at inhalation volumes of 2 and 4 L), with delivered doses remaining with a variation largely within ± 15% of the specification value (although the fine particle fraction, expressed as a percentage of the delivered dose, and the fine particle mass both increased with increasing flow) [8]. This is an important consideration given the impact of disease characteristics on the mean inspiratory flow that we observed in these two studies, and given we observed a range in PIF of 35.0–117.6 L/min in the asthma study and 31.5–135.9 L/min in the COPD study. A similar impact of disease characteristics on inspiratory flow through DPIs has been observed at least one other study [16], whereas two studies in patients with COPD showed no clear correlation using the Diskus, Ellipta, HandiHaler and Easyhaler DPIs [17, 18]. These contrasting results emphasise the importance of flow independency of dose delivery, as it is potentially difficult to predict the flow that a patient can generate through a device purely based on their disease characteristics.

The usability assessments in these two studies were consistent with the results of a previous study conducted in patients with asthma, in which the usability of the NEXThaler was compared with the Diskus and Turbuhaler [10]. In this previous study, in addition to overall ease of use the NEXThaler was rated by patients as superior to the two other DPIs in terms of the time to set up, and to read the instructions for use, and the proportion of participants who completed an error-free successful inhalation was significantly higher for the NEXThaler. We acknowledge that the data from these two studies are from evaluations performed immediately after patients were trained on correct device use. These data therefore don’t necessarily reflect ‘real world’ use. However, given the dose counter only activates following successful inhalation, if a patient did not generate sufficient inspiratory flow during a manoeuvre this will be immediately apparent. The wide variability of flow at BAM firing could be because activation occurred early in the inhalation manoeuvre (after a mean of 0.07 s in the asthma study and 0.13–0.16 s in the COPD study), and which is at the steepest part of the flow–time curve when the flow is increasing rapidly. In addition, given the data were captured after patients had used the device only for the second or third occasion (assuming they only using the training device once), it is possible that the sound of the device actuating may have influenced their manoeuvre.

Conclusions

The NEXThaler multi-dose DPI has a BAM with a dose counter that activates solely following inhalation (so effectively an inhalation counter). The two studies demonstrated that patients with controlled and partly controlled/uncontrolled asthma, and those with a range of severities of COPD were able to trigger the BAM and consequently activate the inhalation counter, thus receiving feedback on inhalation.

Availability of data and materials

Chiesi commits to sharing with qualified scientific and medical researchers, conducting legitimate research, the anonymised patient-level and study-level data, the clinical protocol and the full clinical study report of Chiesi Farmaceutici SpA-sponsored interventional clinical trials in patients for medicines and indications approved by the European Medicines Agency and/or the US Food and Drug Administration after 1st January 2015, following the approval of any received research proposal and the signature of a Data Sharing Agreement. Chiesi provides access to clinical trial information consistently with the principle of safeguarding commercially confidential information and patient privacy. Other information on Chiesi’s data sharing commitment, access and research request’s approval process are available in the Clinical Trial Transparency section of http://www.chiesi.com/en/research-and-development/, including the clinical trial data request portal.

Abbreviations

BAM:

Breath-actuated mechanism

COPD:

Chronic obstructive pulmonary disease

DPI:

Dry-powder inhaler

FEV1 :

Forced expiratory volume in 1 s

GOLD:

Global Initiative for Chronic Obstructive Lung Disease

PEF:

Peak expiratory flow

PIF:

Peak inspiratory flow

References

  1. Global Initiative for Asthma. Global strategy for asthma management and prevention. 2020. https://ginasthma.org/gina-reports/. Accessed 8 Sep 2020.

  2. Global Initiative for Chronic Obstructive Lung Disease. Global strategy for the diagnosis, management, and prevention of chronic obstructive pulmonary disease. 2021. https://goldcopd.org/2021-gold-reports/. Accessed 25 Nov 2020.

  3. Dekhuijzen PNR, Vincken W, Virchow JC, Roche N, Agusti A, Lavorini F, et al. Prescription of inhalers in asthma and COPD: towards a rational, rapid and effective approach. Respir Med. 2013;107:1817–21. https://doi.org/10.1016/j.rmed.2013.09.013.

    Article  CAS  PubMed  Google Scholar 

  4. Virchow JC. What plays a role in the choice of inhaler device for asthma therapy? Curr Med Res Opin. 2005;21(Suppl 4):S19-25. https://doi.org/10.1185/030079905X61758.

    Article  CAS  PubMed  Google Scholar 

  5. Navaie M, Dembek C, Cho-Reyes S, Yeh K, Celli BR. Inhaler device feature preferences among patients with obstructive lung diseases: a systematic review and meta-analysis. Medicine (Baltimore). 2020;99:e20718. https://doi.org/10.1097/MD.0000000000020718.

    Article  Google Scholar 

  6. Laube BL, Janssens HM, de Jongh FHC, Devadason SG, Dhand R, Diot P, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J. 2011;37:1308–31. https://doi.org/10.1183/09031936.00166410.

    Article  CAS  PubMed  Google Scholar 

  7. Buttini F, Pasquali I, Brambilla G, Copelli D, Alberi MD, Balducci AG, et al. Multivariate analysis of effects of asthmatic patient respiratory profiles on the in vitro performance of a reservoir multidose and a capsule-based dry powder inhaler. Pharm Res. 2016;33:701–15. https://doi.org/10.1007/s11095-015-1820-1.

    Article  CAS  PubMed  Google Scholar 

  8. Buttini F, Brambilla G, Copelli D, Sisti V, Balducci AG, Bettini R, et al. Effect of flow rate on in vitro aerodynamic performance of NEXThaler® in comparison with Diskus® and Turbohaler® dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2016;29:167–78. https://doi.org/10.1089/jamp.2015.1220.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Mariotti F, Sergio F, Acerbi D, Meyer T, Herpich C. Lung deposition of the extra fine dry powder fixed combination beclomethasone dipropionate plus formoterol fumarate via the NEXT DPI® in healthy subjects, asthmatic and COPD patients. Eur Respir J. 2011;38(Suppl 55):P830.

    Google Scholar 

  10. Voshaar T, Spinola M, Linnane P, Campanini A, Lock D, Lafratta A, et al. Comparing usability of NEXThaler® with other inhaled corticosteroid/long-acting β2-agonist fixed combination dry powder inhalers in asthma patients. J Aerosol Med Pulm Drug Deliv. 2014;27:363–70. https://doi.org/10.1089/jamp.2013.1086.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Azouz W, Chetcuti P, Hosker HSR, Saralaya D, Stephenson J, Chrystyn H. The inhalation characteristics of patients when they use different dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2015;28:35–42. https://doi.org/10.1089/jamp.2013.1119.

    Article  PubMed  Google Scholar 

  12. Azouz W, Chetcuti P, Hosker H, Saralaya D, Chrystyn H. Inhalation characteristics of asthma patients, COPD patients and healthy volunteers with the Spiromax® and Turbuhaler® devices: a randomised, cross-over study. BMC Pulm Med. 2015;15:47. https://doi.org/10.1186/s12890-015-0043-x.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Farkas Á, Lewis D, Church T, Tweedie A, Mason F, Haddrell AE, et al. Experimental and computational study of the effect of breath-actuated mechanism built in the NEXThaler ® dry powder inhaler. Int J Pharm. 2017;533:225–35. https://doi.org/10.1016/j.ijpharm.2017.09.057.

    Article  CAS  PubMed  Google Scholar 

  14. GlaxoSmithKline UK. Summary of product characteristics: Relvar Ellipta 92 micrograms/22 micrograms inhalation powder, pre-dispensed. Uxbridge, UK; 2018.

  15. AstraZeneca UK Limited. Summary of product characteristics: Symbicort Turbohaler 200/6 Inhalation powder. Luton, UK; 2020.

  16. Al-Showair RAM, Tarsin WY, Assi KH, Pearson SB, Chrystyn H. Can all patients with COPD use the correct inhalation flow with all inhalers and does training help? Respir Med. 2007;101:2395–401. https://doi.org/10.1016/j.rmed.2007.06.008.

    Article  PubMed  Google Scholar 

  17. Ghosh S, Pleasants RA, Ohar JA, Donohue JF, Drummond MB. Prevalence and factors associated with suboptimal peak inspiratory flow rates in COPD. Int J Chron Obstruct Pulmon Dis. 2019;14:585–95. https://doi.org/10.2147/COPD.S195438.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Malmberg LP, Rytilä P, Happonen P, Haahtela T. Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters. Int J Chron Obstruct Pulmon Dis. 2010;5:257–62. https://doi.org/10.2147/copd.s11474.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Writing support was provided by David Young of Young Medical Communications and Consulting Ltd. This support was funded by Chiesi Farmaceutici SpA.

Funding

These studies were funded by Chiesi Farmaceutici SpA.

Author information

Authors and Affiliations

Authors

Contributions

The studies were designed by AC and MS, with data collected by AC. AC, AY, MS, SB, DG and PNRD interpreted the data, and critically revised the manuscript for important intellectual content. All authors read and approved the final manuscript. 

Corresponding author

Correspondence to Alfredo Chetta.

Ethics declarations

Ethics approval and consent to participate

Both studies were approved by an independent ethics committee (Ospedale Maggiore di Parma), and were performed in accordance with the principles of the Declaration of Helsinki, and the International Conference on Harmonisation notes for guidance on Good Clinical Practice (ICH/CPMP/135/95). All patients provided written informed consent prior to any study-related procedure.

Consent for publication

Not applicable.

Competing interests

AC has received grants from Menarini, GlaxoSmithKline and Astra Zeneca, and personal fees from Chiesi. AY has received grants from MSD, AstraZeneca and Sanofi, speakers bureau honoraria from AstraZeneca, Abdi İbrahim, GlaxoSmithKline, Novartis Sandoz, Sanofi, Chiesi, and DEVA Pharmaceutique, and funding from WHO, the Turkish Ministry of Health, GARD, GINA, and ERS. MS, SB and DG are employees of Chiesi Farmaceutici SpA, the sponsor of the two studies. PNRD has received grants and personal fees from Chiesi, AstraZeneca, GlaxoSmithKline, Glenmark, Mylan, Teva, Mundipharma, and Boehringer Ingelheim.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Device usability questionnaire and results.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chetta, A., Yorgancioglu, A., Scuri, M. et al. Inspiratory flow profile and usability of the NEXThaler, a multidose dry powder inhaler, in asthma and COPD. BMC Pulm Med 21, 65 (2021). https://doi.org/10.1186/s12890-021-01430-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12890-021-01430-9

Keywords